Apex™ H7493895940200Recalls & reported problems
Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2
All models of Apex (30)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729475910), received between Dec 19, 2016 and May 1, 2019 — classified as 1 death, 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Entrapment of Device” (1 report).
5 other recalls cover devices cleared under the same FDA submission (P860019); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Entrapment of Device1
- Material Rupture1
- Physical Resistance/Sticking1
Patient problems
- No Consequences Or Impact To Patient2
- Cardiogenic Shock1
- Death1
- Perforation of Vessels1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 1, 2019Physical Resistance/Sticking; Physical Resistance/Sticking
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2019. IT WAS REPORTED THAT LOADING DIFFICULTIES WERE ENCOUNTERED. A 2.00MM X 40MM APEX BALLOON CATHETER WAS SELECTED FOR USE TO DILATE THE LESION. HOWEVER, IT WAS NOTED THAT THE DEVICE WAS UNABLE TO BE LOADED ONTO THE WIRE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED PROLAPSE AND HOLE IN GUIDEWIRE LUMEN AND PUNCTURE MARK IN THE INFLATION LUMEN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8569982 ↗DeathMar 29, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SAME CASE AS MDR ID #: 2134265-2018-02589, 2134265-2018-02590, 2134265-2018-02591, 2134265-2018-02592, 2134265-2018-02593, 2134265-2018-02628, 2134265-2018-03162, 2134265-2018-03163, 2134265-2018-03164, 2134265-2018-03165, 2134265-2018-03166, 2134265-2018-03167, 2134265-2018-03168, 2134265-2018-03169, 2134265-2018-03173. IT WAS REPORTED THAT THE PATIENT DIED POST PROCEDURE. THE PATIENT PRESENTED FROM HOSPICE CARE. THE PATIENT WAS INTUBATED AND HAD A NON-BSC HEART PUMP IN PLACE. THE TARGET LESIONS WERE A CHRONIC TOTAL OCCLUSION OF THE RIGHT CORONARY ARTERY (RCA), THE LEFT ANTERIOR DESCENDING ARTERY (LAD) AND THE SEVERELY TORTUOUS CIRCUMFLEX. SEVENTEEN BOSTON SCIENTIFIC PRODUCTS WERE USED THROUGHOUT THE PROCEDURE. THE PHYSICIAN STARTED WITH THE CIRCUMFLEX AND EXPERIENCED A VESSEL PERFORATION CAUSED BY A NON-BSC GUIDEWIRE. THE PERFORATION WAS MINIMAL AND WAS NOT TREATED. THE LAD AND LEFT MAIN ARTERIES WERE SUCCESSFULLY STENTED WITH 2.50X38MM, 3.00X38MM AND 3.50X16MM SYNERGY STENTS AND BALLOON ANGIOPLASTY WAS PERFORMED. AT THAT POINT, THE PATIENT¿S PRESSURE AND HEART RATE WERE BOTH DROPPING. THE HEART RATE WAS INCREASED WITH PACING AND THE PROCEDURE WAS ENDED. THE PATIENT WAS SENT FROM
Patient problems: Cardiogenic Shock; Death
MAUDE report 7383150 ↗MalfunctionFeb 8, 2017Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT CATHETER ENTRAPMENT OCCURRED. A 2.00MM X 40MM APEX BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, THE BALLOON WAS OVERINFLATED WHICH CAUSED THE INNER HYPOTUBE TO FUSE WITH THE NON-BSC GUIDE WIRE AND BECAME STUCK. THE BALLOON CATHETER AND GUIDE WIRE WERE THEN REMOVED TOGETHER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6311901 ↗InjuryDec 19, 2016Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE AND VESSEL PERFORATION OCCURRED. THE TARGET LESION WAS LOCATED IN THE POSTERIOR DESCENDING ARTERY (PDA). A 2.00MM X 40MM APEX¿ BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 12 ATMOSPHERES, THE BALLOON RUPTURED. A PERFORATION ON THE ARTERY WAS THEN NOTED AND A STENT WAS DEPLOYED TO COVER THE PERFORATION. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS FINE.
Patient problems: Perforation of Vessels
MAUDE report 6185163 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P860019. One submission often covers many models, so check the product and lot details in each record.
- Z-2165-2008 · May 1, 2007 · TerminatedBoston Scientific Scimed Maverick 2 Monorail" PTCA Catheter, 3.0 mm 15 mm REF 38928-1530 Lot 9565194 Use before 2010-04 Indicated for balloon dilation of the stenotic portion o
- Z-2166-2008 · Nov 1, 2005 · TerminatedBoston Scientific Scimed Maverick 2 Monorail 12mm Balloon Catheter, 2.5 mm x 12 mm Made in Ireland Ballybrit Business Park, Galway, Ireland REF 38928-1225 Use before 2007-01 UPN H
- Z-0031-04 · Aug 20, 2003 · TerminatedAce Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 2cm Tip 20 mm Balloon, Catalog Number H74902071051
- Z-0030-04 · Aug 20, 2003 · TerminatedAce Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 2cm Tip 20 mm Balloon, Catalog Number H74902071011
- Z-0032-04 · Aug 20, 2003 · TerminatedAce Percutaneous Transluminal Coronary Angioplasty (PTCA) Dilatation Catheter with 1 cm Tip and 20 mm Balloon, Catalog Number H74903628151
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page