NA M006550340Recalls & reported problems
Boston Scientific Corporation · Adapter for Wolf Hysteroscope · Class 1
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729469773), received between Apr 5, 2024 and Apr 5, 2024 — classified as 1 injury.
The most frequently reported specific device problem is “Degraded” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded1
- Use of Device Problem1
Patient problems
- Burn(s)1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 5, 2024Degraded; Use of Device Problem; Degraded; Use of Device Problem
BOSTON SCIENTIFIC COMPLAINT DEPARTMENT FORWARDED A MEDWATCH REPORT, 2302300000-2024-8002, TO THE MINERVA SURGICAL COMPLAINT DEPARTMENT ON MARCH 21ST, 2024. THE MEDWATCH REPORT STATED "PATIENT PRESENTED TO THE OPERATING ROOM (OR) FOR A PLANNED DILATION AND CURETTAGE (D&C), HYDROTHERMAL ABLATION (HTA) PROCEDURE. DURING THE SURGERY, THE HTA EQUIPMENT FAILED, ALLOWING THE SCOPE TO BECOME INCREDIBLY HOT AND BURNED THIS PATIENT'S VAGINAL OPENING AND CERVIX. WE LEARNED THAT THE WOLF HYSTEROSCOPE ADAPTER WAS SUPPOSED TO BE REPLACED AT EVERY 40 USES WHICH WE DID NOT DO. WE HAVE DISCARDED ALL OF OUR SUPPLIES FROM 2005 AND HAVE ORDERED ALL NEW ONES."
Patient problems: Burn(s)
MAUDE report 19050595 ↗
Other Hysteroscope accessories devices
- Olympus A4832OLYMPUS Winter & Ibe GmbH · Hysteroscope Accessories12 reports
- OMNI OMNI-S-001Hologic, Inc. · Hysteroscope accessories8 reports
- Olympus A4829OLYMPUS Winter & Ibe GmbH · Hysteroscope Accessories8 reports
- Olympus A4825OLYMPUS Winter & Ibe GmbH · Hysteroscope Accessories8 reports
- CLICKLINE 26312Karl Storz GmbH & Co. KG · Hysteroscope accessories6 reports
- n.a. 26153BIKarl Storz GmbH & Co. KG · Hysteroscope accessories3 reports
- OMNI OMNI-G-001Hologic, Inc. · Hysteroscope accessories1 report
- Olympus A4827OLYMPUS Winter & Ibe GmbH · Hysteroscope Accessories1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.