Guider Softip™ XF H965100480Recalls & reported problems
Boston Scientific Corporation · Guide Catheter · Class 2
All models of Guider Softip XF (17)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729202523), received between Dec 10, 2019 and Dec 15, 2023 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 27, 2023Class IIOpen, Classified
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
Z-0014-2024 · Root cause: Employee error · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture2
- Peeled/Delaminated1
Patient problems
- Foreign Body In Patient1
- Muscle Weakness/Atrophy1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- Paralysis1
- Stroke/CVA1
- Syncope/Fainting1
- Thrombosis/Thrombus1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 15, 2023Fracture; Fracture
IT WAS REPORTED THAT PATIENT WITH STROKE WAS TREATED SUCCESSFULLY. HOWEVER, THE NEXT DAY PATIENT HAD SECOND STROKE AND HAD TO BE RETREATED. PHYSICIAN PERFORMED THROMBECTOMY PROCEDURE TO REMOVE THE CLOT. DURING THE PROCEDURE THE OPERATOR FOUND THE RADIOPAQUE MARKER OF THE SUBJECT GUIDE CATHETER IN THE M1 SEGMENT FROM THE FIRST PROCEDURE. ATTEMPTS WERE MADE TO REMOVE THE MARKER BUT WERE UNSUCCESSFUL AND WAS LEFT INSIDE THE PATIENT¿S ANATOMY. THERE WAS SURGICAL DELAY REPORTED DURING THE PROCEDURE.
Patient problems: Foreign Body In Patient; Muscle Weakness/Atrophy; Paralysis; Stroke/CVA; Syncope/Fainting; Thrombosis/Thrombus
MAUDE report 18334427 ↗MalfunctionMay 23, 2022Peeled/Delaminated; Peeled/Delaminated
THE DEVICE WAS RETURNED FOR ANALYSIS AND THE INVESTIGATION OF THE DEVICE REVEALED THAT THE CATHETER (SUBJECT DEVICE) PTFE COATING WAS PEELING. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14468940 ↗InjuryDec 10, 2019Fracture; Fracture
IT WAS REPORTED THAT DURING PROCEDURE, THE GUIDE CATHETER (SUBJECT DEVICE) DISTAL RADIOPAQUE MARKER CAME OFF DURING EMBOLIZATION OF A DURAL FISTULA. THE OPERATOR CHANGED THE INITIAL SHEATH TO A 12F SHEATH AND EXCHANGED THE GUIDEWIRE. THEN THE OPERATOR INFLATED A 8 X 4 BALLOON INSIDE THE TIP RING THAT BROKE OFF AND MANAGED TO RETRIEVE THE TIP FROM THE PATIENT. THIS WAS POSSIBLE AS THE WIRE KEPT THE RADIOPAQUE TIP CONSTRAINED INSIDE THE PATIENT. THERE WAS 1 HOUR SURGICAL DELAY. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9448359 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page