Renegade™ HI-FLO™ Kit M001182980Recalls & reported problems
Boston Scientific Corporation · Microcatheter Kit · Class 2
All models of Renegade HI-FLO Kit (5)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729280736), received between Oct 5, 2018 and Oct 5, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- No Patient Involvement1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 5, 2018Break; Break
IT WAS REPORTED THAT THE DEVICE SNAPPED. A RENEGADE HI FLOW 105/10 KIT WAS SELECTED FOR USE. DURING PREPARATION, THE CARRIER TUBE WAS HYDRATED BEFORE REMOVING THE DEVICE. WHEN IT WAS PULLED OUT OF THE LOOP, IT WAS FOUND OUT THAT THE CATHETER HAD BEEN SNAPPED AND COULD SEE THE BRAIDING INSIDE. THE PROCEDURE WAS COMPLETED WITH ANOTHER RENEGADE DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. DEVICE EVALUATED BY MANUFACTURER: THE DEVICE WAS RETURNED FOR ANALYSIS. THE HUB, SHAFT AND TIP WERE MICROSCOPICALLY EXAMINED. THE DEVICE WAS FRACTURED IN 1 LOCATION. THE LOCATION OF THE FRACTURE WAS 4CM FROM THE HUB. THE SHAFT WAS NOT COMPLETELY SEPARATED THE INNER LINER WAS STILL CONNECTED. THE STRETCHING AND THE FRACTURES THAT WERE NOTICED ON THE DEVICE ARE CONSISTENT WITH THE DEVICE BEING REMOVED FROM THE SHIPPING TUBE WITHOUT BEING COMPLETELY HYDRATED WITH SALINE. INSPECTION OF THE REMAINDER OF THE DEVICE, APART FROM THE OBSERVED DAMAGE, REVEALED NO OTHER DAMAGE OR IRREGULARITIES. NO OTHER ISSUES WERE IDENTIFIED DURING THE PRODUCT ANALYSIS.
Patient problems: No Patient Involvement
MAUDE report 7938922 ↗
FDA clearance and classification
The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Renegade™ HI-FLO™ Kit M001182980
Has the Renegade™ HI-FLO™ Kit M001182980 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08714729280736), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Renegade™ HI-FLO™ Kit M001182980?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Oct 5, 2018. That report lists device problem “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Renegade™ HI-FLO™ Kit M001182980?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Renegade™ HI-FLO™ Kit M001182980, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Renegade™ HI-FLO™ Kit, model M001182980). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths973 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths55 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page