Polaris™ Ultra M0061921530Recalls & reported problems
Boston Scientific Corporation · Ureteral Stent · Class 2 · Implant
All models of Polaris Ultra (40)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729720942), received between Jun 10, 2021 and Feb 11, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Difficult to Advance” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Advance3
- Deformation Due to Compressive Stress2
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 11, 2025Deformation Due to Compressive Stress; Difficult to Advance; Adverse Event Without Identified Device or Use Problem; Deformation Due to Compressive Stress; Difficult to Advance; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING PROCEDURE, WHEN THE STENT WAS PUSHED UP USING THE POSITIONER, THE TAIL WAS BUCKLED. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE AND THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21358503 ↗MalfunctionOct 20, 2023Deformation Due to Compressive Stress; Difficult to Advance; Material Split, Cut or Torn; Deformation Due to Compressive Stress; Difficult to Advance; Material Split, Cut or Torn
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS ULTRA URETERAL STENT WAS USED DURING A URETEROSCOPY PROCEDURE IN THE RIGHT URETER PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE STENT WAS PUSHED UP AS USING POSITIONER, THE PIGTAIL WAS BUCKLED, AND THEY FAILED TO INSERT IT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER POLARIS ULTRA URETERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. A PHOTO OF THE COMPLAINT DEVICE WAS PROVIDED AND SHOWED THAT THE STENT WAS BUCKLED AND TORN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17980719 ↗MalfunctionJun 10, 2021Break; Difficult to Advance; Break; Difficult to Advance
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS ULTRA URETERAL STENT WAS USED DURING A URETERAL STENTING PROCEDURE IN THE RIGHT URETER, PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE AND INSIDE THE PATIENT, WHEN THE PHYSICIAN ATTEMPTED TO INSERT THE STENT, RESISTANCE WAS FELT AND THERE WAS DIFFICULTY IN ADVANCING STENT. ANOTHER POLARIS ULTRA URETERAL STENT USED TO SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF STENT SHAFT BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11979468 ↗
Other STENT, URETERAL devices
- Bard® InLay Optima® Ureteral Stent 788426C. R. Bard, Inc. · STENT, URETERAL1 recall100 reports
- Bard® InLay Optima® Ureteral Stent 788626C. R. Bard, Inc. · STENT, URETERAL122 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777626C. R. Bard, Inc. · STENT, URETERAL118 reports
- Bard® InLay® Ureteral Stent with HydroGlide™ Guidewire 777426C. R. Bard, Inc. · STENT, URETERAL108 reports
- Bard® InLay® Ureteral Stent 778626C. R. Bard, Inc. · STENT, URETERAL100 reports
- Percuflex™ Plus M0061752530Boston Scientific Corporation · STENT, URETERAL1 death99 reports
- Contour VL™ M0061801560Boston Scientific Corporation · STENT, URETERAL1 recall45 reports
- Polaris™ Ultra M0061921330Boston Scientific Corporation · STENT, URETERAL71 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page