Mach1™ H74934356220Recalls & reported problems

Boston Scientific Corporation · Guide Catheter · Class 2

All models of Mach1 (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729351177), received between Jul 31, 2019 and May 30, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1
  • Positioning Problem1
  • Unsealed Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 30, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE GUIDE CATHETER BROKE. A MACH 1 GUIDE CATHETER WAS SELECTED AS PART OF TREATMENT OF THE TARGET LESION. AFTER THE GUIDE CATHETER WAS ADVANCED, IT WAS NOTICED THAT THE DISTAL END HAD COME OFF. THE DISCONNECTED PART OF THE CATHETER WAS STILL ON THE GUIDEWIRE AND WAS ABLE TO BE REMOVED WITH THE GUIDEWIRE. THE DEVICE WAS REMOVED FULLY BY PULLING IT NORMALLY OUT. ANOTHER OF THE SAME DEVICE WAS THEN USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. NO COMPLICATIONS WERE REPORTED, AND PATIENT CONDITION IS NORMAL POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22114797 ↗
  • MalfunctionJun 10, 2024Break; Positioning Problem; Break; Positioning Problem

    IT WAS REPORTED THAT THE DEVICE GOT SEPARATED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY (RCA). A 6F MACH 1 CATHETER-GUIDE WAS SELECTED FOR USE. DURING THE PROCEDURE, A RESISTANCE WAS FELT, AND THE CATHETER COULD NOT BE ADVANCED TO THE DESIRED POSITION. EVENTUALLY THE CATHETER GOT SEPARATED IN THE OSTIUM OF DISTAL RCA. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19498191 ↗
  • MalfunctionJan 8, 2024Unsealed Device Packaging; Unsealed Device Packaging

    IT WAS REPORTED THAT THE INNER POUCH WAS UNSEALED. A 6F MACH 1 GUIDE CATHETER WAS SELECTED FOR USE. DURING UNPACKING, THEY NOTICED THE INNER POUCH WAS UNSEALED AND THE DEVICE STERILITY WAS COMPROMISED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18470974 ↗
  • InjuryJul 31, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE TIP OF THE CATHETER DETACHED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED IN RIGHT CORONARY ARTERY (RCA). DURING THE PERCUTANEOUS CORONARY INTERVENTION (PCI), A MODEL 6F MACH 1 AL.75 WAS USED AND A NON BSC STENT WAS PLACED IN RCA PROXIMAL. WHEN THE DELIVERY BALLOON WAS PULLED AND INFLATED AT 16ATM TO DILATE THE STENT, THE BALLOON GOT PARTIALLY INFLATED INSIDE THE TIP OF THIS DEVICE. WHEN THE MODEL 6F MACH 1 AL.75 WAS PULLED TO REMOVE IT, THE TIP GOT SEPARATED FROM THE MAIN BODY. THE SEPARATED TIP WAS UNABLE TO RETRIEVE BY USING A SNARE. THE PHYSICIAN INSERTED AN UNKNOWN GUIDEWIRE AND ANOTHER UNKNOWN BALLOON CATHETER, THEN IT WAS INFLATED AND RETRIEVED THE SEPARATED TIP TOGETHER WITH THE SYSTEM. WHEN RETRIEVING REACHED TO THE BRACHIAL ARTERY, THE BALLOON RUPTURED AND THE SEPARATED TIP REMAINED INSIDE THE BODY. THE PHYSICIAN WAS HAVING DIFFICULTY IN RETRIEVING PERCUTANEOUSLY, BUT UPON FLUOROSCOPY, IT WAS FIXED IN THE STATE THAT IT ENTERED IN THE SIDE BRANCH. IT WAS REMAINED AS IT WAS AND THE PATIENT UNDERGONE FOLLOW UP OBSERVATION. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENTS STATUS IS STABLE.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8848505 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →