WallFlex™ Colonic M00565080Recalls & reported problems
Boston Scientific Corporation · Stent System with Anchor Lock Delivery System · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729456551), received between Jun 14, 2018 and Dec 11, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Failure1
- Break1
- Material Deformation1
- Material Integrity Problem1
- Positioning Failure1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 11, 2024Use of Device Problem; Material Deformation; Material Integrity Problem; Use of Device Problem; Material Deformation; Material Integrity Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A WALLFLEX ENTERAL COLONIC STENT WAS TO BE USED IN THE INTESTINAL TRACT TO TREAT A BENIGN STENOSIS DURING AN ENTERECTOMY WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING PREPARATION, IT WAS NOTED THAT THE STENT TIP WAS BENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER WALLFLEX ENTERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: IT WAS REPORTED THAT THE WALLFLEX ENTERAL STENT WAS USED TO TREAT A BENIGN INTESTINAL STENOSIS. HOWEVER, PER THE WALLFLEX COLONIC STENT SYSTEM WITH ANCHOR LOCK DELIVERY SYSTEM INSTRUCTIONS FOR USE (IFU), THE DEVICE IS INDICATED FOR THE PALLIATIVE TREATMENT OF COLONIC STRICTURES PRODUCED BY MALIGNANT NEOPLASM AND TO RELIEVE LARGE BOWEL OBSTRUCTION PRIOR TO COLECTOMY IN PATIENTS WITH MALIGNANT STRICTURES. THE DEVICE IS NOT INDICATED FOR THE TREATMENT OF BENIGN STRICTURES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20903031 ↗MalfunctionJun 22, 2023Break; Positioning Failure; Break; Positioning Failure
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6), 2023, THAT A WALLFLEX ENTERAL COLONIC STENT WAS TO BE IMPLANTED IN THE COLON TO TREAT A STRICTURE DURING A COLONIC STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2023. THE PATIENT'S ANATOMY WAS NOT TIGHT AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING STENT DEPLOYMENT, THE STAINLESS-STEEL HANDLE BROKE IN HALF, AND THE STENT COULD NOT BE DEPLOYED. THE STENT WAS REMOVED FROM THE PATIENT, AND THE PROCEDURE WAS NOT COMPLETED BECAUSE ANOTHER OF THE SAME DEVICE WAS UNAVAILABLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17187546 ↗InjuryJun 14, 2018Activation Failure; Activation Failure
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON MAY 23, 2018 THAT A WALLFLEX COLONIC STENT WAS TO BE USED TO TREAT A COLONIC STENOSIS DURING A STENTING PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE STENT WAS ABLE TO BE DEPLOYED. HOWEVER, THE STENT FAILED TO FULLY EXPAND ON ONE END. THE STENT WAS REMOVED WITH FORCEPS AND THE PROCEDURE WAS COMPLETED WITH ANOTHER WALLFLEX COLONIC STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT¿S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. X-RAY IMAGE OF THE WALLFLEX COLONIC DEPLOYED WITHIN THE PATIENT WAS RECEIVED AND SEEMS TO SHOW AN UNEXPANDED PORTION OF THE STENT WITHIN THE ANATOMY. VIDEO OF THE PHYSICIAN MANIPULATING THE STENT AFTER IT WAS REMOVED FROM PATIENT WAS ALSO RECEIVED AND APPEARS TO SHOW THAT THE FLARED END OF THE STENT WAS SLIGHTLY UNEXPANDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7601963 ↗
Other STENT, COLONIC, METALIC, EXPANDABLE devices
- WallFlex™ Colonic M00565050Boston Scientific Corporation · STENT, COLONIC, METALIC, EXPANDABLE2 deaths63 reports
- Evolution G48028Cook Ireland Ltd · STENT, COLONIC, METALIC, EXPANDABLE45 reports
- Evolution G48038Cook Ireland Ltd · STENT, COLONIC, METALIC, EXPANDABLE45 reports
- WallFlex™ Colonic M00565060Boston Scientific Corporation · STENT, COLONIC, METALIC, EXPANDABLE2 deaths33 reports
- WallFlex™ Colonic M00565040Boston Scientific Corporation · STENT, COLONIC, METALIC, EXPANDABLE1 death23 reports
- WALLSTENT™ Enteral Endoprosthesis M00565590Boston Scientific Corporation · STENT, COLONIC, METALIC, EXPANDABLE15 reports
- WallFlex™ Colonic M00565110Boston Scientific Corporation · STENT, COLONIC, METALIC, EXPANDABLE3 deaths14 reports
- WallFlex™ Colonic M00565120Boston Scientific Corporation · STENT, COLONIC, METALIC, EXPANDABLE11 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page