Skip to content

Encore™ 26 M001151050Recalls & reported problems

Boston Scientific Corporation · Inflation Device · Class 2

All models of Encore 26 (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Boston Scientific Encore 26 inflation device is used with balloon dilation catheters to inflate, monitor pressure in, and deflate the balloon. Most of its 54 reports describe contamination inside sealed packaging, damaged packaging that could compromise sterility, and pressure readings that could not be detected. No deaths and no injuries were reported. One recall, started in August 2025 and still open, concerns foreign material particles that can migrate from the device into a balloon catheter.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
54
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 08714729183624), received between Sep 20, 2017 and Mar 31, 2026 — classified as 54 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (16 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 11 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 5, 2025Open, Classified

    The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.

    Z-2533-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Contamination with Chemical or Other Material16
  • Display or Visual Feedback Problem12
  • Packaging Problem12
  • Unsealed Device Packaging9
  • Contamination7
  • Difficult to Open or Remove Packaging Material3
  • Inflation Problem3
  • Physical Resistance/Sticking3

Patient problems

  • No Clinical Signs, Symptoms or Conditions39
  • No Consequences Or Impact To Patient10
  • No Known Impact Or Consequence To Patient2
  • No Patient Involvement2
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 31, 2026Unsealed Device Packaging; Packaging Problem

    IT WAS REPORTED THAT DEVICE CONTAMINATION OCCURRED. AN ENCORE INFLATION DEVICE WAS SELECTED FOR USE. PACKAGING WAS DAMAGED DURING TRANSPORT. THE INNER POUCH WAS ALSO DAMAGED AND THE STERILITY OF THE DEVICE WAS COMPROMISED. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24743766 ↗
  • MalfunctionDec 22, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BALLOON PRESSURE COULD NOT BE DETECTED. THE TARGET LESION WAS LOCATED ON THE CAROTID ARTERY. A ENCORE 26 MT SINGLE 15105 WAS SELECTED FOR CAROTID ARTERY STENOSIS PROCEDURE. DURING THE PROCEDURE, IT WAS FOUND THAT THE BALLOON PRESSURE COULD NOT BE DETECTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT WAS STABLE AFTER THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23872913 ↗
  • MalfunctionDec 18, 2025Contamination; Contamination

    IT WAS REPORTED THAT A FOREIGN MATERIAL WAS PRESENT IN THE DEVICE. DURING RESTOCK, AN ENCORE 26 DEVICE WAS FOUND CONTAMINATED - WITH 2 HAIRS VISIBLE WITHIN THE SEALED, UNOPENED PACKAGE. THERE WAS NO PATIENT INVOLVEMENT, AND THE DEVICE WAS NOT USED IN ANY PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23842377 ↗
  • MalfunctionOct 31, 2025Unsealed Device Packaging; Packaging Problem; Unsealed Device Packaging; Packaging Problem

    IT WAS REPORTED THAT PACKAGING SEAL WAS COMPROMISED. AN ENCORE 26 PERIPHERAL INTERVENTION INFLATION DEVICE WAS SELECTED FOR USE. PRIOR PROCEDURE, IT WAS NOTED THAT THE PROTECTIVE PAPER TEARS EASILY DUE TO THE BLISTER PACKAGE OF THE BOSTON SCIENTIFIC INFLATION PUMPS. THERE WAS NO VISIBLE DAMAGE TO THE OUTER PACKAGING, HOWEVER, A POTENTIAL RISK OF STERILITY BEING COMPROMISED DUE TO THE TORN PAPER LAYER THAT MAY HAVE COME INTO CONTACT WITH BOTH THE OUTER AND INNER SURFACES OF THE PACKAGING AND THE DEVICE ITSELF. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23437178 ↗
  • MalfunctionOct 29, 2025Contamination; Packaging Problem; Contamination; Packaging Problem

    IT WAS REPORTED THAT A STERILITY ISSUE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. AN ENCORE 26 INFLATION DEVICE WAS SELECTED FOR DEBULK PROCEDURE. PRIOR TO THE PROCEDURE, THE PACKAGING DID NOT OPEN PROPERLY, AND THE STERILITY OF THE DEVICE WAS COMPROMISED. THE TEAR TAB REMAINED INTACT. NO ISSUES WERE NOTED WITH THE PACKAGING PRIOR TO OPENING, AND NO PHYSICAL DAMAGE WAS OBSERVED. THE DEVICE WAS NOT USED DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE DEVICE WITH NO REPORTED PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23415781 ↗
  • MalfunctionOct 29, 2025Contamination; Packaging Problem; Contamination; Packaging Problem

