Skip to content

Mach1™ H749343566860Recalls & reported problems

Boston Scientific Corporation · Guide Catheter · Class 2

All models of Mach1 (28)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729351986), received between Mar 28, 2019 and Dec 8, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Contamination1
  • Nonstandard Device1
  • Physical Resistance/Sticking1
  • Positioning Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Vascular Dissection1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 8, 2025Nonstandard Device; Nonstandard Device

    REPORTABLE BASED ON THE DEVICE ANALYSIS COMPLETED ON 17NOV2025. IT WAS REPORTED THAT THE DEVICE WAS DEFECTIVE. A 6F MACH 1 WAS SELECTED FOR USE. DURING INSPECTION OF THE PULLED-OUT ITEMS, IT WAS NOTED THAT THE UNIT WAS MARKED AS DEFECTIVE. HOWEVER, THE DEVICE ANALYSIS REVEALED THAT THE SHAFT WAS DETACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23743402 ↗
  • MalfunctionJun 10, 2024Break; Positioning Problem; Break; Positioning Problem

    IT WAS REPORTED THAT THE DEVICE GOT SEPARATED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY (RCA). A 6F MACH 1 CATHETER-GUIDE WAS SELECTED FOR USE. DURING THE PROCEDURE, A RESISTANCE WAS FELT, AND THE CATHETER COULD NOT BE ADVANCED TO THE DESIRED POSITION. EVENTUALLY THE CATHETER GOT SEPARATED IN THE OSTIUM OF DISTAL RCA. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19498189 ↗
  • MalfunctionApr 30, 2021Contamination; Contamination

    IT WAS REPORTED THAT THE STERILITY OF THE DEVICE WAS COMPROMISED. A MACH1 CATHETER-GUIDE WAS DELIVERED AND UPON REMOVING FROM PACKAGING, THE STERILE PACKAGING OF THE PRODUCT WAS OPENED. NO PATIENT INVOLVEMENT WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11751016 ↗
  • InjuryMar 28, 2019Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT VESSEL DISSECTION OCCURED. THE LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY. A ROTALINK ADVANCER, 1.25 MM ROTALINK BURR, 330 CM ROTAWIRE AND 6F MACH1 GUIDING CATHETER WERE SELECTED FOR USE. THE ROTAWIRE WAS GUIDED THROUGH A NON-BSC MICROCATHETER TO THE DISTAL END OF THE LESION IN THE PL BRANCH. IT WAS NOTED THAT IT WAS DIFFICULT TO GUIDE THE ROTAWIRE THROUGH THE LESION DUE TO THE OCCLUSION, SO THE WIRE WAS SPUN HEAVILY TO GET IT TO THE DISTAL POINT IN THE PL BRANCH. SUBSEQUENTLY, THE BURR WAS ADVANCED IN DYNAGLIDE MODE OVER THE ROTAWIRE WITHIN THE MACH1 GUIDE CATHETER BUT THERE WAS RESISTANCE APPROXIMATELY 20CM INTO THE GUIDING CATHETER. THE ROTAWIRE STARTED TO SPIN QUITE HEAVILY AND THE DISTAL TIP OF THE WIRE BROKE OFF WHILE THE BURR WAS STILL IN THE GUIDING AND RESULTED TO A LIMITED LOCAL DISSECTION. AS THE BURR WAS COMPLETELY STUCK WHILE BEING IN THE GUIDING CATHETER, THE BURR AND WIRE WERE RETIREVED BUT THE WIRE POSITION WAS LOST. AN ATTEMPT TO REWIRE THE LESION WAS DONE USING A NON-BSC WIRE WHICH WAS GUIDED THROUGH THE MICROCATHETER BUT THE WIRE BECAME STUCK IN THE DISSECTION. THE PROCEDURE WAS DISCONTINUED AND THE DISSECTION WAS GIVEN TIME TO HEAL; SAM

    Patient problems: Vascular Dissection

    MAUDE report 8460916 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page