WallFlex™ Colonic M00565080Recalls & reported problems

Boston Scientific Corporation · Stent System with Anchor Lock Delivery System · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729456551), received between Jun 14, 2018 and Dec 11, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure1
  • Break1
  • Material Deformation1
  • Material Integrity Problem1
  • Positioning Failure1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 11, 2024Use of Device Problem; Material Deformation; Material Integrity Problem; Use of Device Problem; Material Deformation; Material Integrity Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A WALLFLEX ENTERAL COLONIC STENT WAS TO BE USED IN THE INTESTINAL TRACT TO TREAT A BENIGN STENOSIS DURING AN ENTERECTOMY WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING PREPARATION, IT WAS NOTED THAT THE STENT TIP WAS BENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER WALLFLEX ENTERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: IT WAS REPORTED THAT THE WALLFLEX ENTERAL STENT WAS USED TO TREAT A BENIGN INTESTINAL STENOSIS. HOWEVER, PER THE WALLFLEX COLONIC STENT SYSTEM WITH ANCHOR LOCK DELIVERY SYSTEM INSTRUCTIONS FOR USE (IFU), THE DEVICE IS INDICATED FOR THE PALLIATIVE TREATMENT OF COLONIC STRICTURES PRODUCED BY MALIGNANT NEOPLASM AND TO RELIEVE LARGE BOWEL OBSTRUCTION PRIOR TO COLECTOMY IN PATIENTS WITH MALIGNANT STRICTURES. THE DEVICE IS NOT INDICATED FOR THE TREATMENT OF BENIGN STRICTURES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20903031 ↗
  • MalfunctionJun 22, 2023Break; Positioning Failure; Break; Positioning Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6), 2023, THAT A WALLFLEX ENTERAL COLONIC STENT WAS TO BE IMPLANTED IN THE COLON TO TREAT A STRICTURE DURING A COLONIC STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2023. THE PATIENT'S ANATOMY WAS NOT TIGHT AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING STENT DEPLOYMENT, THE STAINLESS-STEEL HANDLE BROKE IN HALF, AND THE STENT COULD NOT BE DEPLOYED. THE STENT WAS REMOVED FROM THE PATIENT, AND THE PROCEDURE WAS NOT COMPLETED BECAUSE ANOTHER OF THE SAME DEVICE WAS UNAVAILABLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17187546 ↗
  • InjuryJun 14, 2018Activation Failure; Activation Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON MAY 23, 2018 THAT A WALLFLEX COLONIC STENT WAS TO BE USED TO TREAT A COLONIC STENOSIS DURING A STENTING PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE STENT WAS ABLE TO BE DEPLOYED. HOWEVER, THE STENT FAILED TO FULLY EXPAND ON ONE END. THE STENT WAS REMOVED WITH FORCEPS AND THE PROCEDURE WAS COMPLETED WITH ANOTHER WALLFLEX COLONIC STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT¿S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. X-RAY IMAGE OF THE WALLFLEX COLONIC DEPLOYED WITHIN THE PATIENT WAS RECEIVED AND SEEMS TO SHOW AN UNEXPANDED PORTION OF THE STENT WITHIN THE ANATOMY. VIDEO OF THE PHYSICIAN MANIPULATING THE STENT AFTER IT WAS REMOVED FROM PATIENT WAS ALSO RECEIVED AND APPEARS TO SHOW THAT THE FLARED END OF THE STENT WAS SLIGHTLY UNEXPANDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7601963 ↗

Other STENT, COLONIC, METALIC, EXPANDABLE devices

All STENT, COLONIC, METALIC, EXPANDABLE devices →