Maverick²™ H7493892820150Recalls & reported problems
Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2
All models of Maverick² (30)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Maverick balloon catheter is a percutaneous coronary catheter. Reports mostly describe the balloon rupturing or failing to cross narrowed, hardened arteries, with procedures usually completed using another device and no patient complications. There are 41 reports naming this identifier, including 1 death and no injuries. There are no recalls naming this device. Reports rose to 2 in the latest 12 months from 0 in the previous 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 40
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 08714729370154), received between Sep 6, 2017 and Feb 27, 2026 — classified as 40 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (34 reports).
5 other recalls cover devices cleared under the same FDA submission (P860019); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture34
- Failure to Advance10
- Adverse Event Without Identified Device or Use Problem2
- Material Deformation2
- Break1
- Inflation Problem1
- Nonstandard Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions25
- No Consequences Or Impact To Patient15
Recent adverse event reports
The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 27, 2026Nonstandard Device; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND MODERATELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 1.50 MM X 20 MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, UPON INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THERE WERE NO COMPLICATIONS REPORTED POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24475238 ↗MalfunctionAug 21, 2024Failure to Advance; Failure to Advance
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 31JUL2024. IT WAS REPORTED THAT CROSSING DIFFICULTIES WERE ENCOUNTERED. THE 75% STENOSED TARGET LESION WAS LOCATED IN A SEVERELY TORTUOUS AND SEVERELY CALCIFIED LESION IN THE LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK WAS ADVANCED FOR TREATMENT BUT COULD NOT PASS THROUGH THE LESION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE WITHOUT ANY PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS WAS FINE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A BALLOON LONGITUDINAL TEAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20019000 ↗MalfunctionAug 6, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS, AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THIRD INFLATION AT 6 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT CONDITION WAS GOOD POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19917328 ↗MalfunctionAug 6, 2024Material Rupture; Failure to Advance; Material Rupture; Failure to Advance
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED MID RIGHT CORONARY ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE PROCEDURE, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE, AND NO PATIENT COMPLICATIONS WERE REPORTED. THE PATIENT WAS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19917326 ↗MalfunctionJul 2, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 85% STENOSED TARGET LESION WAS LOCATED IN A SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT CIRCUMFLEX ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE PROCEDURE, THE BALLOON BURST. THE DEVICE WAS REMOVED FROM THE PATIENT'S BODY AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE WITHOUT ANY PATIENT COMPLICATIONS. THE PATIENT WAS IN GOOD CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19656195 ↗MalfunctionApr 25, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED MID-LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE SECOND INFLATION AT 12 ATMOSPHERES FOR 21-22 SECONDS, THE BALLOON BURST. THE DEVICE WAS COMPLETELY REMOVED AND THERE WERE NO FRAGMENTS LEFT INSIDE THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19182870 ↗MalfunctionDec 27, 2023Material Rupture; Failure to Advance; Material Rupture; Failure to Advance
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, THE DEVICE FAILED TO CROSS THE LESION, AND WHEN THE BALLOON WAS INFLATED AT NOMINAL PRESSURE, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED, AND THE PATIENT CONDITION WAS GOOD POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18401295 ↗MalfunctionNov 27, 2023Material Rupture; Failure to Advance; Material Rupture; Failure to Advance
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY (LAD). A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, THE DEVICE WAS UNABLE TO CROSS THE LESION AND DURING INFLATION AT NOMINAL PRESSURE, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18212432 ↗MalfunctionMar 21, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. INFLATION WAS PERFORMED THREE TIMES. HOWEVER, DURING THE THIRD INFLATION AT 6 ATMOSPHERES, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED, AND THE PATIENT CONDITION WAS STABLE POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16587292 ↗MalfunctionMar 8, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING PROCEDURE, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16506390 ↗MalfunctionFeb 6, 2023Material Rupture; Failure to Advance; Material Rupture; Failure to Advance
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED MID LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE PROCEDURE, THE BALLOON FAILED TO CROSS THE LESION AND BURST. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16303388 ↗MalfunctionFeb 6, 2023Material Rupture; Failure to Advance; Material Rupture; Failure to Advance
