Polaris™ Loop M0061552310Recalls & reported problems
Boston Scientific Corporation · Ureteral Stent · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 08714729421078), received between Aug 6, 2020 and Apr 21, 2026 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Break2
- Detachment of Device or Device Component2
- Difficult to Remove2
- Material Split, Cut or Torn2
- Migration2
- Defective Device1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient1
- Urinary Tract Infection1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 21, 2026Break
IT WAS REPORTED THAT DURING A LITHOTRIPSY FOR LEFT RENAL CALCULI, A POLARIS URETERAL STENT WAS TO BE USED. AFTER UNPACKING, IT WAS FOUND THAT THE STENT WAS FRACTURED INTO TWO PARTS. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24947744 ↗MalfunctionJan 6, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT, DURING A URETEROSCOPY PROCEDURE USING A POLARIS LOOP URETERAL STENT, THE LOOP WAS BROKEN WHEN THE PHYSICIAN PUSHED THE DEVICE INTO THE BLADDER. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23983542 ↗MalfunctionNov 3, 2025Nonstandard Device; Defective Device; Material Integrity Problem; Material Split, Cut or Torn; Nonstandard Device; Defective Device; Material Integrity Problem; Material Split, Cut or Torn
IT WAS REPORTED THAT A POLARIS LOOP URETERAL STENT WAS USED IN A URETEROSCOPIC LITHOTRIPSY PROCEDURE. DURING THE PROCEDURE, THE LOOP WAS ENTANGLED WITH THE SUTURE THREAD CAUSING IT TO BREAK WHILE BEING UNTIED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23459999 ↗MalfunctionJul 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE PROCEDURE, THE LOOP WAS ENTANGLED WITH COTTON THREAD, CAUSING IT TO BREAK WHEN BEING UNTIED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THE PATIENT CONDITION FOLLOWING PROCEDURE WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22555173 ↗MalfunctionOct 25, 2024Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, DURING PREPARATION PRIOR TO TRANSURETEROSCOPIC LITHOTRIPSY, IT WAS NOTICED THAT THE STENT WAS FRACTURED. ANOTHER OF THE SAME STENT WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED. THE CONDITION OF THE PATIENT AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20533824 ↗InjuryApr 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS LOOP URETERAL STENT WAS IMPLANTED DURING A STENT PLACEMENT FOR STONE MANAGEMENT, LASER LITHOTRIPSY PROCEDURE IN THE RIGHT URETER ON (B)(6) 2022, AS PART OF THE U0652 DOUBLE-J REGISTRY CLINICAL STUDY. ON (B)(6) 2022, THE STENT WAS SUCCESSFULLY IMPLANTED, AND A MEDICATION OF ALPHA BLOCKER, ANTIBIOTIC, AND NARCOTIC ANALGESICS (OPIOIDS) WAS GIVEN FOR THE PATIENT. ON (B)(6) 2022, IT WAS NOTED THAT THE PATIENT EXPERIENCED URINARY TRACT INFECTION AND WAS HOSPITALIZED FOR NINE DAYS. ON (B)(6) 2022, THE EVENT WAS CONSIDERED RESOLVED. THERE WAS NO NEW STENT IMPLANTED.
Patient problems: Urinary Tract Infection
MAUDE report 16714282 ↗MalfunctionApr 1, 2022Difficult to Remove; Migration; Difficult to Remove; Migration
IT WAS REPORTED THAT A POLARIS LOOP URETERAL STENT WAS IMPLANTED DURING AN URETEROSCOPIC LITHOTRIPSY PROCEDURE PERFORMED ON (B)(6) 2022. THE PATIENT WAS STENTED WITH A POLARIS LOOP URETERAL STENT 6X22CM AND FOR AROUND ONE MONTH, A REMOVAL PROCEDURE WAS PERFORMED ON (B)(6) 2022. DURING PROCEDURE, WHEN THE PHYSICIAN ATTEMPTED TO REMOVED THE IMPLANTED POLARIS LOOP URETERAL STENT, IT WAS FOUND THAT THE COIL PART MIGRATED TO THE PATIENT'S URETERIC OFFICE. THEY WERE UNABLE TO REMOVE THE MIGRATED COIL WITH FORCEPS, THEREFORE THE PHYSICIAN INSERTED CYSTOSCOPE IN ORDER TO VIEW AND WAS ABLE TO REMOVED THE MIGRATED COIL SUCCESSFULLY. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE AS "PATIENT DOES NOT FEEL ANY UNUSUAL PAIN". THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13981930 ↗MalfunctionApr 1, 2022Difficult to Remove; Migration; Difficult to Remove; Migration
IT WAS REPORTED THAT A POLARIS LOOP URETERAL STENT WAS IMPLANTED DURING AN RETROGRADE INTRA RENAL SURGERY PROCEDURE PERFORMED ON (B)(6) 2022. IT WAS REPORTED THAT THE PATIENT WAS STENTED WITH POLARIS LOOP 6X22CM FOR AROUND 4 DAYS AND A REMOVAL PROCEDURE WAS PERFORMED ON (B)(6) 2022. DURING PROCEDURE, WHEN THE PHYSICIAN ATTEMPTED TO REMOVED THE IMPLANTED POLARIS LOOP URETERAL STENT, IT WAS FOUND THAT THE COIL PART MIGRATED TO THE PATIENT'S URETERIC OFFICE. THEY WERE UNABLE TO REMOVE THE MIGRATED COIL WITH FORCEPS, THEREFORE THE PHYSICIAN INSERTED CYSTOSCOPE IN ORDER TO VIEW AND WAS ABLE TO REMOVED THE MIGRATED COIL SUCCESSFULLY. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE AS "PATIENT DOES NOT FEEL ANY UNUSUAL PAIN". THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13972221 ↗MalfunctionAug 6, 2020Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS LOOP URETERAL STENT WAS USED DURING A LITHOTOMY UNDER URETEROSCOPE PROCEDURE FOR A STONE IN THE URETER, PERFORMED ON (B)(6), 2020. ACCORDING TO THE COMPLAINANT, DURING WITHDRAWAL, THE PHYSICIAN NOTICED THAT THE STENT WAS FRACTURED WHEN REMOVED FROM INSIDE THE PATIENT'S BODY. THE FRACTURED STENT WAS PULLED OUT AND NO FRAGMENTS WERE LEFT INSIDE THE PATIENT. ANOTHER URETERAL STENT WAS IMPLANTED AND SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10375894 ↗
Other STENT, URETERAL devices
- Bard® InLay Optima® Ureteral Stent 788426C. R. Bard, Inc. · STENT, URETERAL1 recall100 reports
- Bard® InLay Optima® Ureteral Stent 788626C. R. Bard, Inc. · STENT, URETERAL122 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777626C. R. Bard, Inc. · STENT, URETERAL118 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777426C. R. Bard, Inc. · STENT, URETERAL108 reports
- Bard® InLay® Ureteral Stent 778626C. R. Bard, Inc. · STENT, URETERAL100 reports
- Percuflex™ Plus M0061752530Boston Scientific Corporation · STENT, URETERAL1 death99 reports
- Contour VL™ M0061801560Boston Scientific Corporation · STENT, URETERAL1 recall45 reports
- Polaris™ Ultra M0061921330Boston Scientific Corporation · STENT, URETERAL71 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page