    IT WAS REPORTED THAT A STERILITY ISSUE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. AN ENCORE 26 INFLATION DEVICE WAS SELECTED FOR DEBULK PROCEDURE. PRIOR TO THE PROCEDURE, THE PACKAGING DID NOT OPEN PROPERLY, AND THE STERILITY OF THE DEVICE WAS COMPROMISED. THE TEAR TAB REMAINED INTACT. NO ISSUES WERE NOTED WITH THE PACKAGING PRIOR TO OPENING, AND NO PHYSICAL DAMAGE WAS OBSERVED. THE DEVICE WAS NOT USED DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE DEVICE WITH NO REPORTED PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23415775 ↗
  • MalfunctionOct 24, 2025Unsealed Device Packaging; Packaging Problem; Unsealed Device Packaging; Packaging Problem

    IT WAS REPORTED THAT PACKAGING SEAL WAS COMPROMISED. THE TARGET LESION WAS LOCATED BELOW THE KNEE. AN ENCORE 26 PERIPHERAL INTERVENTION INFLATION DEVICE WAS SELECTED FOR USE. PRIOR PROCEDURE, IT WAS NOTED THAT THE PROTECTIVE PAPER TEARS EASILY AND SOME PUMP WAS STICKING TO THE PLASTIC BLISTER. THERE WERE NO VISIBLE DAMAGE TO THE OUTER PACKAGING. HOWEVER, A POTENTIAL RISK OF STERILITY BEING COMPROMISED DUE TO THE TORN PAPER LAYER THAT MAY HAVE COME INTO CONTACT WITH BOTH THE OUTER AND INNER SURFACES OF THE PACKAGING AND THE DEVICE ITSELF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23381438 ↗
  • MalfunctionOct 24, 2025Unsealed Device Packaging; Packaging Problem; Unsealed Device Packaging; Packaging Problem

    IT WAS REPORTED THAT PACKAGING SEAL WAS COMPROMISED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. AN ENCORE 26 PERIPHERAL INTERVENTION INFLATION DEVICE WAS SELECTED FOR USE. PRIOR PROCEDURE, IT WAS NOTED THAT THE PROTECTIVE PAPER TEARS EASILY AND SOME PUMP WAS STICKING TO THE PLASTIC BLISTER. THERE WERE NO VISIBLE DAMAGE TO THE OUTER PACKAGING. HOWEVER, A POTENTIAL RISK OF STERILITY BEING COMPROMISED DUE TO THE TORN PAPER LAYER THAT MAY HAVE COME INTO CONTACT WITH BOTH THE OUTER AND INNER SURFACES OF THE PACKAGING AND THE DEVICE ITSELF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23379258 ↗
  • MalfunctionOct 24, 2025Unsealed Device Packaging; Packaging Problem; Unsealed Device Packaging; Packaging Problem

    IT WAS REPORTED THAT PACKAGING SEAL WAS COMPROMISED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. AN ENCORE 26 PERIPHERAL INTERVENTION INFLATION DEVICE WAS SELECTED FOR USE. PRIOR PROCEDURE, IT WAS NOTED THAT THE PROTECTIVE PAPER TEARS EASILY AND SOME PUMP WAS STICKING TO THE PLASTIC BLISTER. THERE WERE NO VISIBLE DAMAGE TO THE OUTER PACKAGING. HOWEVER, A POTENTIAL RISK OF STERILITY BEING COMPROMISED DUE TO THE TORN PAPER LAYER THAT MAY HAVE COME INTO CONTACT WITH BOTH THE OUTER AND INNER SURFACES OF THE PACKAGING AND THE DEVICE ITSELF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23379256 ↗
  • MalfunctionOct 24, 2025Unsealed Device Packaging; Packaging Problem; Unsealed Device Packaging; Packaging Problem

    IT WAS REPORTED THAT PACKAGING SEAL WAS COMPROMISED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. AN ENCORE 26 PERIPHERAL INTERVENTION INFLATION DEVICE WAS SELECTED FOR USE. PRIOR PROCEDURE, IT WAS NOTED THAT THE PROTECTIVE PAPER TEARS EASILY AND SOME PUMP WAS STICKING TO THE PLASTIC BLISTER. THERE WERE NO VISIBLE DAMAGE TO THE OUTER PACKAGING. HOWEVER, A POTENTIAL RISK OF STERILITY BEING COMPROMISED DUE TO THE TORN PAPER LAYER THAT MAY HAVE COME INTO CONTACT WITH BOTH THE OUTER AND INNER SURFACES OF THE PACKAGING AND THE DEVICE ITSELF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23379255 ↗
  • MalfunctionOct 24, 2025Unsealed Device Packaging; Packaging Problem; Unsealed Device Packaging; Packaging Problem

