Skip to content

WALLSTENT™ RP Endoprosthesis M001712380Recalls & reported problems

Boston Scientific Corporation · Self-Expanding Stent · Class 3

All models of WALLSTENT RP Endoprosthesis (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729345923), received between Oct 2, 2017 and Mar 23, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Device Markings/Labelling Problem1

Patient problems

  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 23, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2020. A 10X69X75/ ILIAC 6F UNISTEP PLUS WALLSTENT RP ENDOPROSTHESIS WAS USED WITH NO KNOWN ISSUES REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED STENT DETACHMENT AND SHAFT BREAK.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9865591 ↗
  • MalfunctionOct 2, 2017Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT THE PRODUCT WAS MISLABELED. THE TARGET LESION WAS LOCATED IN THE ILIAC ARTERY. A 10X69X75 WALLSTENT¿ RP ENDOPROSTHESIS WAS SELECTED TO TREAT THE LESION. DURING ADVANCING INTO THE INTRODUCER SHEATH, IT WAS NOTED THAT THE STENT WAS FLAWED AS IT WAS MUCH LONGER THAN WHAT WAS LABELED. THE ACTUAL STENT LENGTH MEASURED AT 130MM. THE STENT WAS DEPLOYED AS THERE WAS NO SPARE STENT FOR REPLACEMENT. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6907507 ↗

Other STENT, ILIAC devices

All STENT, ILIAC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page