Quantum™ Maverick™ H7493808008500Recalls & reported problems

Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2

All models of Quantum Maverick (36)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729392620), received between Oct 9, 2018 and Aug 27, 2020 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

5 other recalls cover devices cleared under the same FDA submission (P860019); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break2
  • Failure to Advance2

Patient problems

  • No Consequences Or Impact To Patient4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 27, 2020Break; Break

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED.THE TARGET LESION WAS LOCATED IN THE RIGHT CORONARY ARTERY. A 5.0MM X 8MM QUANTUM MAVERICK BALLOON CATHETER WAS USED; HOWEVER, THE DEVICE WAS FRACTURED. THE DEVICE WAS SIMPLY PULLED OUT AND REMOVED FROM THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND PATIENT'S STATUS WAS STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10462682 ↗
  • MalfunctionJun 25, 2019Failure to Advance; Failure to Advance

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 29-MAY-2019. IT WAS REPORTED THAT CROSSING DIFFICULTIES WERE ENCOUNTERED. THE 90% STENOSED, 8MMX3.5MM, TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND CALCIFIED RIGHT CORONARY ARTERY. A 5.0MM X 8MM QUANTUM MAVERICK BALLOON CATHETER WAS ADVANCED BUT FAILED TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED SHAFT DETACHMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8733055 ↗
  • MalfunctionNov 26, 2018Break; Break

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. A 5.0MM X 8MM QUANTUM MAVERICK BALLOON CATHETER WAS SELECTED BUT DELIVERY SHAFT OF THE BALLOON WAS FRACTURED NEAR THE HUB. THE DEVICE WAS SIMPLY REMOVED AS A WHOLE ALL TOGETHER. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8103598 ↗
  • MalfunctionOct 9, 2018Failure to Advance; Failure to Advance

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 14-SEP-2018. IT WAS REPORTED THAT CROSSING DIFFICULTIES WERE ENCOUNTERED. A 5.0MM X 8MM QUANTUM MAVERICK BALLOON CATHETER WAS ADVANCED BUT FAILED TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED SHAFT BREAK. RETURNED PRODUCT CONSISTED OF A QUANTUM MAVERICK BALLOON CATHETER WITH A MANDREL. THE OUTER SHAFT, INNER SHAFT, BALLOON AND TIP WERE VISUALLY AND MICROSCOPICALLY EXAMINED. VISUAL EXAMINATION REVEALED SEVERAL KINKS ALONG THE HYPOTUBE AND SHAFT. THE HYPOTUBE IS SEPARATED 90.8CM FROM THE HUB. THE BALLOON IS TIGHTLY FOLDED AND THERE IS BLOOD ON THE BALLOON FOLDS. MICROSCOPIC EXAMINATION REVEALED THAT THE TIP IS DAMAGED. INSPECTION OF THE REMAINDER OF THE DEVICE PRESENTED NO OTHER DAMAGE OR IRREGULARITIES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7948564 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P860019. One submission often covers many models, so check the product and lot details in each record.

  • Z-2165-2008 · May 1, 2007 · Terminated
    Boston Scientific Scimed Maverick 2 Monorail" PTCA Catheter, 3.0 mm 15 mm REF 38928-1530 Lot 9565194 Use before 2010-04 Indicated for balloon dilation of the stenotic portion o
  • Z-2166-2008 · Nov 1, 2005 · Terminated
    Boston Scientific Scimed Maverick 2 Monorail 12mm Balloon Catheter, 2.5 mm x 12 mm Made in Ireland Ballybrit Business Park, Galway, Ireland REF 38928-1225 Use before 2007-01 UPN H
  • Z-0030-04 · Aug 20, 2003 · Terminated
    Ace Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 2cm Tip 20 mm Balloon, Catalog Number H74902071011
  • Z-0031-04 · Aug 20, 2003 · Terminated
    Ace Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 2cm Tip 20 mm Balloon, Catalog Number H74902071051
  • Z-0032-04 · Aug 20, 2003 · Terminated
    Ace Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 1 cm Tip and 20 mm Balloon, Catalog Number H74903628151

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →