Express® LD Iliac / Biliary H74938046730750Recalls & reported problems
Boston Scientific Corporation · Premounted Stent System · Class 3
All models of Express LD Iliac / Biliary (34)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 08714729579663), received between May 15, 2018 and May 1, 2025 — classified as 7 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated8
- Failure to Advance3
- Activation Failure1
- Difficult to Remove1
- Entrapment of Device1
- Material Separation1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Consequences Or Impact To Patient3
- No Known Impact Or Consequence To Patient2
- No Code Available1
- No Patient Involvement1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
InjuryMay 1, 2025Failure to Advance; Device Dislodged or Dislocated; Failure to Advance; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE STENT DISLODGED, REQUIRING INTERVENTION. THE TARGET LESION WAS LOCATED IN THE TORTUOUS AND SEVERELY CALCIFIED OSTIAL STENOSIS OF THE CELIAC ARTERY. VASCULAR ACCESS WAS OBTAINED VIA THE RIGHT COMMON FEMORAL ARTERY TO THE CELIAC ARTERY. A 6F NON-BOSTON SCIENTIFIC GUIDE SHEATH WAS PLACED, AND A 7.0 X 30 X 75 CM EXPRESS LD STENT BALLOON EXPANDABLE WAS ADVANCED FOR TREATMENT OVER .035 WIRE. IT WAS NOTED THAT THE GUIDE SHEATH WAS ONLY PARTIALLY INTO THE CELIAC ARTERY, AND AN ATTEMPT WAS MADE TO ADVANCE THE BALLOON/STENT OVER THE WIRE AND PAST THE LESION. HOWEVER, WHEN ADVANCING, THE BALLOON PUSHED FORWARD, CAUSING THE WIRE TO LOSE ACCESS IN THE CELIAC ARTERY, AND MOVE FORWARD IN THE AORTA. WHEN ATTEMPTING TO PULL BACK THE STENT INTO THE SHEATH, THE STENT DISLODGED FROM THE BALLOON SHAFT. A SNARE WAS USED TO PULL THE STENT BACK TO THE COMMON FEMORAL ARTERY; HOWEVER, A SURGICAL CUTDOWN WAS PERFORMED TO REMOVE THE STENT. THE PROCEDURE WAS CANCELLED, AND THE PATIENT WAS UNDER LOCAL ANESTHESIA. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21945293 ↗MalfunctionSep 23, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE STENT MOVED ON BALLOON THE TARGET LESION WAS LOCATED IN THE CALCIFIED ILIAC ARTERY. A 7.0 X 30 X 75 CM EXPRESS LD STENT BALLOON EXPANDABLE WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE STENT MOVED ON BALLOON. NO PATIENT COMPLICATION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20284364 ↗InjuryApr 17, 2023Material Separation; Activation Failure; Material Separation; Activation Failure
IT WAS REPORTED THAT STENT DID NOT FULLY DEPLOY, AND BALLOON DETACHMENT OCCURRED. THE PATIENT PRESENTED WITH PERIPHERAL ARTERY DISEASE (PAD). THE STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED ILIAC ARTERY. A 7.0X30X75CM EXPRESS LD ILIAC / BILIARY STENT WAS ADVANCED FOR TREATMENT. HOWEVER, THE STENT DID NOT FULLY DEPLOY WHEN THE PHYSICIAN REMOVED THE STENT DELIVERY SYSTEM (SDS). THE BALLOON OF THE SDS TORE ABOUT LEAVING THE BALLOON IN THE BODY. THE PHYSICIAN THEN GAINED ACCESS ON THE OTHER SIDE AND WAS ABLE TO REMOVE THE BALLOON FROM THE BODY WITH A SNARE AND THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND PATIENT FULLY RECOVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16756006 ↗MalfunctionSep 14, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT MOVED ON BALLOON. THE NARROWING TARGET LESION WAS LOCATED IN THE MODERATELY CALCIFIED AND STRAIGHT VESSEL IN PROXIMAL ILIAC. A 7.0X30X75CM EXPRESS LD ILIAC / BILIARY STENT WAS ADVANCED FOR TREATMENT. HOWEVER, AFTER DELIVERING THE STENT TO THE ILIAC ARTERY USING UP AND OVER APPROACH, IT WAS NOTED THAT THE STENT MOVED FORWARD ON THE BALLOON DELIVERY SYSTEM. THE DEVICE WAS REMOVED INTACT AND ONCE IT WAS SAFELY OUT THE SHEATH, IT WAS PHYSICALLY OBSERVED THAT THE STENT SLID SLIGHTLY FORWARD ON BALLOON AND NOT OFF BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER 7.0X30X75CM EXPRESS LD ILIAC / BILIARY STENT. THE CASE OUTCOME WENT GREAT, WITH NO PATIENT CONSEQUENCES AT ALL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15422589 ↗InjuryMar 15, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE STENT DISLODGEMENT OCCURRED. TWO 7.0X30X75CM EXPRESS LD ILIAC / BILIARY STENTS WERE SELECTED TO TREAT THE LESION. HOWEVER, DURING DELIVERY, BOTH STENTS FELL OFF THE BALLOON. SUBSEQUENTLY, A SNARE WAS USED TO REMOVE THE STENTS. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS FINE. IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND TORTUOUS COMMON ILLIAC ARTERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11485379 ↗InjuryMar 15, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE STENT DISLODGEMENT OCCURRED. TWO 7.0X30X75CM EXPRESS LD ILIAC / BILIARY STENTS WERE SELECTED TO TREAT THE LESION. HOWEVER, DURING DELIVERY, BOTH STENTS FELL OFF THE BALLOON. SUBSEQUENTLY, A SNARE WAS USED TO REMOVE THE STENTS. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11485344 ↗InjurySep 17, 2020Failure to Advance; Device Dislodged or Dislocated; Failure to Advance; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED. THE STENOSED TARGET LESION WAS LOCATED IN THE ILIAC ARTERY. A 7.0X30X75CM EXPRESS LD ILIAC / BILIARY STENT WAS ADVANCED FOR TREATMENT BUT COULD NOT CROSS THE LESION. THE PHYSICIAN DECIDED TO WITHDRAW THE STENT TO PRE-DILATE THE LESION. HOWEVER, AFTER REMOVING THE DEVICE OUT OF THE SHEATH, IT WAS NOTED THAT THE STENT WAS NO LONGER ATTACHED TO THE STENT DELIVERY SYSTEM BALLOON. UPON IMAGING, THE STENT WAS FOUND LODGED AT THE END OF THE SHEATH. THE PHYSICIAN WAS NOT ABLE TO REMOVE THE STENT FROM THE ARTERY. THE PHYSICIAN ADVANCED THE STENT NEAR THE LESION TO FULLY DEPLOY IT. THE STENT WAS THEN FULLY DEPLOYED BUT NOT IN THE OPTIMAL LOCATION. SUBSEQUENTLY, A SECOND STENT WAS PLACED TO FULLY COVER THE LESION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 10547119 ↗MalfunctionMay 13, 2020Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT REMOVAL DIFFICULTIES WERE ENCOUNTERED. AFTER A 7FR SHEATH WAS INSERTED, A 7.0X30X75CM EXPRESS LD ILIAC / BILIARY STENT WAS ADVANCED FOR TREATMENT. HOWEVER, IT WAS NOTED THAT THE STENT BECAME STUCK IN THE SHEATH AND WAS UNABLE TO DEPLOY UPON INSERTION. UPON REMOVAL OF THE BALLOON, THE STENT WAS CAPTURED BY THE BALLOON. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10051998 ↗MalfunctionJul 10, 2019Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT CATHETER ENTRAPMENT OCCURRED. A 7.0X30X75CM EXPRESS LD ILIAC/BILIARY STENT WAS SELECTED FOR USE. HOWEVER, DURING PREPARATION, IT WAS NOTED THAT THE STENT GOT CAUGHT OVER THE WIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 8776148 ↗MalfunctionJun 18, 2019Failure to Advance; Physical Resistance/Sticking
