Ultraflex™ Esophageal NG M00513850Recalls & reported problems
Boston Scientific Corporation · Proximal Release Stent System · Class 2 · Implant
All models of Ultraflex Esophageal NG (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 08714729716075), received between Apr 24, 2018 and Dec 20, 2023 — classified as 4 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation, Positioning or Separation Problem8
- Material Deformation2
- Positioning Failure2
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Difficult to Remove1
- Failure to Advance1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient3
- Dysphagia/ Odynophagia1
- Fistula1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
InjuryDec 20, 2023Adverse Event Without Identified Device or Use Problem; Positioning Problem; Adverse Event Without Identified Device or Use Problem; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL COVERED STENT WAS TO BE IMPLANTED IN THE MIDDLE THIRD OF THE ESOPHAGUS TO TREAT MALIGNANT STRICTURE SECONDARY TO ESOPHAGEAL CANCER DURING AN ESOPHAGEAL STENT IMPLANTATION PROCEDURE PERFORMED ON (B)(6) 2023. THE PATIENT'S ANATOMY WAS TIGHT AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THE STENT WAS DIFFICULT TO POSITION, AND IT WAS DEPLOYED NEAR THE PHARYNX. THE STENT WAS REMOVED WITH FORCEPS AND ANOTHER ULTRAFLEX ESOPHAGEAL STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18367045 ↗MalfunctionMay 9, 2023Positioning Failure; Failure to Advance; Material Deformation; Positioning Failure; Failure to Advance; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2023, THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT WAS IMPLANTED TO TREAT ESOPHAGEAL STENOSIS DUE TO ESOPHAGEAL CANCER DURING AN ESOPHAGEAL STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2023. THE PATIENT ANATOMY WAS TORTUOUS AND NOT DILATED PRIOR TO STENT PLACEMENT, DURING THE PROCEDURE, THE DEVICE COULD NOT BE ADVANCED THROUGH THE STENOSIS. DILATION WAS ATTEMPTED AND A NEW MANEUVER WAS PERFORMED TO IMPLANT THE STENT. HOWEVER, THE STENT WAS UNABLE TO CROSS THE STENOSIS EVEN AFTER SEVERAL ATTEMPTS AND THE DISTAL PART OF THE DELIVERY SHAFT WAS NOTED TO BE BENT. THE STENT WAS REMOVED FROM THE PATIENT FULLY COVERED BY THE STENT DEPLOYMENT SUTURE, AND THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16896525 ↗MalfunctionMar 31, 2023Break; Activation, Positioning or Separation Problem; Break; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON MARCH 18, 2023 THAT AN ULTRAFLEX ESOPHAGEAL PROXIMAL COVERED STENT WAS USED TO TREAT A 7CM MALIGNANT MIDDLE ESOPHAGEAL CARCINOMA DURING A STENT PLACEMENT PROCEDURE PERFORMED ON MARCH 17, 2023. THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS DILATED PRIOR TO STENT PLACEMENT PROCEDURE. DURING THE PROCEDURE, THE BLACK DEPLOYMENT SUTURE STOPPED UNRAVELLING AND THE STENT WAS UNABLE TO FULLY DEPLOY. THE STENT WAS PARTIALLY DEPLOYED WHEN IT WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS CANCELLED AND WAS RESCHEDULED AFTER A WEEK. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. **ADDITIONAL INFORMATION RECEIVED ON OCTOBER 10, 2023 AND OCTOBER 15, 2023** THE STENT WAS PARTIALLY RELEASED AND THE BLACK DEPLOYMENT SUTURE WAS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16657213 ↗InjurySep 15, 2022Material Separation; Material Separation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON AUGUST 22, 2022, THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT WAS IMPLANTED TO TREAT A STRICTURE DUE TO ESOPHAGEAL CANCER DURING AN ESOPHAGEAL STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6), 2022. DURING THE PROCEDURE, THE ULTRAFLEX ESOPHAGEAL STENT WAS SUCCESSFULLY DEPLOYED. HOWEVER, IT WAS NOTED THAT THE SILASTIC BAND ON THE DELIVERY SYSTEM FELL INSIDE THE PATIENT. THE SILASTIC BAND WAS REMOVED USING FORCEPS. THE STENT REMAINED IMPLANTED AND THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15433028 ↗MalfunctionJan 26, 