Express® LD Iliac / Biliary H74938046860750Recalls & reported problems

Boston Scientific Corporation · Premounted Stent System · Class 3

All models of Express LD Iliac / Biliary (34)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729392293), received between Feb 21, 2019 and Dec 21, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation2
  • Device Dislodged or Dislocated1
  • Difficult to Advance1
  • Difficult to Insert1
  • Unsealed Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Patient Involvement1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 21, 2025Difficult to Insert; Difficult to Insert

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 25NOV2025. IT WAS REPORTED THAT STENT WAS DIFFICULT TO INSERT. AN 8.0X60X75CM EXPRESS LD STENT WAS SELECTED FOR USE IN AN ILIAC STENTING PROCEDURE. DURING THE PROCEDURE, STENT WAS PREPARED ACCORDING TO STANDARD PROCEDURE BUT COULD NOT BE INTRODUCED INTO THE NON-BOSTON SCIENTIFIC SHEATH. THERE WAS NO VISIBLE DAMAGE TO THE DEVICE UPON OPENING THE PACKAGE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATION AS THE RESULT OF THIS EVENT. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE FIRST TWO ROWS OF THE PROXIMAL STENT STRUTS WERE FOUND TO BE LIFTED/DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23861096 ↗
  • MalfunctionNov 13, 2024Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation

    IT WAS REPORTED THAT STENT DAMAGE OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA RETROGRADE APPROACH. THE 50% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILD TO MODERATELY CALCIFIED EXTERNAL ILIAC ARTERY. A 7F SHEATH AND 0.35 NON-BOSTON SCIENTIFIC (BSC) GUIDEWIRE WAS ADVANCED, AND THE DISEASED LESION WAS DETERMINED USING A NON-BSC IVUS CATHETER THROUGH COMMON AND EXTERNAL ILIAC ARTERY. AN 8.0X60X75CM EXPRESS LD ILIAC/BILIARY STENT BALLOON EXPANDABLE CATHETER WAS ADVANCED FOR TREATMENT. HOWEVER, RESISTANCE WAS FELT WHEN THE STENT WAS ADVANCED INTO COMMON ILIAC WITH POSSIBLE PLAQUE. THE STENT WAS REMOVED, AND THE LESION WAS DILATED WITH A 6MM BALLOON CATHETER. THE STENT WAS REINSERTED AGAIN, BUT STILL WOULD NOT ADVANCE. THE PHYSICIAN PLACED A 7MMX80 NON-BSC STENT IN THE EXTERNAL ILIAC, AND POST-DILATED WITH 6MM BALLOON. THE PHYSICIAN REINSERTED THIS STENT AGAIN, BUT STILL WOULD NOT ADVANCE. THIS STENT WAS REMOVED AND SUSPECTED THAT THE PROXIMAL END WAS DAMAGE BUT COULD NOT CONFIRM. A NEW 7X45 SHEATH WAS PLACED IN THE COMMON ILIAC AND TRIED AGAIN TO REINSERT THIS EXPRESS LD STENT INTO THE SHEATH, BUT IT WAS DIFFICULT TO ADVANCE THROUGH THE HUB. THE PHYSICIAN INSPECTED A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20675098 ↗
  • MalfunctionMay 2, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT STENT DAMAGE OCCURRED. THE STENOSED TARGET LOCATION WAS LOCATED IN THE ILIAC ARTERY. AN 8.0X60X75CM EXPRESS LD ILIAC / BILIARY STENT WAS SELECTED FOR USE. HOWEVER, DURING PREPARATION AND PRIOR TO INSERTION, THE STENT APPEARED TO BE FLARED ON THE PROXIMAL END. THE PROCEDURE WAS COMPLETED WITH A NEW STENT. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19234524 ↗
  • MalfunctionAug 26, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED. THE TARGET LESION WAS LOCATED IN THE NON-TORTUOUS, MODERATELY CALCIFIED ON RIGHT, AND SEVERELY CALCIFIED ON LEFT BILATERAL COMMON ILIAC ARTERIES. A 8.0X60X75CM EXPRESS LD ILIAC/BILIARY STENT WAS SELECTED FOR USE. HOWEVER, DURING PREPARATION, THE STENT DISMOUNTED FROM THE BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15306459 ↗
  • MalfunctionFeb 21, 2019Unsealed Device Packaging; Unsealed Device Packaging

    IT WAS REPORTED THAT UNSEALED DEVICE PACKING WAS ENCOUNTERED. DURING PREPARATION OF A 8.0X60X75CM EXPRESS LD ILIAC / BILIARY STENT, IT WAS NOTED THAT THE INNER POUCH WAS NOT COMPLETELY SEALED. THERE WAS NO VISIBLE HOLE OR TEAR IN THE POUCH. THE DEVICE WAS NOT USED INSIDE THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH DIFFERENT DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.

    Patient problems: No Patient Involvement

    MAUDE report 8358514 ↗