Express® LD Iliac / Biliary H74938046106070Recalls & reported problems
Boston Scientific Corporation · Premounted Stent System · Class 3
All models of Express LD Iliac / Biliary (34)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729392118), received between Jul 31, 2023 and Oct 7, 2024 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Difficult or Delayed Positioning1
- Difficult to Advance1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryOct 7, 2024Difficult or Delayed Positioning; Difficult to Advance; Device Dislodged or Dislocated; Difficult or Delayed Positioning; Difficult to Advance; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT MOVED ON BALLOON OCCURRED, REQUIRING PLACEMENT OF ANOTHER STENT. THE OVER THAN 80% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED VESSEL. THE PATIENT PRESENTED WITH BILATERAL COMMON ILIAC ARTERY (CIA) DISEASE, UNDERWENT A PROCEDURE USING BILATERAL FEMORAL ACCESS. TWO 10.0X60X75CM EXPRESS LD STENTS WERE PLACED IN THE OSTIUM OF THE COMMON ILIAC ARTERY. DURING THE INSERTION ON THE RIGHT SIDE, THE PHYSICIAN ENCOUNTERED RESISTANCE. UNDER ANGIOGRAPHY, THE STENT WAS STILL ON THE CATHETER, SO THE STENT WAS INFLATED SIMULTANEOUSLY. HOWEVER, IT WAS NOTED THAT ONLY THE TOP HALF OF THE STENT WAS EXPANDED, WHILE THE BOTTOM PART REMAINED UNEXPANDED. THE BALLOON WAS REMOVED, AND A NEW BALLOON WAS ADVANCED TO INFLATE THE UNEXPANDED PART BUT WAS UNSUCCESSFUL. THE PHYSICIAN LOST WIRE ACCESS AND HAD TO COME CONTRALATERAL SIDE WITH A NEW WIRE BUT WAS UNABLE TO GET TRUE LUMEN THROUGH THE STENT. SUBSEQUENTLY, THE PHYSICIAN PLACED ANOTHER WIRE BESIDE THE STENT, CRUSHED IT AGAINST THE ARTERY WALL, AND PLACED ANOTHER EXPRESS LD OVER IT TO LOCK IT IN PLACE. AS PER PHYSICIAN'S OPINION, THE STENT MAY HAVE PARTIALLY SLID OFF THE BALLOON DURING THE PR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20385421 ↗MalfunctionJan 23, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 10.0X60X75CM EXPRESS LD ILIAC / BILIARY STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING DEPLOYMENT, THE BALLOON SEEMED TO HAVE BURST. UPON REMOVAL, IT APPEARED AS THOUGH THERE WAS A SMALL HOLE IN THE PART OF THE BALLOON. THE DEVICE WAS REMOVED FROM STERILE FIELD, WIPED TWICE, AND BAGGED TWICE. IT WAS NOTED THAT THE PHYSICIAN WAS NOT CONCERNED ABOUT ANY PIECES OF THE BALLOON REMAINING IN THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18565235 ↗InjuryJul 31, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA CONTRALATERAL APPROACH. THE TARGET LESION WAS LOCATED IN ILIAC ARTERY. A 10.0X60X75CM EXPRESS LD ILIAC STENT WAS ADVANCED FOR TREATMENT. HOWEVER, DURING THE PROCEDURE, THE STENT CAME OFF THE BALLOON AT THE BIFURCATION. THE PHYSICIAN OPTED TO TRANSFER THE PATIENT TO A DIFFERENT HEALTHCARE FACILITY. IT WAS DETERMINED THE OTHER HEALTHCARE FACILITY WOULD HAVE BETTER EQUIPMENT AND BACKUP TO RETRIEVE THE STENT. BEFORE TRANSFERRING THE PATIENT TO THE NEW HEALTHCARE FACILITY, THE STENT WAS MOVED FROM OUT OF THE BIFURCATION TO THE AORTA. IT WAS THEN SNARED IN ORDER TO SECURE THE STENT DURING TRANSPORT TO THE NEW HEALTHCARE FACILITY. AFTER ARRIVAL AT THE NEW HEALTHCARE FACILITY, A LARGE DRY SHEATH AND SNARE WERE USED TO PULL THE STENT OUT. FINALLY. BOTH ILIAC ARTERIES WERE THEN STENTED. THE PROCEDURE WAS COMPLETED, AND NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17425623 ↗
Other STENT, ILIAC devices
- Cordis S.M.A.R.T. Control Vascular Stent System C10040MLCordis US Corp. · STENT, ILIAC1 recall3 reports
- Absolute Pro 1012534-100Abbott Vascular Inc. · STENT, ILIAC51 reports
- Protégé™ GPS™ SERB65-09-60-80Ev3, Inc · STENT, ILIAC1 recall0 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C09060MLCordis US Corp. · STENT, ILIAC1 recall0 reports
- Absolute Pro 1012535-100Abbott Vascular Inc. · STENT, ILIAC1 death41 reports
- Omnilink Elite 1012631-39Abbott Vascular Inc. · STENT, ILIAC27 reports
- Omnilink Elite 1012632-29Abbott Vascular Inc. · STENT, ILIAC25 reports
- Absolute Pro 1012536-100Abbott Vascular Inc. · STENT, ILIAC21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page