    IT WAS REPORTED THAT PACKAGING SEAL WAS COMPROMISED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. AN ENCORE 26 PERIPHERAL INTERVENTION INFLATION DEVICE WAS SELECTED FOR USE. PRIOR PROCEDURE, IT WAS NOTED THAT THE PROTECTIVE PAPER TEARS EASILY AND SOME PUMP WAS STICKING TO THE PLASTIC BLISTER. THERE WERE NO VISIBLE DAMAGE TO THE OUTER PACKAGING. HOWEVER, A POTENTIAL RISK OF STERILITY BEING COMPROMISED DUE TO THE TORN PAPER LAYER THAT MAY HAVE COME INTO CONTACT WITH BOTH THE OUTER AND INNER SURFACES OF THE PACKAGING AND THE DEVICE ITSELF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23379252 ↗
  • MalfunctionOct 24, 2025Unsealed Device Packaging; Packaging Problem; Unsealed Device Packaging; Packaging Problem

    IT WAS REPORTED THAT PACKAGING SEAL WAS COMPROMISED. THE TARGET LESION WAS LOCATED BELOW THE KNEE. AN ENCORE 26 PERIPHERAL INTERVENTION INFLATION DEVICE WAS SELECTED FOR USE. PRIOR PROCEDURE, IT WAS NOTED THAT THE PROTECTIVE PAPER TEARS EASILY AND SOME PUMP WAS STICKING TO THE PLASTIC BLISTER. THERE WERE NO VISIBLE DAMAGE TO THE OUTER PACKAGING. HOWEVER, A POTENTIAL RISK OF STERILITY BEING COMPROMISED DUE TO THE TORN PAPER LAYER THAT MAY HAVE COME INTO CONTACT WITH BOTH THE OUTER AND INNER SURFACES OF THE PACKAGING AND THE DEVICE ITSELF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23379251 ↗
  • MalfunctionOct 24, 2025Unsealed Device Packaging; Packaging Problem; Unsealed Device Packaging; Packaging Problem

    IT WAS REPORTED THAT PACKAGING SEAL WAS COMPROMISED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. AN ENCORE 26 PERIPHERAL INTERVENTION INFLATION DEVICE WAS SELECTED FOR USE. PRIOR PROCEDURE, IT WAS NOTED THAT THE PROTECTIVE PAPER TEARS EASILY AND SOME PUMP WAS STICKING TO THE PLASTIC BLISTER. THERE WERE NO VISIBLE DAMAGE TO THE OUTER PACKAGING. HOWEVER, A POTENTIAL RISK OF STERILITY BEING COMPROMISED DUE TO THE TORN PAPER LAYER THAT MAY HAVE COME INTO CONTACT WITH BOTH THE OUTER AND INNER SURFACES OF THE PACKAGING AND THE DEVICE ITSELF. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23379249 ↗
  • MalfunctionOct 15, 2025Contamination; Difficult to Open or Remove Packaging Material; Contamination; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT A STERILITY ISSUE OCCURRED. AN ENCORE 26 INFLATION DEVICE WAS SELECTED FOR USE DURING AN ENDOVASCULAR PROCEDURE. PRIOR TO PROCEDURE, IT WAS NOTED THAT THE PROTECTIVE PAPER COVERING THE INNER TRAY OF THE DEVICE IS DIFFICULT TO TEAR. THERE WAS NO VISIBLE DAMAGING TO THE PACKAGING, HOWEVER THE STERILITY OF THE DEVICE WAS COMPROMISED. THE DEVICE WAS NOT USED DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE DEVICE WITH NO REPORTED PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23293600 ↗
  • MalfunctionSep 16, 2025Defective Device; Output Problem

    FLUID NOTED IN TUBING.

    Patient problems: Insufficient Information

    MAUDE report 23061916 ↗
  • MalfunctionJun 4, 2025Display or Visual Feedback Problem; Physical Resistance/Sticking; Display or Visual Feedback Problem; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE GAUGE WAS READING INACCURATELY, AND THE NEEDLE WAS STICKING. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS, NON-CALCIFIED LATERAL CUTANEOUS VEIN. AN ENCORE PI INFLATION DEVICE WAS USED FOR TREATMENT. DURING THE PROCEDURE, THE BALLOON WAS INFLATED UNDER FLUOROSCOPY. HOWEVER, IT WAS NOTED THAT THE ENCORE METER DID NOT CHANGE FROM 0, THE GAUGE NEEDLE WAS STICKING AND DID NOT MOVE. DESPITE THE INACCURATE GAUGE READING AND NEEDLE STICKING, THE BALLOON ITSELF DID NOT FAIL TO INFLATE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22146950 ↗
  • MalfunctionMay 13, 2025Moisture Damage; Device Contamination with Chemical or Other Material

    UPON OPENING AN ENCORE 26 INFLATOR, IT WAS NOTED TO HAVE FLUID PRESENT IN THE TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22008486 ↗
  • MalfunctionJul 16, 2024Device Contamination with Chemical or Other Material; Material Integrity Problem; Device Contamination with Chemical or Other Material; Material Integrity Problem

    IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND. A BOSTON SCIENTIFIC EMPLOYEE NOTED THAT THERE WAS A HAIR-LIKE OBJECT INSIDE THE UNOPENED INNER POUCH OF AN ENCORE 26 INFLATION DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19756051 ↗
  • MalfunctionDec 28, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT A FOREIGN MATERIAL WAS PRESENT IN THE DEVICE. AN ENCORE 26 INFLATION DEVICE WAS SELECTED FOR USE. DURING PREPARATION, IT WAS NOTED THAT THE DEVICE CONTAINED WHAT APPEARS TO BE A HAIR. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18410433 ↗
  • MalfunctionNov 21, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT DEVICE CONTAMINATION OCCURRED. AN ENCORE 26 WAS SELECTED FOR USE. DURING PREPARATION, IT WAS NOTED THAT A HAIR WAS FOUND ON THE DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER ENCORE 26. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18183102 ↗
  • MalfunctionNov 15, 2023Material Integrity Problem; Material Integrity Problem

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2023. IT WAS REPORTED THAT THERE IS A DROP OF WATER INSIDE THE TUBE. THE TARGET LESION WAS IN THE SHUNT LIMB. AN ENCORE 26 INFLATION DEVICE WAS SELECTED FOR USED. DURING PREPARATION, IT WAS NOTED THAT THERE IS A DROP OF WATER INSIDE THE TUBE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE IS NO PATIENT INJURY. HOWEVER, DEVICE INVESTIGATIONS REVEALED THAT A FOREIGN MATERIAL WAS FOUND INSIDE THE POUCH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18141633 ↗
  • MalfunctionSep 22, 2023Contamination; Tear, Rip or Hole in Device Packaging; Difficult to Open or Remove Packaging Material; Contamination; Tear, Rip or Hole in Device Packaging; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT DEVICE CONTAMINATION OCCURRED. AN ENCORE 26 WAS SELECTED FOR USE. DURING PREPARATION, DIFFICULTY OPENING AN ENCORE INFLATION SYRINGE WAS NOTED. WHILE ATTEMPTING TO OPEN THE PACKAGING, THE STERILE PAPER TORE AND CONTAMINATED THE PRODUCT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17804554 ↗
  • MalfunctionSep 22, 2023Contamination; Tear, Rip or Hole in Device Packaging; Difficult to Open or Remove Packaging Material; Contamination; Tear, Rip or Hole in Device Packaging; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT DEVICE CONTAMINATION OCCURRED. AN ENCORE 26 WAS SELECTED FOR USE. DURING PREPARATION, DIFFICULTY OPENING AN ENCORE INFLATION SYRINGE WAS NOTED. WHILE ATTEMPTING TO OPEN THE PACKAGING, THE STERILE PAPER TORE AND CONTAMINATED THE PRODUCT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17802960 ↗
  • MalfunctionJun 27, 2023Material Separation; Defective Device

    ENCORE 26 INFLATION DEVICE BEING USED DURING CASE APPEARS TO HAVE A DEFECT IN THE PLASTIC. SMALL ROUND AREA WAS NOTICED AFTER USING DEVICE. WAS NOT SURE IF PART OF PLASTIC CAME OFF INSIDE THE SYRINGE/TUBING PART OF THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17208338 ↗
  • MalfunctionMay 4, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT THERE WAS FOREIGN OBJECT IN THE DEVICE. AN ENCORE 26 INFLATION DEVICE WAS NOTED TO HAVE A WHITE FLOATING OBJECT INSIDE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16872963 ↗

Other Syringe, balloon inflation devices

All Syringe, balloon inflation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page