Skip to content

Expo™ H74908526501Recalls & reported problems

Boston Scientific Corporation · Angiographic Catheter · Class 2

All models of Expo (47)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729249399), received between Sep 12, 2022 and Sep 12, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 21, 2024Class IOpen, Classified

    An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.

    Z-1671-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Mar 21, 2024Class IOpen, Classified

    An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.

    Z-1672-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Difficult to Advance1
  • Difficult to Remove1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 12, 2022Difficult to Remove; Difficult to Advance; Material Deformation; Difficult to Remove; Difficult to Advance; Material Deformation

    IT WAS REPORTED VIA USER MEDWATCH MW5111200 THAT REMOVAL DIFFICULTIES OCCURRED. THE PATIENT PRESENTED WITH ST ELEVATION MYOCARDIAL INFARCTION (STEMI). THE TARGET LESION WAS LOCATED IN THE RIGHT FEMORAL ARTERY. THE 5F EXPO WR (5PK) CATHETER WAS INSERTED INTO THE SHEATH. THE PHYSICIAN WAS HAVING DIFFICULTY ENGAGING THE ARTERY AND HAD TO TWIST THE DEVICE TO GET THE CATHETER ENGAGED. WHEN THEY PANNED OUT TO LOOK AT THE ARTERIES THE CATHETER WAS KINKED RIGHT OUTSIDE THE SHEATH AND INSIDE THE ARTERY. UPON WITHDRAWAL, THE CATHETER WOULD NOT COME OUT. THE PATIENT WAS SENT TO SURGERY TO REMOVE THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15401968 ↗

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page