Blazer® II M0045031TK20Recalls & reported problems
Boston Scientific Corporation · Temperature Ablation Catheter · Class 3
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729199939), received between May 1, 2018 and Mar 18, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Device Contaminated During Manufacture or Shipping” (1 report).
3 other recalls cover devices cleared under the same FDA submission (P920047); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Device Contaminated During Manufacture or Shipping1
- Device Operates Differently Than Expected1
- Failure to Deliver Energy1
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 18, 2022Failure to Deliver Energy; Failure to Deliver Energy
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 24FEB2022. IT WAS REPORTED THAT DURING A CARDIAC RADIOFREQUENCY ABLATION PROCEDURE, A BLAZER II CATHETER HAD POOR CONTACT AND COULD NOT DISCHARGE. THERE WERE NO ISSUES REGARDING TEMPERATURE. THE CATHETER WAS DISCHARGED INTERMITTENTLY AND COULD NOT BE USED NORMALLY. NO ERROR MESSAGES. TEMPERATURE CONTROL MODE WAS IN USE. PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT INJURY REPORTED. DEVICE WAS RETURNED TO BOSTON SCIENTIFIC FOR ANALYSIS. HOWEVER, DEVICE ANALYSIS REVEALED THAT THERE IS A CUT IN THE DISTAL PART OF THE CATHETER SHAFT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13805346 ↗MalfunctionNov 27, 2019Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT FEORGIN MATERIAL WAS FOUND INSIDE THE PACKAGING. DURING PREPARATION WHEN THEY OPENED THE BLAZER CATHETER, THEY FOUND THAT THE PACKAGE CONTAINED DIRT/TRASH INSIDE THE BAG. ANOTHER OF THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATION OCCURRED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9385150 ↗MalfunctionMay 1, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 04-APRIL-2018. IT WAS REPORTED THAT TEMPERATURE DOES NOT RISE ANYMORE AT 37DEGREE. A 7/110/2.5/7-4 BLAZER® II ABLATION CATHETER WAS SELECTED FOR TREATMENT OF PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIA (PSVT). DURING PROCEDURE, IT WAS NOTED THAT THE TEMPERATURE WOULD NOT RISE MORE THAN 37 DEGREES. THE GENERATOR WAS CHECKED; HOWEVER, THE TEMPERATURE STILL WOULD NOT RISE. THE PROCEDURE WAS THEN COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S CONDITION WAS GOOD. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED THE THERMISTOR WAS SEPARATED FROM THE TIP AND WAS RETRACTED APPROXIMATELY 4CM BACK.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7475212 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P920047. One submission often covers many models, so check the product and lot details in each record.
- Z-0202-2019 · Jun 20, 2018 · Open, ClassifiedBoston Scientific Maestro 4000 Cardiac Ablation System, Firmware Version 5.14, Model Numbers: M00440000, M0044000D0, M0044000H0,M0044000R0, M0044000Z0,
- Z-0973-2011 · Jul 14, 2009 · TerminatedRF Ablation System Foot Switch, used with the EPT-1000 RF Generator, Foot Switch Model Number UPN M0048400, and in kits with UPNs M0044810 and M0044811. Intended use: The EPT-1000
- Z-1617-2008 · Feb 14, 2008 · TerminatedBoston Scientific SteeroCath-T Ablation Catheter. This catheter is indicated for general indications, treatment of supraventricular tachycardia (svt) and ventricular tachycardia (
Other Cardiac ablation percutaneous catheter devices
- THERMOCOOL SMARTTOUCH SF D134805Biosense Webster Inc · Cardiac ablation percutaneous catheter1 recall176 deaths4,106 reports
- THERMOCOOL SMARTTOUCH SF D134804Biosense Webster Inc · Cardiac ablation percutaneous catheter44 deaths784 reports
- THERMOCOOL SMARTTOUCH SF D134801Biosense Webster Inc · Cardiac ablation percutaneous catheter15 deaths608 reports
- ThermoCool SmartTouch D132705Biosense Webster Inc · Cardiac ablation percutaneous catheter25 deaths574 reports
- THERMOCOOL SMARTTOUCH SF D134702Biosense Webster Inc · Cardiac ablation percutaneous catheter19 deaths366 reports
- THERMOCOOL SMARTTOUCH SF D134701Biosense Webster Inc · Cardiac ablation percutaneous catheter11 deaths278 reports
- ThermoCool SmartTouch D133602Biosense Webster Inc · Cardiac ablation percutaneous catheter8 deaths225 reports
- SMARTABLATE SAT001Biosense Webster Inc · Cardiac ablation percutaneous catheter185 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.