2022Positioning Failure; Activation, Positioning or Separation Problem; Positioning Failure; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ULTRALFEX ESOPHAGEAL PROXIMAL RELEASE COVERED STENT WAS TO BE IMPLANTED TO TREAT AN ESOPHAGEAL CANCER DURING AN ESOPHAGEAL STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE STENT COULD NOT BE DEPLOYED. THE STENT WAS FULLY COVERED BY THE DEPLOYMENT SUTURE WHEN IT WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER ULTRAFLEX ESOPHAGEAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED DUE TO THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: THIS COMPLAINT HAS BEEN DEEMED MDR REPORTABLE BASED ON THE INVESTIGATION RESULT OF STENT PARTIALLY DEPLOYED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13361560 ↗MalfunctionJan 13, 2022Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON DECEMBER 21, 2021 THAT AN ULTRAFLEX ESOPHAGEAL COVERED PROXIMAL RELEASE STENT WAS TO BE IMPLANTED TO TREAT A 5-6 CM MALIGNANT STRICTURE DUE TO ESOPHAGEAL CANCER DURING AN ESOPHAGEAL STENTING PROCEDURE PERFORMED ON (B)(6) 2021. THE PATIENT'S ANATOMY WAS TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THE STENT COULD NOT BE DEPLOYED. THE STENT WAS REMOVED PARTIALLY DEPLOYED ON THE DELIVERY SYSTEM AND ANOTHER ULTRAFLEX ESOPHAGEAL STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13250980 ↗MalfunctionSep 13, 2021Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON AUGUST 20, 2021 THAT AN ULTRAFLEX ESOPHAGEAL NG COVERED PROXIMAL RELEASE STENT WAS TO BE USED TO TREAT A 4-5 CM STRICTURE DUE TO TUMORS IN THE ESOPHAGUS DURING AN ESOPHAGEAL STENT IMPLANTATION PROCEDURE PERFORMED ON (B)(6) 2021. THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THE STENT PARTIALLY DEPLOYED. THE STENT WAS REMOVED PARTIALLY DEPLOYED ON THE DELIVERY SYSTEM AND ANOTHER ULTRAFLEX ESOPHAGEAL STENT WAS IMPLANTED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12459669 ↗MalfunctionSep 1, 2021Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2021 THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT WAS TO BE IMPLANTED TO TREAT AN ESOPHAGEAL CANCER DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6), 2021. DURING THE PROCEDURE, WHEN THE STENT WAS ATTEMPTED TO BE DEPLOYED, RESISTANCE WAS FELT AND THE STENT WAS UNABLE TO DEPLOY. THE STENT WAS REMOVED FROM THE PATIENT PARTIALLY DEPLOYED ON THE DELIVERY SYSTEM AND THE PROCEDURE WAS CANCELLED AND RESCHEDULED TO USE A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12408938 ↗MalfunctionJun 23, 2020Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2020 THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT WAS TO BE IMPLANTED TO TREAT CANCER IN THE LOWER ESOPHAGUS DURING AN UPPER ENDOSCOPY WITH ESOPHAGEAL STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, POST PATIENT SEDATION, IT WAS NOTICED THAT THE STENT WAS PARTIALLY DEPLOYED WITHOUT MANIPULATION. REPORTEDLY, THERE WERE NO ISSUES NOTED WITH THE PACKAGING. THE PROCEDURE WAS COMPLETED WITH ANOTHER ULTRAFLEX ESOPHAGEAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. A PHOTO OF THE DEVICE PROVIDED BY THE COMPLAINANT BEFORE THE PROCEDURE SHOWS THE STENT WAS PARTIALLY DEPLOYED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10184094 ↗MalfunctionMar 26, 2019Activation, Positioning or Separation Problem; Material Deformation; Activation, Positioning or Separation Problem; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON MARCH 2, 2019 THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT WAS TO BE USED TO TREAT A MALIGNANT STRICTURE IN THE ESOPHAGUS DURING AN ESOPHAGEAL STENTING PROCEDURE PERFORMED ON (B)(6) 2019. REPORTEDLY, THE PATIENT'S ANATOMY WAS NOT DILATED PRIOR TO STENT PLACEMENT. