Skip to content

Polaris™ Ultra M0061921310Recalls & reported problems

Boston Scientific Corporation · Ureteral Stent · Class 2 · Implant

All models of Polaris Ultra (40)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 08714729124405), received between Dec 19, 2017 and Sep 23, 2024 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Material Deformation2
  • Detachment of Device or Device Component1
  • Device Contaminated During Manufacture or Shipping1
  • Difficult to Advance1
  • Entrapment of Device1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING PREPARATION, WHEN THE PACKAGE WAS OPENED, THE POLARIS ULTRA URETERAL STENT WAS FOUND BROKEN. ANOTHER OF THE SAME STENT WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED. THE CONDITION OF THE PATIENT AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20285800 ↗
  • MalfunctionApr 15, 2024Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS ULTRA URETERAL STENT WAS OPENED. THE EVENT DATE WAS NOT REPORTED. IT WAS REPORTED THAT WHEN THE DEVICE WAS UNPACKED, THE STENT WAS FOUND BROKEN. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19111428 ↗
  • MalfunctionMar 23, 2024Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS ULTRA URETERAL STENT WAS OPENED TO BE USED DURING A URETEROSCOPY PROCEDURE PERFORMED ON (B)(6) 2024. DURING PREPARATION. THE DEVICE TIP WAS FOUND STICKY. IT WAS NOTED THAT THE PACKAGING WAS NORMAL. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER POLARIS ULTRA URETERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18965351 ↗
  • MalfunctionFeb 14, 2023Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS ULTRA URETERAL STENT WAS USED DURING A ENDOSCOPIC COMBINED INTRA RENAL SURGERY PROCEDURE IN THE URINARY TRACT PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, IT WAS NOTED THAT THE STENT COULD NOT BE INSERTED THROUGH TRANSURETHRAL. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. INVESTIGATION RESULTS REVEALED THAT THE STENT WAS BUCKLED, INSIDE THE PATIENT, THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16375899 ↗
  • MalfunctionOct 14, 2022Entrapment of Device; Material Separation; Entrapment of Device; Material Separation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS ULTRA URETERAL STENT WAS USED DURING A URETERAL STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2022. DURING WITHDRAWAL OF THE STENT PUSHER, THE RADIO OPAQUE (RO) MARKER BAND ATTACHED AT THE TIP OF THE STENT PUSHER FELL OFF. THE RO MARKER BAND GOT STUCK IN THE WORKING CHANNEL OF THE CYSTOSCOPE. AFTER THE PROCEDURE, THE RO MARKER BAND WAS ABLE TO BE REMOVED FROM THE SCOPE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15611144 ↗
  • MalfunctionNov 23, 2020Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS ULTRA URETERAL STENT WAS USED DURING A URETEROSCOPY PROCEDURE IN THE RIGHT URETER, PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, OUTSIDE THE PATIENT, WHEN THE PHYSICIAN PULLED THE SUTURE FROM THE BLADDER COIL, IT WAS NOTICED THAT THE STENT SHAFT BROKE. ANOTHER POLARIS ULTRA STENT WAS OPENED AND SUCCESSFULLY COMPLETED THE PROCEDURE. PHOTO OF THE COMPLAINT DEVICE WAS PROVIDED BY THE COMPLAINANT AND SHOWED THE COIL DEFORMED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10885440 ↗
  • MalfunctionNov 27, 2019Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS ULTRA URETERAL STENT WAS USED DURING A URETEROSCOPY PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, THE STENT SHAFT WAS FRACTURED WHILE THE STENT WAS BEING PLACED OVER THE GUIDEWIRE. THE PROCEDURE WAS COMPLETED WITH ANOTHER POLARIS ULTRA URETERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9386442 ↗
  • MalfunctionDec 19, 2017Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS¿ ULTRA STENT WAS TO BE IMPLANTED IN AN URETEROSCOPIC LITHOTRIPSY (URSL) PROCEDURE PERFORMED IN THE UTERINE ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, THE PHYSICIAN NOTED THAT THE STENT WAS BROKEN INTO TWO PIECES ALONG THE SHAFT. THIS DEVICE WAS NOT USED IN THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER POLARIS¿ ULTRA STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7128927 ↗

Other STENT, URETERAL devices

All STENT, URETERAL devices →