Equalizer™ M001171070Recalls & reported problems
Boston Scientific Corporation · Occlusion Balloon Catheter · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729402374), received between Sep 11, 2018 and Sep 11, 2018 — classified as 1 injury.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Material Rupture1
Patient problems
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 11, 2018Material Rupture; Detachment of Device or Device Component; Material Rupture; Detachment of Device or Device Component
IT WAS REPORTED THAT BALLOON RUPTURE AND BALLOON DETACHMENT OCCURRED. THE TARGET LESION WAS A STENT GRAFT LOCATED IN THE MODERATELY TORTUOUS VESSEL IN THE ABDOMINAL AORTA. A 27/7/2/65 EQUALIZER BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE THIRD INFLATION FOR 80 SECONDS, THE BALLOON RUPTURED. UPON DEVICE REMOVAL, IT WAS NOTED THAT A QUARTER OF THE BALLOON PART GOT SEPARATED. WHEN THE PHYSICIAN CHECKED INSIDE THE SHEATH, THE DETACHED BALLOON MATERIAL COULD NOT BE FOUND. ANGIOGRAPHY WAS PERFORMED ALL THE WAY TO THE PERIPHERAL AND IT WAS CONFIRMED THAT THERE WERE NO ISSUE WITH THE BLOOD FLOW. THE DETACHED FRAGMENTS OF THE BALLOON WERE NOT FOUND. THE PROCEDURE WAS COMPLETED WITH 17-107 EQUALIZER BALLOON CATHETER. NO FURTHER PATIENT COMPLCATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 7866001 ↗
Other Catheter, intravascular occluding, temporary devices
- Pruitt F3 Carotid Shunt 2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary1 death158 reports
- GORE Molding Occlusion Balloon Catheter MOB37W. L. Gore & Associates, Inc. · Catheter, Intravascular Occluding, Temporary6 deaths152 reports
- Bridge Occlusion Balloon 590-001Spectranetics Corporation · Catheter, Intravascular Occluding, Temporary4 deaths57 reports
- Pruitt F3 Carotid Shunt e2012-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary53 reports
- Pruitt F3 Carotid Shunt 2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary46 reports
- Equalizer™ M001171100Boston Scientific Corporation · Catheter, intravascular occluding, temporary18 reports
- TRANSFORM M003SSC07070Stryker Corporation · Catheter, intravascular occluding, temporary17 reports
- Pruitt F3 Carotid Shunt e2013-10Lemaitre Vascular, Inc. · Catheter, Intravascular Occluding, Temporary17 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.