Skip to content

Express® LD Iliac / Biliary H74938046660750Recalls & reported problems

Boston Scientific Corporation · Premounted Stent System · Class 3

All models of Express LD Iliac / Biliary (34)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729392170), received between Jul 5, 2017 and Jul 5, 2017 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 5, 2017Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT STENT DISLODGMENT OCCURRED. THE TARGET LESION WAS LOCATED IN THE PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA). A 6.0X60X75CM EXPRESS LD ILIAC / BILIARY STENT WAS ADVANCED TO TREAT THE LESION. THE PHYSICIAN MOVED THE STENT PAST THE LESION AND DECIDED TO PULL IT BACK. HOWEVER, WHEN THE PHYSICIAN TRIED TO PULL THE STENT BACK TO A VERY TIGHT STRICTURE, THE STENT SPLIT OFF THE STENT DELIVERY BALLOON. SUBSEQUENTLY, THE DISLODGED STENT WAS RE-WIRED AND A BALLOON CATHETER WAS USED TO DEPLOY THE STENT, AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6687713 ↗

Other STENT, ILIAC devices

All STENT, ILIAC devices →

Sources and updates