Quantum™ Maverick™ H7493808012220Recalls & reported problems

Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2

All models of Quantum Maverick (36)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729392644), received between Jul 13, 2017 and Jul 9, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

5 other recalls cover devices cleared under the same FDA submission (P860019); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1
  • Failure to Advance1

Patient problems

  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 9, 2019Break; Break

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. DURING UNPACKING OF A 2.25MM X 12MM QUANTUM MAVERICK BALLOON CATHETER, IT WAS NOTED THAT THE CATHETER OF THE BALLOON WAS BROKEN. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE.

    Patient problems: No Patient Involvement

    MAUDE report 8773284 ↗
  • MalfunctionJul 13, 2017Failure to Advance; Failure to Advance

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 28-JUN-2017. IT WAS REPORTED THAT CROSSING DIFFICULTIES WERE ENCOUNTERED. THE TARGET LESION WAS LOCATED IN A CORONARY ARTERY. A 2.25 MM X 12 MM QUANTUM¿ MAVERICK¿ BALLOON CATHETER WAS ADVANCED BUT FAILED TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A LONGITUDINAL BALLOON TEAR.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6712215 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P860019. One submission often covers many models, so check the product and lot details in each record.

  • Z-2165-2008 · May 1, 2007 · Terminated
    Boston Scientific Scimed Maverick 2 Monorail" PTCA Catheter, 3.0 mm 15 mm REF 38928-1530 Lot 9565194 Use before 2010-04 Indicated for balloon dilation of the stenotic portion o
  • Z-2166-2008 · Nov 1, 2005 · Terminated
    Boston Scientific Scimed Maverick 2 Monorail 12mm Balloon Catheter, 2.5 mm x 12 mm Made in Ireland Ballybrit Business Park, Galway, Ireland REF 38928-1225 Use before 2007-01 UPN H
  • Z-0031-04 · Aug 20, 2003 · Terminated
    Ace Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 2cm Tip 20 mm Balloon, Catalog Number H74902071051
  • Z-0032-04 · Aug 20, 2003 · Terminated
    Ace Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 1 cm Tip and 20 mm Balloon, Catalog Number H74903628151
  • Z-0030-04 · Aug 20, 2003 · Terminated
    Ace Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 2cm Tip 20 mm Balloon, Catalog Number H74902071011

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →

Sources and updates