Skip to content

NA M006550340Recalls & reported problems

Boston Scientific Corporation · Adapter for Wolf Hysteroscope · Class 1

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729469773), received between Apr 5, 2024 and Apr 5, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Degraded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded1
  • Use of Device Problem1

Patient problems

  • Burn(s)1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 5, 2024Degraded; Use of Device Problem; Degraded; Use of Device Problem

    BOSTON SCIENTIFIC COMPLAINT DEPARTMENT FORWARDED A MEDWATCH REPORT, 2302300000-2024-8002, TO THE MINERVA SURGICAL COMPLAINT DEPARTMENT ON MARCH 21ST, 2024. THE MEDWATCH REPORT STATED "PATIENT PRESENTED TO THE OPERATING ROOM (OR) FOR A PLANNED DILATION AND CURETTAGE (D&C), HYDROTHERMAL ABLATION (HTA) PROCEDURE. DURING THE SURGERY, THE HTA EQUIPMENT FAILED, ALLOWING THE SCOPE TO BECOME INCREDIBLY HOT AND BURNED THIS PATIENT'S VAGINAL OPENING AND CERVIX. WE LEARNED THAT THE WOLF HYSTEROSCOPE ADAPTER WAS SUPPOSED TO BE REPLACED AT EVERY 40 USES WHICH WE DID NOT DO. WE HAVE DISCARDED ALL OF OUR SUPPLIES FROM 2005 AND HAVE ORDERED ALL NEW ONES."

    Patient problems: Burn(s)

    MAUDE report 19050595 ↗

Other Hysteroscope accessories devices

All Hysteroscope accessories devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.