DURING PRE-OP INSPECTION, STENT GOT CAUGHT OVER WIRE. UNABLE TO ADVANCE OR REMOVE STENT ON WIRE. THIS DID NOT REACH THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8707084 ↗InjurySep 27, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED. A 7.0X30X75CM EXPRESS LD WAS SELECTED TO TREAT A STENOSED LESION IN THE ILIAC VEIN. DURING THE PROCEDURE WHILE INSIDE THE PATIENT, THE STENT CAME OFF THE BALLOON. THE STENT WAS RECOVERED AND ANOTHER STENT WITH A 135CM SHAFT WAS SELECTED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS FINE POST PROCEDURE. DEVICE EVALUATED BY MANUFACTURER: A VISUAL AND MICROSCOPIC EXAMINATION WAS PERFORMED ON THE RETURNED DEVICE, THE FOLLOWING ATTRIBUTES WERE EXAMINED: THE DELIVERY SYSTEM WAS RETURNED FOR ANALYSIS WITH THE STENT FULLY MOUNTED ONTO THE DELIVERY SYSTEM. A VISUAL AND TACTILE EXAMINATION ON THE CATHETER IDENTIFIED ONE KINK ON THE CATHETER OF THE DEVICE APPROXIMATELY 1MM DISTAL TO THE STRAIN RELIEF. THIS TYPE OF DAMAGE IS CONSISTENT WITH EXCESSIVE FORCE BEING APPLIED TO THE DELIVERY SYSTEM. NO ISSUES WERE NOTED WITH THE SHAFT THAT COULD HAVE CONTRIBUTED TO THE COMPLAINT INCIDENT. A VISUAL AND MICROSCOPIC EXAMINATION WAS PERFORMED ON THE TIP AND MARKERBANDS OF THE DEVICE. NO DAMAGE OR ANY ISSUES WERE NOTED WITH THE TIP OR MARKERBANDS OF THE DEVICE THAT COULD HAVE CONTRIBUTED TO THE COMPLAINT IN
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7915197 ↗InjuryMay 15, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT STENT DISLODGEMENT OCCURRED. A 7.0X30X75CM EXPRESS® LD ILIAC / BILIARY STENT WAS ADVANCED TO TREAT THE LESION. WHEN THE PHYSICIAN ADVANCED THE DEVICE THROUGH A SHEATH OR A GUIDE CATHETER, THE STENT CAME OFF THE BALLOON. THE PHYSICIAN ATTEMPTED TO SNARE THE DISLODGED STENT BUT THE SNARING DEVICE TANGLED IN THE STENT. THE PATIENT THEN WAS TAKEN TO THE OPERATING ROOM FOR A NON-SCHEDULED SURGERY TO REMOVE THE DISLODGED STENT. THE STENT WAS REMOVED AND A DIFFERENT STENT WAS DEPLOYED. THE PROCEDURE WAS THEN COMPLETED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS OKAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7515286 ↗
Other STENT, ILIAC devices
- Cordis S.M.A.R.T. Control Vascular Stent System C10040MLCordis US Corp. · STENT, ILIAC1 recall3 reports
- Absolute Pro 1012534-100Abbott Vascular Inc. · STENT, ILIAC51 reports
- Protégé™ GPS™ SERB65-09-60-80Ev3, Inc · STENT, ILIAC1 recall0 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C09060MLCordis US Corp. · STENT, ILIAC1 recall0 reports
- Absolute Pro 1012535-100Abbott Vascular Inc. · STENT, ILIAC1 death41 reports
- Omnilink Elite 1012631-39Abbott Vascular Inc. · STENT, ILIAC27 reports
- Omnilink Elite 1012632-29Abbott Vascular Inc. · STENT, ILIAC25 reports
- Absolute Pro 1012536-100Abbott Vascular Inc. · STENT, ILIAC21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page