The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 08714729716082), received between Nov 23, 2016 and Jan 12, 2024 — classified as 6 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation, Positioning or Separation Problem5
- Positioning Problem3
- Activation Failure2
- Break2
- Failure to Advance2
- Material Twisted/Bent2
- Component Missing1
- Difficult or Delayed Positioning1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient4
- Dysphagia/ Odynophagia1
- Fistula1
- Insufficient Information1
- No Code Available1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
InjuryJan 12, 2024Migration; Migration
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ULTRAFLEX ESOPHAGEAL PROXIMAL RELEASE COVERED STENT WAS IMPLANTED IN THE ESOPHAGUS FOR THE TREATMENT OF A 2CM STENOSIS DUE TO ESOPHAGEAL CANCER DURING A GASTROSCOPY PROCEDURE PERFORMED ON (B)(6) 2023. THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. ON SEPTEMBER 7, 2023, THE PATIENT EXPERIENCED DISTURBED PASSAGE OF SOLID FOOD AND IT WAS LATER NOTED THAT THE STENT MIGRATED. ON (B)(6) 2023, THE STENT WAS REMOVED USING BIOPSY FORCEPS. ON (B)(6) 2023, ANOTHER ULTRAFLEX ESOPHAGEAL STENT WAS IMPLANTED.
Patient problems: Dysphagia/ Odynophagia
MAUDE report 18506871 ↗InjuryMay 18, 2022Difficult or Delayed Positioning; Positioning Problem; Difficult or Delayed Positioning; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2022 THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT WAS USED TO TREAT AN ESOPHAGEAL STRICTURE DURING AN OESOPHAGO-GASTRO-DUODENOSCOPY (OGD) WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2022. REPORTEDLY, THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THE STENT DEPLOYMENT SUTURE COULD NOT BE PULLED OUT AND IT WAS NOTICED THROUGH IMAGING THAT THE STENT WAS DEPLOYED AT THE INCORRECT LOCATION. THE STENT WAS REMOVED FROM THE PATIENT WITH FORCEPS AND THE PROCEDURE WAS COMPLETED WITH ANOTHER ULTRAFLEX ESOPHAGEAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14428369 ↗InjuryDec 8, 2021Failure to Advance; Material Twisted/Bent; Positioning Problem; Activation Failure; Failure to Advance; Material Twisted/Bent; Positioning Problem; Activation Failure
THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT # 3005099803-2021-06336 AND 3005099803-2021-06337 FOR THE ASSOCIATED DEVICE INFORMATION. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2021 THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT WAS TO BE IMPLANTED TO TREAT A 12 CM MALIGNANT ESOPHAGEAL STENOSIS DURING A GASTROSCOPY PROCEDURE PERFORMED ON (B)(6) 2021. REPORTEDLY, THE PATIENT'S ANATOMY WAS TIGHT AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THERE WAS DIFFICULTY ADVANCING THE DELIVERY SYSTEM OF A 15CM ULTRAFLEX ESOPHAGEAL STENT (SUBJECT OF THIS REPORT) THROUGH THE STRICTURE. THE DISTAL PART OF THE DELIVERY SYSTEM WAS BENT AND THE STENT WAS DEPLOYED IN AN INCORRECT LOCATION. IT WAS OBSERVED UNDER DIRECT VISUALIZATION WITH A PEDIATRIC GASTROSCOPE THAT THE STENT DID NOT FULLY EXPAND. THE STENT WAS REMOVED FROM THE PATIENT USING A RE-USABLE GRASPING FORCEPS. A 10 CM ULTRAFLEX ESOPHAGEAL STENT (THE SUBJECT OF MFR. REPORT # 3005099803-2021-06337) WAS ATTEMPTED TO BE DEPLOYED, BUT SIMILAR PROBLEMS OCCURRED; THE DELIVERY SYSTEM HAD DIFFICULTY ADVANCING THROUGH THE STRICTURE, THE DI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12960100 ↗MalfunctionDec 8, 2021Failure to Advance; Material Twisted/Bent; Positioning Problem; Activation Failure; Failure to Advance; Material Twisted/Bent; Positioning Problem; Activation Failure
