Skip to content

Maverick²™ H7493892815250Recalls & reported problems

Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2

All models of Maverick² (30)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Boston Scientific Maverick balloon catheter has 110 adverse event reports, with no deaths and 2 injuries. Most reports describe the balloon bursting or the shaft breaking during artery-opening procedures, often in heavily calcified blockages. Many reports say the patient was fine afterward. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
111
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 111 adverse event reports that list this exact device (DI 08714729370086), received between Nov 16, 2016 and Sep 24, 2026 — classified as 2 injuries, 109 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (61 reports).

5 other recalls cover devices cleared under the same FDA submission (P860019); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 3 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 3 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 3 reportsNov 2025: 2 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture61
  • Break31
  • Failure to Advance12
  • Nonstandard Device8
  • Adverse Event Without Identified Device or Use Problem4
  • Material Deformation3
  • Inflation Problem2
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions56
  • No Consequences Or Impact To Patient45
  • No Patient Involvement7
  • No Known Impact Or Consequence To Patient4
  • Vascular Dissection1

Recent adverse event reports

The 25 most recent of 111 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Material Rupture

    It was reported that balloon rupture occurred.The 60% target lesion was located in the moderate to severely tortuous and mildly calcified left anterior descending artery. The first 2.00mm x 20mm Maverick 2 balloon catheter was advanced for dilation. However, at 8 - 12 atmospheres, the balloon ruptured. Then, second 2.50mm x 15mm Maverick 2 balloon catheter was used to pre-dilate the left anterior descending artery lesion at 8 - 12 atmospheres, but balloon rupture was observed. The non-BSC guidewire from the third diagonal branch was exchanged to the second diagonal branch, and the 2.5mm x 15mm Maverick balloon was used to pre-dilate the second diagonal branch lesion at 8 - 12 atmospheres. An attempt to advance the 2.5mm x 10mm Wolverine cutting balloon to the lesion in the anterior descending artery was unsuccessful. The diagonal branch guidewire was withdrawn, the microcatheter was advanced past the stenotic lesion in the anterior descending artery, and a rotational atherectomy guidewire was exchanged to the distal segment of the anterior descending artery. The 1.5 BSC rotational atherectomy tip was advanced to the calcified lesions in the proximal and mid segments and rotated at

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884197 ↗
  • MalfunctionAug 11, 2026Material Rupture

    It was reported that balloon rupture occurred.A 90% stenosed target lesion was located in the mildly tortuous and severely calcified proximal to mid right coronary artery. A 2.50mm x 15mm Maverick? balloon catheter was used for pre-dilatation at 8 to 12 atmospheres. However, during the pre-dilatation, the proximal bend first and the balloon ruptured. The procedure was completed and another device was used to complete the procedure. Intraoperative ECG and hemodynamic parameters remained stable, and the patient experienced no specific discomfort. He was returned to the ward postoperatively.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50214650 ↗
  • MalfunctionMar 30, 2026Break; Nonstandard Device

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE PATIENT WAS UNDERGOING PERCUTANEOUS CORONARY INTERVENTION. THE 75% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED RIGHT CORONARY ARTERY. A 2.50MM X 15MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE PROCESS OF DELIVERING THE GUIDE WIRE INTO THE PATIENT'S BODY, THE DELIVERY SHAFT OF THE DEVICE FRACTURED. THE DEVICE WAS PROMPTLY REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THE PATIENT CONDITION FOLLOWING THE PROCEDURE WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24729069 ↗
  • MalfunctionFeb 17, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 85% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK 2 BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, THE BALLOON BURST DUE TO SEVERE CALCIFICATION. THE DEVICE WAS REMOVED AND COMPLETED THE PROCEDURE WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS. PATIENT WAS IN GOOD CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24366758 ↗
  • MalfunctionJan 19, 2026Nonstandard Device; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24109010 ↗
  • MalfunctionDec 8, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT UNDERWENT PERCUTANEOUS CORONARY INTERVENTION. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK 2 BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, WHILE TRYING TO INSERT THE DEVICE INTO THE LESION, THE BALLOON BURST DUE TO SEVERE CALCIFICATION. THE DEVICE WAS REMOVED SUCCESSFULLY, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23739425 ↗
  • MalfunctionNov 26, 2025Nonstandard Device; Material Rupture; Nonstandard Device; Material Rupture

