Mach1™ M001196480Recalls & reported problems
Boston Scientific Corporation · Guide Catheter · Class 2
All models of Mach1 (28)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729403531), received between May 4, 2020 and Aug 23, 2022 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Material Integrity Problem2
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Entrapment of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 23, 2022Detachment of Device or Device Component; Material Integrity Problem; Detachment of Device or Device Component; Material Integrity Problem
IT WAS REPORTED THAT A DISTAL TIP BREAK OCCURRED. THE 100% STENOSED, TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A MACH 1 GUIDE CATHETER AND A NON-BSC DILATION CATHETER WERE SELECTED FOR USE. THE DISTAL TIP BROKE AS BOTH DEVICES WERE BEING ADVANCED TO THE LESION THAT WAS DESCRIBED TO BE HARD. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT CONDITION FOLLOWING THE PROCEDURE WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15279239 ↗MalfunctionAug 23, 2022Detachment of Device or Device Component; Material Integrity Problem; Detachment of Device or Device Component; Material Integrity Problem
IT WAS REPORTED THAT A DISTAL TIP BREAK OCCURRED. THE 100% STENOSED, TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A MACH 1 GUIDE CATHETER AND A NON-BSC DILATION CATHETER WERE SELECTED FOR USE. THE DISTAL TIP BROKE AS BOTH DEVICES WERE BEING ADVANCED TO THE LESION THAT WAS DESCRIBED TO BE HARD. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT CONDITION FOLLOWING THE PROCEDURE WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15279207 ↗MalfunctionJul 20, 2021Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A TIP DETACHMENT OCCURRED. A 6F MACH1 GUIDE CATHETER WAS SELECTED FOR USE IN A TREATMENT PROCEDURE. THE TARGET 50% STENOSED TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS RIGHT SUPERFICIAL FEMORAL ARTERY. IT WAS OBSERVED THAT THE TIP WAS DETACHED. THE DEVICE WAS NOT USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12200027 ↗MalfunctionMay 4, 2020Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6F MACH1 WAS ADVANCED VIA A CONTRALATERAL APPROACH OVER THE STEEP ILIAC BIFURCATION TO ASSIST IN INTRAVASCULAR ULTRASOUND (IVUS) SUPPORTED BY A NON-BSC SHEATH. HOWEVER, DURING IVUS CATHETER REMOVAL, IT FELT LIKE IT WAS STUCK AND WAS REMOVED FORCIBLY. THE MACH1 WAS THEN ATTEMPTED TO BE REMOVED BUT IT WAS STUCK INSIDE THE NON-BSC SHEATH. IT COULD NOT BE REMOVED EVEN BY PULLING FORCIBLY, AND THE MACH 1 BROKE. THE DEVICE WAS COMPLETELY REMOVED FROM THE PATIENT'S BODY TOGETHER WITH THE NON-BSC SHEATH. THE PROCEDURE WAS COMPLETED WITH ANOTHER SYSTEM. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10020502 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page