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE STENT WAS UNABLE TO BE FULLY DEPLOYED AS THE DEPLOYMENT SUTURE WAS KNOTTED. THE PARTIALLY DEPLOYED STENT WAS REMOVED AND ANOTHER ULTRAFLEX ESOPHAGEAL STENT WAS PLACED TO COMPLETE THE PROCEDURE. THERE WERE NO WERE PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8453829 ↗MalfunctionOct 25, 2018Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2018 THAT AN ULTRAFLEX ESOPHAGEAL PROXIMAL RELEASE COVERED STENT WAS TO BE USED TO TREAT A MALIGNANT STRICTURE IN THE ESOPHAGUS DURING ESOPHAGEAL STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2018. REPORTEDLY, THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE THE STENT WAS ABLE TO CROSS THE STENOSIS ALONG WITH THE GUIDEWIRE. HOWEVER, IT WAS REPORTED THAT GREAT RESISTANCE WAS MET DURING DEPLOYMENT, AND THE STENT WAS DEPLOYED PARTIALLY ON THE DELIVERY SYSTEM. THE STENT WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER ULTRAFLEX ESOPHAGEAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8005618 ↗InjuryOct 15, 2018Migration; Migration
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2018 THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT WAS IMPLANTED TO TREAT A MALIGNANT STRICTURE IN THE DISTAL ESOPHAGUS DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2018. REPORTEDLY, THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. ACCORDING TO THE COMPLAINANT, 50 DAYS POST STENT PLACEMENT, THE PATIENT EXPERIENCED DYSPHAGIA. DURING A GASTROSCOPY PROCEDURE, THE PHYSICIAN NOTED THAT THE STENT COVER WAS PARTIALLY DETACHED FROM THE STENT. THE DETACHED COVER WAS REMOVED WITH A BIOPSY FORCEPS. REPORTEDLY, THE STENT REMAINS IMPLANTED AND THE PHYSICIAN WAS COMFORTABLE LEAVING THE STENT IN PLACE. THE PATIENT CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE. THE PATIENT'S DYSPHAGIA WAS NOTED TO HAVE RESOLVED.
Patient problems: Dysphagia/ Odynophagia
MAUDE report 7965757 ↗InjuryApr 24, 2018Difficult to Remove; Unraveled Material; Material Integrity Problem; Difficult to Remove; Unraveled Material; Material Integrity Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2018 THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT HAD BEEN IMPLANTED SIX WEEKS PRIOR TO (B)(6) 2018 TO TREAT AN IATROGENIC PERFORATION IN THE ESOPHAGUS CAUSED DURING A PREVIOUS ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE. ACCORDING TO THE COMPLAINANT, SIX WEEKS POST STENT PLACEMENT ON (B)(6) 2018, THE PATIENT EXPERIENCED DISCOMFORT WHICH PROMPTED THE PHYSICIAN TO CHECK THE ULTRAFLEX ESOPHAGEAL STENT. DURING THIS CHECK OF THE STENT, A FISTULA WAS NOTED AT THE PROXIMAL END OF THE STENT. ON (B)(6) 2018, A STENT REMOVAL PROCEDURE WAS PERFORMED. DURING THE PROCEDURE, THERE WAS DIFFICULTY REMOVING THE STENT AND IT BECAME UNRAVELED. THE STENT COVER WAS ALSO NOTED TO BE DAMAGED AT ONE END. THE STENT WAS ABLE TO BE REMOVED AND THE PATIENT UNDERWENT AN ESOPHAGOGRAM TEST. PER THE PHYSICIAN, THE TEST RESULTS WERE GOOD AND REVEALED NO ABNORMALITIES. IT IS UNKNOWN HOW THE FISTULA WAS TREATED BUT PER THE PHYSICIAN, EVERYTHING HAS HEALED AND THE PATIENT DOES NOT HAVE ANY OF THE PAST SYMPTOMS OR FISTULA GROWTH. THE PHYSICIAN PERFORMED THE FIRST DILATION ON (B)(6) 2018. NOTE: ACCORDING TO THE COMPLAINANT
Patient problems: Fistula
MAUDE report 7454900 ↗
Other PROSTHESIS, ESOPHAGEAL devices
- WallFlex™ Esophageal M00516730Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall29 reports
- WallFlex™ Esophageal M00516740Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall1 death22 reports
- WallFlex™ Esophageal M00516750Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall1 death19 reports
- WallFlex™ Esophageal M00516700Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall18 reports
- WallFlex™ Esophageal M00516910Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall15 reports
- WallFlex™ Esophageal M00516930Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall14 reports
- WallFlex™ Esophageal M00516710Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall13 reports
- WallFlex™ Esophageal M00516940Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall12 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page