Navigator M0062502030Recalls & reported problems
Boston Scientific Corporation · Ureteral Access Sheath Set · Class 2
All models of Navigator (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729751502), received between Nov 22, 2017 and Sep 21, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Peeled/Delaminated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Peeled/Delaminated2
- Material Deformation1
- Material Fragmentation1
- Material Split, Cut or Torn1
- Torn Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2026Peeled/Delaminated; Material Deformation
It was reported that a Navigator Access Sheath device was set to be used during a Lithotripsy procedure. Reportedly, upon opening it was found that the sheath crumbled and began to chip before use. Additionally, the sheath was also noticed to be kinked. The procedure was completed with another of the same device with no patient complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50842276 ↗MalfunctionAug 27, 2026Material Fragmentation
Sheath crumbled and began to chip before the item was ever used.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50482721 ↗InjuryDec 18, 2025Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED THAT A NAVIGATOR ACCESS SHEATH DEVICE WAS USE DURING A PROCEDURE. DURING STONE LASER TREATMENT IN THE KIDNEY'S UPPER POLE, THEY NOTICED BLUE FLECKS IN THE PATIENT'S BLADDER. REPORTEDLY, IT WAS FOUND THAT THE NAVIGATOR BREAKS APART AND THE PHYSICIAN RE-ENTERED THE URETER SIX TIMES TO ENSURE ALL PLASTIC PIECES WERE RETRIEVED. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23838091 ↗MalfunctionJun 18, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT A NAVIGATOR ACCESS SHEATH DEVICE WAS ABOUT TO BE USED DURING A FLEXIBLE URETEROSCOPY PROCEDURE. PRIOR TO PROCEDURE, THE PHYSICIAN PLACES THE ACCESS SHEATH OVER THE WIRE AND NOTICED THAT THE CONNECTION OF THE SHEATH TO THE HUB WAS FRAYED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19556751 ↗MalfunctionNov 22, 2017Torn Material; Torn Material
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NAVIGATOR ACCESS SHEATH WAS USED IN THE KIDNEY DURING A URETEROSCOPY AND LASER LITHOTRIPSY PROCEDURE PERFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE INNER LINING OF THE DEVICE STARTED TO PEEL OFF/SHEARED EXPOSING THE METAL COIL. IT WAS ALSO REPORTED THAT NO PARTS DETACHED FROM THE SHEATH. THE PROCEDURE WAS COMPLETED WITH ANOTHER NAVIGATOR ACCESS SHEATH. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7052236 ↗
Other CATHETER, UROLOGICAL devices
- Magic3™ Coude Intermittent Catheter with Sure-Grip™ Sleeve 50614C. R. Bard, Inc. · CATHETER, UROLOGICAL883 reports
- Magic3™ Intermittent Catheter with Sure-Grip™ Sleeve 53614GC. R. Bard, Inc. · CATHETER, UROLOGICAL692 reports
- Magic3™ Coude Intermittent Catheter with Sure-Grip™ Sleeve 50616C. R. Bard, Inc. · CATHETER, UROLOGICAL2 deaths443 reports
- GENTLECATH GLIDE 421567Convatec, Purchasing Department · Catheter, Urological342 reports
- Magic3™ Intermittent Catheter with Sure-Grip™ Sleeve 53616GC. R. Bard, Inc. · CATHETER, UROLOGICAL301 reports
- Bardex® Foley Catheter Silicone Coated 0167V22SC. R. Bard, Inc. · CATHETER, UROLOGICAL187 reports
- Bardex® Foley Catheter Silicone Coated 0167V20SC. R. Bard, Inc. · CATHETER, UROLOGICAL175 reports
- Bardia® Coude Olive Tip Urethral Foley Catheter 802516C. R. Bard, Inc. · CATHETER, UROLOGICAL170 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