RIGIFLEX II M00554500Recalls & reported problems
Boston Scientific Corporation · Single Use Achalasia Ballon Dilator · Class 2
All models of RIGIFLEX II (3)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 08714729719106), received between Dec 9, 2019 and Jan 24, 2025 — classified as 1 injury, 16 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Burst Container or Vessel15
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
- No Consequences Or Impact To Patient3
- Perforation1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 24, 2025Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING A LOWER ESOPHAGEAL SPHINCTER (LES) DILATATION PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE BALLOON BURST. IT WAS REPORTED THAT NO PIECE OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER RIGIFLEX II DILATATION BALLOON. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21233639 ↗MalfunctionJan 10, 2025Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II DILATATION BALLOON WAS USED DURING AN ENDOSCOPIC DILATATION PROCEDURE. THE EXACT PROCEDURE DATE IS UNKNOWN. DURING THE PROCEDURE, THE BALLOON BURST. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21127686 ↗MalfunctionJul 25, 2023Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II BALLOON WAS USED IN THE LOWER ESOPHAGEAL SPHINCTER (LES) DURING A DILATATION PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE BALLOON BURST. THE CUSTOMER STATED THAT NO PIECES OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH THE ORIGINAL DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE OKAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17393704 ↗MalfunctionJan 25, 2023Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II DILATATION BALLOON WAS USED IN THE ESOPHAGUS DURING AN ESOPHAGEAL DILATION PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE FOR TEST PERIOD, THE STAFF OPENED THE PACKAGED AND NOTICED THAT THE CATHETER WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE FIXED WIRE DILATATION BALLOON DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16243705 ↗MalfunctionDec 16, 2022Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II DILATATION BALLOON WAS USED IN THE LOWER ESOPHAGEAL SPHINCTER DURING AN ENDOSCOPIC DILATION PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, THE BALLOON RUPTURED AT 20 PSI AFTER ONLY ONE MINUTE OF DILATING. THE CUSTOMER STATED THAT NO PIECES OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE HAD ALREADY BEEN COMPLETED SUCCESSFULLY WITH THE ORIGINAL DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15998007 ↗MalfunctionNov 21, 2022Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II BALLOON WAS USED IN A PROCEDURE ON (B)(6) 2022. DURING THE PROCEDURE, THE BALLOON BURST. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15840288 ↗MalfunctionAug 16, 2022Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II DILATATION BALLOON WAS USED IN THE ESOPHAGUS DURING AN ACHALASIA DILATATION PROCEDURE PERFORMED ON (B)(6) 2022. DURING PREPARATION, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER RIGIFLEX II DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION WAS REPORTED TO BE FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15235315 ↗InjuryJun 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II DILATATION BALLOON WAS USED IN THE ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2022. AFTER THE PROCEDURE, WHEN THE PHYSICIAN LOOKED IN THE ESOPHAGUS POST DILATION, PERFORATION WAS NOTICED. ENDOSCOPIC CLOSURE WAS ATTEMPTED BUT A SURGICAL INTERVENTION WAS NEEDED. SURGERY WAS SUCCESSFUL; HOWEVER, THE PATIENT WAS ADMITTED TO THE HOSPITAL BEYOND THE STANDARD OF CARE. THERE WAS NO ALLEGED PROBLEM WITH THE BALLOON. THE PATIENT CONDITION FOLLOWING THE INTERVENTION WAS REPORTED TO BE STABLE.
Patient problems: Perforation
MAUDE report 14738865 ↗MalfunctionJan 5, 2022Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II BALLOON WAS PREPARED FOR USE IN AN ACHALASIA BALLOONING PROCEDURE PERFORMED ON (B)(6) 2021. DURING PREPARATION, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER RIGIFLEX II BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13171629 ↗MalfunctionDec 23, 2021Burst Container or Vessel; Use of Device Problem; Burst Container or Vessel; Use of Device Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II DILATATION BALLOON WAS USED IN THE ESOPHAGUS DURING AN ACHALASIA BALLOONING PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER RIGIFLEX II DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13082036 ↗MalfunctionAug 13, 2021Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II DILATATION BALLOON WAS USED IN THE COMMON BILE DUCT (CBD) DURING A BALLOON DILATATION PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER RIGIFLEX II DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12321605 ↗MalfunctionAug 13, 2021Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II DILATATION BALLOON WAS USED IN THE COMMON BILE DUCT (CBD) DURING A BALLOON DILATATION PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER RIGIFLEX II DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12321181 ↗MalfunctionMay 18, 2021Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II DILATATION BALLOON WAS UNPACKED TO BE USED IN THE ESOPHAGUS FOR AN ACHALASIA DILATATION PROCEDURE PERFORMED ON (B)(6) 2021. DURING PREPARATION, IT WAS NOTED THAT THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER RIGIFLEX II DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11843357 ↗MalfunctionMar 11, 2021Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II DILATATION BALLOON WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ACHALASIA PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER RIGIFLEX II DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11463652 ↗MalfunctionAug 7, 2020Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II DILATATION BALLOON WAS USED DURING AN ACHALASIA BALLOONING PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, IT WAS NOTED THAT THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER RIGIFLEX II DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10379013 ↗MalfunctionMay 5, 2020Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II DILATATION BALLOON WAS USED IN THE MID COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED ANOTHER RIGIFLEX II DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10024429 ↗MalfunctionDec 9, 2019Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RIGIFLEX II DILATATION BALLOON WAS USED IN THE LOWER ESOPHAGUS AND CARDIA DURING A DILATION PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON SUDDENLY BURST WHEN ATTEMPTED TO BE INFLATED. THE PROCEDURE WAS NOT COMPLETED AS THE SAME DEVICE WAS UNAVAILABLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE OKAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9440340 ↗
Other esophageal dilator balloon with or without electrode sensors devices
- EsoFLIP ES-330Covidien LP · esophageal dilator balloon with or without electrode sensors1 recall91 reports
- RIGIFLEX II M00554510Boston Scientific Corporation · esophageal dilator balloon with or without electrode sensors7 reports
- RIGIFLEX II M00554520Boston Scientific Corporation · esophageal dilator balloon with or without electrode sensors2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page