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED MID LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE PROCEDURE, THE BALLOON CATHETER FAILED TO CROSS THE LESION AND THE BALLOON BURST. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16302919 ↗MalfunctionAug 24, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED MID LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, ON THE SECOND INFLATION AT 12 ATMOSPHERES FOR 21 TO 22 SECONDS, THE BALLOON RUPTURED. THERE WERE NO BALLOON PIECES LEFT INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15287618 ↗MalfunctionMay 23, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR TREATMENT. HOWEVER, ON THE THIRD INFLATION AT 6 ATMOSPHERES, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14476921 ↗MalfunctionFeb 25, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THIRD INFLATION AT 6 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13602995 ↗MalfunctionJan 7, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE THIRD INFLATION AT 6 ATMOSPHERES, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13195528 ↗MalfunctionNov 29, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 85% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED MID LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE PRESSURE DID NOT RISE TO MORE THAN 10 ATMOSPHERES. UPON CHECKING, A SMALL HOLE IN THE MARKER WAS FOUND. THE DEVICE WAS COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT CONDITION WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12884729 ↗MalfunctionOct 6, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, ON THE THIRD INFLATION AT 6 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12587180 ↗MalfunctionOct 1, 2021Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED AT THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED RIGHT CORONARY ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION AT NOMINAL PRESSURE, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WAS NO PATIENT INJURY REPORTED AND THE PATIENT'S CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12565598 ↗MalfunctionJul 27, 2021Material Rupture; Failure to Advance; Material Rupture; Failure to Advance
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 92% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED MID LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED BUT FAILED TO CROSS THE LESION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12230292 ↗MalfunctionMay 26, 2021Material Rupture; Material Deformation; Material Rupture; Material Deformation
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 85% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED MID LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK 2 BALLOON CATHETER WAS ADVANCED FOR DILATION BUT THE BALLOON WAS KINKED DUE TO ANGULATION WHEN CROSSING THE LESION. FURTHERMORE, DURING SECOND INFLATION AT 12 ATMOSPHERES FOR 21-22 SECONDS, THE BALLOON RUPTURED DUE TO RESISTANCE AND CALCIFICATION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11890142 ↗MalfunctionMay 5, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED DISTAL RIGHT CORONARY ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON DID NOT EXPAND AND WAS NOTED TO BE RUPTURED. ANOTHER 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED, BUT ALSO RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11781694 ↗MalfunctionMay 5, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED DISTAL RIGHT CORONARY ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON DID NOT EXPAND AND WAS NOTED TO BE RUPTURED. ANOTHER 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED, BUT ALSO RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11781670 ↗MalfunctionApr 8, 2021Material Rupture; Material Deformation; Material Rupture; Material Deformation
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 85% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED MID LEFT ANTERIOR DESCENDING ARTERY. A 1.50MM X 20MM MAVERICK BALLOON CATHETER WAS ADVANCED FOR DILATATION WHERE A KINK WAS OBSERVED IN THE DEVICE. DURING SECOND INFLATION AT 12 ATMOSPHERES FOR 21-22 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11638401 ↗MalfunctionFeb 1, 2021Inflation Problem; Inflation Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 14-JAN-2021. IT WAS REPORTED THAT BALLOON INFLATION FAILURE OCCURRED. DURING PREPARATION, A 1.50 MM X 20 MM MAVERICK BALLOON CATHETER WAS SELECTED FOR DILATATION. HOWEVER, THE BALLOON FAILED TO INFLATE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS WAS STABLE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED HYPOTUBE SEPARATION. IT WAS FURTHER REPORTED THAT THE BREAK OCCURRED OUTSIDE THE PATIENT'S BODY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11261150 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P860019. One submission often covers many models, so check the product and lot details in each record.
- Z-2165-2008 · May 1, 2007 · TerminatedBoston Scientific Scimed Maverick 2 Monorail" PTCA Catheter, 3.0 mm 15 mm REF 38928-1530 Lot 9565194 Use before 2010-04 Indicated for balloon dilation of the stenotic portion o
- Z-2166-2008 · Nov 1, 2005 · TerminatedBoston Scientific Scimed Maverick 2 Monorail 12mm Balloon Catheter, 2.5 mm x 12 mm Made in Ireland Ballybrit Business Park, Galway, Ireland REF 38928-1225 Use before 2007-01 UPN H
- Z-0030-04 · Aug 20, 2003 · TerminatedAce Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 2cm Tip 20 mm Balloon, Catalog Number H74902071011
- Z-0031-04 · Aug 20, 2003 · TerminatedAce Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 2cm Tip 20 mm Balloon, Catalog Number H74902071051
- Z-0032-04 · Aug 20, 2003 · TerminatedAce Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 1 cm Tip and 20 mm Balloon, Catalog Number H74903628151
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page