Polaris™ Ultra M0061921350Recalls & reported problems
Boston Scientific Corporation · Ureteral Stent · Class 2 · Implant
All models of Polaris Ultra (40)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 08714729124498), received between May 3, 2018 and Oct 2, 2025 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation4
- Break2
- Deformation Due to Compressive Stress2
- Difficult to Advance2
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 2, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE PREPARATION, THE STENT WAS ALREADY BROKEN WHEN THE PACKAGING WAS OPENED JUST BEFORE USE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THERE WERE NO KNOWN PATIENT COMPLICATIONS REPORTED. NOTE: ADDITIONAL INFORMATION WAS RECEIVED FROM THE COMPLAINANT ON OCTOBER 16, 2025, THE REPORTED EVENT OF STENT COIL DETACHED OCCURRED OUTSIDE THE PATIENT. THEREFORE, THIS IS CONSIDERED NON-REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23197352 ↗MalfunctionAug 18, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING INSERTION, THE PHYSICIAN ATTEMPTED TO PLACE EACH POLARIS ULTRA URETERAL STENT, BUT THE COIL TIP WAS CRUMPLED AND MALFORMED. THE THIRD STENT ATTEMPTED AND COMPLETED WITH INSERTION JUST FINE. TWO SEPARATE LOTS, WITH THE SAME ISSUE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22830430 ↗MalfunctionAug 18, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING INSERTION, THE PHYSICIAN ATTEMPTED TO PLACE EACH POLARIS ULTRA URETERAL STENT, BUT THE COIL TIP WAS CRUMPLED AND MALFORMED. THE THIRD STENT ATTEMPTED AND COMPLETED WITH INSERTION JUST FINE. TWO SEPARATE LOTS, WITH THE SAME ISSUE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22829624 ↗MalfunctionFeb 7, 2024Deformation Due to Compressive Stress; Difficult to Advance; Deformation Due to Compressive Stress; Difficult to Advance
NOTE: THIS REPORT IS ONE OF TWO COMPLAINTS THAT PERTAIN TO THE SAME EVENT (MFR REPORT # 2124215-2024-04553 AND MFR. REPORT # 2124215-2024-04544). IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS ULTRA URETERAL STENT WAS USED DURING A PERCUTANEOUS NEPHROLITHOTOMY PROCEDURE IN THE URETER PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE STENT WAS ATTEMPTED TO PLACE OVER THE WIRE, BUT THEY COULD NOT GET IT THROUGH THE URETER. IT WAS NOTED THAT THE PATIENT HAD A VERY NARROW URETEROPELVIC JUNCTION AND THE STENT WAS FOUND BUCKLED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER POLARIS ULTRA URETERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18660326 ↗MalfunctionFeb 7, 2024Deformation Due to Compressive Stress; Difficult to Advance; Deformation Due to Compressive Stress; Difficult to Advance
NOTE: THIS REPORT IS ONE OF TWO COMPLAINTS THAT PERTAIN TO THE SAME EVENT (MFR REPORT # 2124215-2024-04553 AND MFR. REPORT # 2124215-2024-04544). IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS ULTRA URETERAL STENT WAS USED DURING A PERCUTANEOUS NEPHROLITHOTOMY PROCEDURE IN THE URETER PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE STENT WAS ATTEMPTED TO PLACE OVER THE WIRE, BUT THEY COULD NOT GET IT THROUGH THE URETER. IT WAS NOTED THAT THE PATIENT HAD A VERY NARROW URETEROPELVIC JUNCTION AND THE STENT WAS FOUND BUCKLED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER POLARIS ULTRA URETERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18660325 ↗MalfunctionJul 12, 2023Material Deformation; Adverse Event Without Identified Device or Use Problem; Material Deformation; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS ULTRA URETERAL STENT WAS USED DURING A URETEROSCOPY PROCEDURE, PERFORMED ON (B)(6), 2023. DURING THE PROCEDURE, WHEN THE STENT WAS PUSHED USING A POSITIONER, IT WAS NOTICED THAT THE TAIL OF THE STENT BUCKLED AND IT FAILED TO BE INSERTED. ANOTHER POLARIS ULTRA URETERAL STENT WAS OPENED AND USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. ADDITIONAL INFORMATION RECEIVED ON JULY 19. 2023 STATING THAT THE TARGET ANATOMY LOCATION FOR THIS PROCEDURE WAS THE KIDNEY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17305631 ↗MalfunctionAug 18, 2021Material Deformation; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS ULTRA URETERAL STENT WAS USED DURING A URETERAL STENTING PROCEDURE IN THE RIGHT URETER PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, WHEN THE PHYSICIAN PUSHED THE STENT USING THE POSITIONER, THE STENT COIL BECAME BUCKLED AND WAS UNABLE TO INSERT THE STENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER POLARIS ULTRA URETERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12338803 ↗MalfunctionMay 3, 2018Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS¿ ULTRA STENT WAS TO BE USED FOR A PROCEDURE PERFORMED IN THE URETER ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING UNPACKING, THE STENT WAS NOTICED TO BE BROKEN INTO TWO PIECES ALONG THE SHAFT INSIDE OF THE PACKAGING. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7485507 ↗
Other STENT, URETERAL devices
- Bard® InLay Optima® Ureteral Stent 788426C. R. Bard, Inc. · STENT, URETERAL1 recall100 reports
- Bard® InLay Optima® Ureteral Stent 788626C. R. Bard, Inc. · STENT, URETERAL122 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777626C. R. Bard, Inc. · STENT, URETERAL118 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777426C. R. Bard, Inc. · STENT, URETERAL108 reports
- Bard® InLay® Ureteral Stent 778626C. R. Bard, Inc. · STENT, URETERAL100 reports
- Percuflex™ Plus M0061752530Boston Scientific Corporation · STENT, URETERAL1 death99 reports
- Contour VL™ M0061801560Boston Scientific Corporation · STENT, URETERAL1 recall45 reports
- Polaris™ Ultra M0061921330Boston Scientific Corporation · STENT, URETERAL71 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page