RX ERCP Cannula M00545780Recalls & reported problems

Boston Scientific Corporation · ERCP Cannula · Class 1

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729268673), received between Aug 13, 2018 and Jan 29, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Contamination with Chemical or Other Material2
  • Difficult to Advance2
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 29, 2026Difficult to Advance; Device Contamination with Chemical or Other Material

    NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES THAT WERE USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN RX ERCP CANNULA WAS ATTEMPTED TO BE USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT 22008078 PERFORMED ON (B)(6) 2026. DURING THE PREPARATION, IT WAS REPORTED THAT THERE WAS PLASTIC (NOT PART OF THE CANNULA) INSIDE THE LUMEN OF THE CANNULA WHERE THE LUMEN GETS NARROWER. IT PROHIBITED A 0.035 OR A 0.025 WIRE FROM MAKING IT THROUGH THE LUMEN TO THE TIP OF THE CANNULA. THE PROCEDURE WAS COMPLETED WITH THE ORIGINAL DEVICE UTILIZING A 0.018 GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24218177 ↗
  • MalfunctionJan 29, 2026Difficult to Advance; Device Contamination with Chemical or Other Material

    NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES THAT WERE USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN RX ERCP CANNULA WAS ATTEMPTED TO BE USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT 22008078 PERFORMED ON (B)(6) 2026. DURING THE PREPARATION, IT WAS REPORTED THAT THERE WAS PLASTIC (NOT PART OF THE CANNULA) INSIDE THE LUMEN OF THE CANNULA WHERE THE LUMEN GETS NARROWER. IT PROHIBITED A 0.035 OR A 0.025 WIRE FROM MAKING IT THROUGH THE LUMEN TO THE TIP OF THE CANNULA. THE PROCEDURE WAS COMPLETED WITH THE ORIGINAL DEVICE UTILIZING A 0.018 GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24218170 ↗
  • MalfunctionAug 13, 2018Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN RX ERCP CANNULA WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING PROCEDURE CLOSURE, THE X-RAY IMAGE SHOWED A DETACHED RADIOPAQUE (RO) MARKER FROM THE RX ERCP CANNULA WHICH WAS CONFIRMED AFTER THEY INSPECTED THE RX ERCP CANNULA AFTER USE AND NOTICED DAMAGE TO THE TIP OF THE CANNULA. REPORTEDLY, THERE WAS NO ATTEMPT MADE TO RETRIEVE THE RO MARKER AS IT WAS LOCATED IN THE LIVER TREE WHICH WAS FOUND TO BE VERY DIFFICULT TO RETRIEVE. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7778664 ↗