Percuflex™ M001245500Recalls & reported problems
Boston Scientific Corporation · Ureteral Stent System · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729178613), received between Jan 17, 2024 and Sep 25, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation2
- Break1
- Difficult to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2025Material Separation; Difficult to Advance; Material Separation; Difficult to Advance
IT WAS REPORTED THAT DEVICE SEPARATION OCCURRED. A PERCUFLEX? DEVICE WAS SELECTED FOR USE DURING A NEPHROSTOMY INSERTION AND/OR URETERIC STENT INSERTION PROCEDURE. DURING THE PROCEDURE, WHEN THE STENT WAS INSERTED OVER THE WIRE, IT WAS DIFFICULT TO ADVANCE. THE THREAD WAS PULLED IN AN ATTEMPT TO REPOSITION THE STENT TOWARDS THE INSERTION SITE; HOWEVER, THE STRING SNAPPED OFF THE STENT (WHICH REMAINED INTACT), AND THE STENT BECAME DEPLOYED IN A LESS-THAN-IDEAL POSITION. GIVEN THE STRICTURE, THE PHYSICIAN FELT THAT THE STENT SHOULD NOT MIGRATE. THE PATIENT WOULD RETURN FOR A SECOND PROCEDURE TO ASSESS THE STENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23151490 ↗InjuryJan 17, 2024Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT THE STENT BROKE OFF AND WAS REMOVED WITH A SNARE. THE TARGET LESION WAS IN THE LEFT KIDNEY. A PERCUFLEX TEMPERATURE ABLATION CATHETER URETERAL STENT WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE STENT WAS PLACED IN A NORMAL FASHION. THE STABILIZER WAS BEING REMOVED WHEN THE PROXIMAL LOOP BROKE OFF A FEW CENTIMETERS FROM THE TAPERED END. THE TWO PIECES OF THE DEVICE WERE IN THE PATIENT AT THIS POINT. THE DOCTOR DECIDED TO ATTEMPT TO RETRIEVE THE PIECES THE FOLLOWING DAY, SO A NEPHROSTOMY TUBE WAS KEPT IN PLACE SO THAT THEY HAD ACCESS FOR THE RETRIEVAL PROCEDURE AND THE BROKEN DEVICE WAS RETRIEVED VIA SNARE THE NEXT DAY. HENCE, THE PHYSICIAN FELT THAT THE TUBE BROKE FROM THE PULLING OF THE STRING. THE PROCEDURE WAS COMPLETED WITH ANOTHER PERCUFLEX STENT. THE PATIENT HAS FULLY RECOVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18527609 ↗
Other STENT, URETERAL devices
- Bard® InLay Optima® Ureteral Stent 788426C. R. Bard, Inc. · STENT, URETERAL1 recall100 reports
- Bard® InLay Optima® Ureteral Stent 788626C. R. Bard, Inc. · STENT, URETERAL122 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777626C. R. Bard, Inc. · STENT, URETERAL118 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777426C. R. Bard, Inc. · STENT, URETERAL108 reports
- Bard® InLay® Ureteral Stent 778626C. R. Bard, Inc. · STENT, URETERAL100 reports
- Percuflex™ Plus M0061752530Boston Scientific Corporation · STENT, URETERAL1 death99 reports
- Contour VL™ M0061801560Boston Scientific Corporation · STENT, URETERAL1 recall45 reports
- Polaris™ Ultra M0061921330Boston Scientific Corporation · STENT, URETERAL71 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.