Sterling™ H74939032806080Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Sterling (153)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729720713), received between Sep 23, 2019 and Sep 23, 2019 — classified as 1 injury.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Material Integrity Problem1
- Material Rupture1
Patient problems
- No Code Available1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 23, 2019Material Rupture; Detachment of Device or Device Component; Material Integrity Problem; Material Rupture; Detachment of Device or Device Component; Material Integrity Problem
IT WAS REPORTED THAT THE TIP DETACHED. AN 8.0MMX60MMX80CM STERLING BALLOON WAS SELECTED FOR A SHUNT PROCEDURE IN THE UPPER LEG. DURING THE PROCEDURE, THE BALLOON WAS DAMAGED DURING DILATATION AND THE TIP OF THE DEVICE DETACHED. THERE WERE NO REPORTED PATIENT COMPLICATIONS. IT WAS FURTHER REPORTED THAT WHEN DILATING THE BALLOON, IT SUDDENLY BURST. THE BALLOON PRESSURE WAS 13 MILLIBARS WHEN IT BURST. RADIOGRAPH SHOWED THAT A TORN PART OF THE BALLOON REMAINED IN THE VESSEL. THE PHYSICIAN WAS ABLE TO RETRIEVE THE DETACHED PIECE OF THE BALLOON WITH A SNARE CATHETER. AFTER REMOVING THE BALLOON, IT WAS NOTED THAT THE BALLOON HAD BEEN TORN CIRCULARLY 5MM IN FRONT OF THE TIP. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS DISCHARGED AFTER THE PROCEDURE.
Patient problems: No Code Available; No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 9105401 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page