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2021 THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT WAS TO BE IMPLANTED TO TREAT A MALIGNANT ESOPHAGEAL STENOSIS DURING A GASTROSCOPY PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THERE WAS DIFFICULTY ADVANCING THE DELIVERY SYSTEM THROUGH THE STRICTURE. THE DISTAL PART OF THE DELIVERY SYSTEM WAS BENT AND THE STENT WAS DEPLOYED IN AN INCORRECT LOCATION. IT WAS OBSERVED UNDER DIRECT VISUALIZATION WITH A PEDIATRIC GASTROSCOPE THAT THE STENT DID NOT FULLY EXPAND. THE STENT WAS REMOVED FROM THE PATIENT USING A RE-USABLE GRASPING FORCEPS AND THE PROCEDURE WAS CANCELLED. A NON-SERIOUS INJURY WAS REPORTED RELATED TO THE EVENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 12960099 ↗MalfunctionDec 7, 2021Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2021 THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT WAS TO BE IMPLANTED TO TREAT A 6-7CM MALIGNANT STRICTURE DUE TO ESOPHAGEAL CANCER DURING AN ESOPHAGEAL STENT IMPLANTATION PROCEDURE PERFORMED ON (B)(6) 2021. REPORTEDLY, THE PATIENT'S ANATOMY WAS TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THE STENT COULD NOT BE DEPLOYED. THE STENT WAS REMOVED FROM THE PATIENT PARTIALLY COVERED BY THE STENT DEPLOYMENT SUTURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER ULTRAFLEX ESOPHAGEAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12951700 ↗MalfunctionNov 17, 2021Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON OCTOBER 29, 2021 THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT WAS TO BE IMPLANTED TO TREAT A 6-7CM MALIGNANT STRICTURE DUE TO ESOPHAGEAL CANCER DURING AN ESOPHAGEAL STENT IMPLANTATION PROCEDURE PERFORMED ON (B)(6) 2021. REPORTEDLY, THE PATIENT'S ANATOMY WAS TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THE STENT COULD NOT BE DEPLOYED. REPORTEDLY, THE STENT WAS REMOVED FROM THE PATIENT PARTIALLY COVERED BY THE STENT DEPLOYMENT SUTURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER ULTRAFLEX ESOPHAGEAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12832709 ↗MalfunctionOct 12, 2020Break; Activation, Positioning or Separation Problem; Break; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2020 THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT WAS TO BE IMPLANTED TO TREAT A MALIGNANT STRICTURE IN THE ESOPHAGUS DURING A STENTING PROCEDURE PERFORMED ON (B)(6) 2020. REPORTEDLY, THE PATIENT'S ANATOMY WAS DILATED PRIOR TO STENT PLACEMENT. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE PHYSICIAN ATTEMPTED TO DEPLOY THE STENT BY PULLING THE DEPLOYMENT SUTURE; HOWEVER, THE DEPLOYMENT SUTURE BROKE AND THE STENT FAILED TO DEPLOY. REPORTEDLY, THE STENT WAS REMOVED PARTIALLY DEPLOYED ON THE DELIVERY SYSTEM AND ANOTHER ULTRAFLEX ESOPHAGEAL STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10666611 ↗MalfunctionJul 7, 2020Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2020 THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT WAS TO BE IMPLANTED IN THE ESOPHAGUS TO TREAT ESOPHAGEAL CANCER DURING AN ESOPHAGEAL STENTING PROCEDURE PERFORMED ON (B)(6) 2020. REPORTEDLY, THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS DILATED PRIOR TO STENT PLACEMENT. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BLACK STENT DEPLOYMENT SUTURE COULD NOT BE COMPLETELY RELEASED AND THE STENT DID NOT FULLY DEPLOY. THE STENT WAS REMOVED FROM THE PATIENT PARTIALLY DEPLOYED ON THE DELIVERY SYSTEM AND ANOTHER ULTRAFLEX ESOPHAGEAL STENT WAS IMPLANTED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10239011 ↗InjuryMar 30, 2020Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ULTRAFLEX ESOPHAGEAL NG STENT WAS IMPLANTED IN THE ESOPHAGUS DURING A STENTING PROCEDURE PERFORMED ON (B)(6), 2020. ACCORDING TO THE COMPLAINANT, POST STENT PLACEMENT, ON (B)(6), 2020 THE STENT WAS FOUND TO BE SLIGHTLY BROKEN IN THE MIDDLE. ANOTHER ULTRAFLEX ESOPHAGEAL NG STENT WAS PLACED STENT-IN-STENT TO COVER THE BROKEN PORTION OF THE STENT TO RESOLVE THE ISSUE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. ADDITIONAL INFORMATION RECEIVED ON (B)(6), 2020 REPORTEDLY, THE PATIENT DID NOT EXPERIENCE SYMPTOMS OR ADVERSE EVENTS RELATED TO THE STENT BREAK CAUSING THE PATIENT TO BE HOSPITALIZED. THE PATIENT WAS ALREADY HOSPITALIZED WHEN THE PHYSICIAN DISCOVERED A LEAK IN THE STENT.
Patient problems: No Code Available
MAUDE report 9898489 ↗InjuryDec 14, 2017Component Missing; Component Missing
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2017 THAT AN ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE COVERED STENT WAS IMPLANTED TO TREAT AN ESOPHAGEAL ANASTOMOTIC DEHISCENCE DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2017. REPORTEDLY, THE PATIENT¿S ANATOMY WAS NOT DILATED PRIOR TO STENT PLACEMENT AND THE PATIENT UNDERWENT CHEMOTHERAPY AND RADIATION PRIOR TO THE INITIAL SURGERY. ACCORDING TO THE COMPLAINANT, ON (B)(6) 2017, 8 DAYS POST STENT PLACEMENT, BILE WAS COMING OUT THROUGH THE PATIENT¿S INTUBATION TUBE, WHICH PROMPTED THE PHYSICIAN TO CHECK THE STENT. THE PHYSICIAN PERFORMED AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE AND FOUND THAT A PART OF THE STENT COVER HAD DISAPPEARED. ACCORDING TO THE PHYSICIAN, THE STENT HAD FAILED TO RESOLVE THE DEHISCENCE AND A 4 CM BRONCHO-ESOPHAGIC FISTULA WAS NOTED. THE PATIENT WAS SENT TO URGENT SURGERY AND THE PHYSICIAN PERFORMED CERVICAL ESOPHAGECTOMY, GASTROSTOMY, AND MUSCLE PLASTY OVER THE BRONCHO-ESOPHAGIC FISTULAE. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: Fistula
MAUDE report 7119103 ↗InjuryAug 4, 2017Kinked; Kinked
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON JULY 16, 2017 THAT AN ULTRAFLEX ESOPHAGEAL PROXIMAL RELEASE COVERED STENT WAS TO BE USED TO TREAT A 12 CM STRICTURE IN THE MID-ESOPHAGUS DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2017. REPORTEDLY, THE PATIENT'S ANATOMY HAD BEEN DILATED UP TO 11 MM PRIOR TO STENT PLACEMENT. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE PHYSICIAN WAS ABLE TO DEPLOY THE STENT. HOWEVER, WHEN THE STENT WAS VISUALIZED WITH A SCOPE, THE STENT WAS NOTED TO BE BENT AND THE ESOPHAGUS WAS COMPLETELY BLOCKED. THE STENT WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER ULTRAFLEX ESOPHAGEAL STENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6768581 ↗MalfunctionNov 23, 2016Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2016 THAT AN ULTRAFLEX ESOPHAGEAL PROXIMAL RELEASE STENT WAS TO BE USED IN THE ESOPHAGUS DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2016. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE STENT WAS ATTEMPTED TO BE DEPLOYED; HOWEVER, ONCE THE STENT WAS ALMOST FULLY DEPLOYED (75% DEPLOYED) THE STENT WAS NO LONGER ABLE TO BE RELEASED. THE PHYSICIAN SUSPECTED THAT THE DEPLOYMENT SUTURE WAS ENTANGLED. THE PARTIALLY DEPLOYED STENT WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER ULTRAFLEX ESOPHAGEAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT¿S CONDITION POST- PROCEDURE WAS REPORTED TO BE ¿NO PROBLEM¿.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6125387 ↗
Other PROSTHESIS, ESOPHAGEAL devices
- WallFlex™ Esophageal M00516730Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall29 reports
- WallFlex™ Esophageal M00516740Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall1 death22 reports
- WallFlex™ Esophageal M00516750Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall1 death19 reports
- WallFlex™ Esophageal M00516700Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall18 reports
- WallFlex™ Esophageal M00516910Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall15 reports
- WallFlex™ Esophageal M00516930Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall14 reports
- WallFlex™ Esophageal M00516710Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall13 reports
- WallFlex™ Esophageal M00516940Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall12 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page