Ultraflex™ Esophageal NG M00513830Recalls & reported problems
Boston Scientific Corporation · Proximal Release Stent System · Class 2 · Implant
All models of Ultraflex Esophageal NG (15)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729716204), received between Jul 13, 2021 and Feb 25, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Activation, Positioning or Separation Problem2
- Appropriate Device Problem Term/Code Not Available1
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 25, 2022Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
THIS ULTRAFLEX ESOPHAGEAL NG PROXIMAL RELEASE STENT WAS RETURNED TO BOSTON SCIENTIFIC CORPORATION WITHOUT ANY REPORT OF PERFORMANCE ISSUES OR ADVERSE PATIENT EFFECTS. THE RETURNED DEVICE WAS ANALYZED AND THE STENT WAS PARTIALLY DEPLOYED AND THE SHAFT WAS SLIGHTLY BENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF STENT PARTIALLY DEPLOYED. NOTE: BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13599261 ↗MalfunctionJul 13, 2021Positioning Failure; Activation, Positioning or Separation Problem; Appropriate Device Problem Term/Code Not Available; Positioning Failure; Activation, Positioning or Separation Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2021 THAT AN ULTRAFLEX ESOPHAGEAL NG UNCOVERED PROXIMAL RELEASE STENT WAS TO BE IMPLANTED IN THE ESOPHAGUS DURING A STENT PLACEMENT PROCEDURE FOR POST-RADIATION ESOPHAGUS SQUAMOUS CELL CARCINOMA PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE STENT COULD NOT BE DEPLOYED. THE PROCEDURE WAS NOT COMPLETED AS ANOTHER STENT OF THE SAME SIZE WAS UNAVAILABLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JULY 13, 2021. THE ULTRAFLEX ESOPHAGEAL NG UNCOVERED PROXIMAL RELEASE STENT WAS PARTIALLY COVERED BY THE BLACK STENT DEPLOYMENT SUTURE WHEN IT WAS REMOVED FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12159337 ↗
FDA clearance and classification
The FDA regulates product code ESW (prosthesis, esophageal) under 21 CFR 878.3610 as a Class II device. 21 CFR 878.3610 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Ultraflex™ Esophageal NG M00513830
Has the Ultraflex™ Esophageal NG M00513830 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08714729716204), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Ultraflex™ Esophageal NG M00513830?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jul 13, 2021 and Feb 25, 2022. In those reports, the most common specific device problems are “Activation, Positioning or Separation Problem” (2) and “Positioning Failure” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Ultraflex™ Esophageal NG M00513830?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Ultraflex™ Esophageal NG M00513830 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Ultraflex™ Esophageal NG M00513830, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Ultraflex™ Esophageal NG, model M00513830). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other PROSTHESIS, ESOPHAGEAL devices
- WallFlex™ Esophageal M00516730Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall29 reports
- WallFlex™ Esophageal M00516740Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall1 death22 reports
- WallFlex™ Esophageal M00516750Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall1 death19 reports
- WallFlex™ Esophageal M00516700Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall18 reports
- WallFlex™ Esophageal M00516910Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall15 reports
- WallFlex™ Esophageal M00516710Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall13 reports
- WallFlex™ Esophageal M00516930Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall13 reports
- WallFlex™ Esophageal M00516940Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall12 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page