    IT WAS REPORTED THAT THE BALLOON BURST. THE 85% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK 2 BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, WHILE TRYING TO INSERT THE DEVICE INTO THE LESION, THE BALLOON BURST DUE TO SEVERE CALCIFICATION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS, AND THE PATIENT CONDITION WAS GOOD POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23657446 ↗
  • MalfunctionNov 7, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 85% STENOSED TARGET LESION WAS LOCATED IN A SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK? BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE PROCEDURE, IT WAS NOTICED THAT THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE WITHOUT ANY PATIENT COMPLICATIONS REPORTED. THE PATIENT WAS IN GOOD CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23498208 ↗
  • MalfunctionOct 31, 2025Nonstandard Device; Material Rupture; Nonstandard Device; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK 2 BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, THE BALLOON BURST DUE TO SEVERAL CALCIFICATION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS. PATIENT WAS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23442542 ↗
  • MalfunctionOct 16, 2025Nonstandard Device; Material Rupture; Nonstandard Device; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT UNDERWENT PERCUTANEOUS CORONARY INTERVENTION. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK 2 BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, WHILE TRYING TO INSERT THE DEVICE INTO THE LESION, THE BALLOON BURST DUE TO SEVERE CALCIFICATION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23310109 ↗
  • MalfunctionOct 7, 2025Nonstandard Device; Material Rupture; Nonstandard Device; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK? BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT CONDITION WAS GOOD POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23240375 ↗
  • MalfunctionSep 23, 2025Nonstandard Device; Material Rupture; Nonstandard Device; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 85% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK? BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS IN GOOD CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23121839 ↗
  • MalfunctionSep 8, 2025Nonstandard Device; Material Rupture; Nonstandard Device; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT UNDERWENT PERCUTANEOUS CORONARY INTERVENTION. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK? BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS GOOD POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22996473 ↗
  • MalfunctionJul 21, 2025Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. FOLLOWING PRE-DILATATION, A 2.50MM X 15MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE PROCEDURE, THE BALLOON BURST DUE TO SEVERE CALCIFICATION. THE DEVICE WAS REMOVED SUCCESSFULLY FROM THE PATIENT'S BODY, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22555180 ↗
  • MalfunctionJul 18, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 85% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE INITIAL INFLATION AT NOMINAL PRESSURE, THE BALLOON RUPTURED DUE TO SEVERE CALCIFICATION. THE DEVICE WAS SIMPLY PULLED OUT FROM THE PATIENT'S BODY, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED, AND PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22543468 ↗
  • MalfunctionJul 1, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A SHAFT BREAK OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY (LAD). A 2.50 MM X 15 MM MAVERICK 2 CATHETER WAS SELECTED FOR DILATATION. WHILE ADVANCING THE BALLOON INTO THE GUIDING CATHETER, THE SHAFT OF THE DEVICE BROKE AT THE MIDDLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT REMAINED GOOD WITH NO COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22377520 ↗
  • MalfunctionJun 12, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK 2 BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, THE BALLOON BURST DUE TO SEVERAL CALCIFICATION. THE DEVICE WAS REMOVED AND COMPLETED THE PROCEDURE WITH A DIFFERENT DEVICE. THERE WAS NO PATIENT COMPLICATIONS REPORTED. PATIENT WAS IN GOOD CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22198076 ↗
  • MalfunctionDec 26, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 85% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT CONDITION WAS GOOD POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21021149 ↗
  • MalfunctionNov 14, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY (LAD). A 2.50MM X 15MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AT NOMINAL ATMOSPHERES. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE AND NO PATIENT COMPLICATIONS WERE REPORTED. THE PATIENT REMAINED IN STABLE CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20684316 ↗
  • MalfunctionNov 12, 2024Material Rupture; Failure to Advance; Material Rupture; Failure to Advance

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION AT NOMINAL ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND PATIENT IS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20667873 ↗
  • MalfunctionNov 5, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT CONDITION WAS GOOD POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20615938 ↗
  • MalfunctionOct 9, 2024Break; Break

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK 2 BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, THE SHAFT WAS FRACTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED. PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20409357 ↗
  • MalfunctionSep 24, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION, WHICH WAS 85% STENOSED, WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY (LAD). A 2.50MM X 15MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE PROCEDURE, THE BALLOON RUPTURED AT ITS RATED BURST PRESSURE. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20290644 ↗
  • MalfunctionMay 21, 2024Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT THE DEVICE WAS DIFFICULT TO REMOVE. A 2.50MM X 15MM MAVERICK 2 BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, IT WAS FOUND THAT THE PRE-DILATION BALLOON CATHETER WAS DIFFICULT TO WITHDRAW AFTER THE NON-BOSTON SCIENTIFIC GUIDEWIRE PENETRATED IT. THE PHYSICIAN IMMEDIATELY REPLACED IT WITH A NEW PTCA BALLOON DILATION CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19367178 ↗
  • MalfunctionMay 9, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 2.50MM X 15MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE PROCEDURE, THE BALLOON RUPTURED AT RATED BURST PRESSURE DUE TO HEAVY TIGHT CALCIFICATION. THE DEVICE WAS REMOVED AND COMPLETED THE PROCEDURE WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED. PATIENT WAS IN GOOD CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19278010 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P860019. One submission often covers many models, so check the product and lot details in each record.

  • Z-2165-2008 · May 1, 2007 · Terminated
    Boston Scientific Scimed Maverick 2 Monorail" PTCA Catheter, 3.0 mm 15 mm REF 38928-1530 Lot 9565194 Use before 2010-04 Indicated for balloon dilation of the stenotic portion o
  • Z-2166-2008 · Nov 1, 2005 · Terminated
    Boston Scientific Scimed Maverick 2 Monorail 12mm Balloon Catheter, 2.5 mm x 12 mm Made in Ireland Ballybrit Business Park, Galway, Ireland REF 38928-1225 Use before 2007-01 UPN H
  • Z-0032-04 · Aug 20, 2003 · Terminated
    Ace Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 1 cm Tip and 20 mm Balloon, Catalog Number H74903628151
  • Z-0031-04 · Aug 20, 2003 · Terminated
    Ace Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 2cm Tip 20 mm Balloon, Catalog Number H74902071051
  • Z-0030-04 · Aug 20, 2003 · Terminated
    Ace Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 2cm Tip 20 mm Balloon, Catalog Number H74902071011

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →