Ureteral Catheter M0064002110Recalls & reported problems
Boston Scientific Corporation · Cone Tip Ureteral Catheter · Class 2
All models of Ureteral Catheter (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729751663), received between Jun 9, 2017 and May 9, 2023 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break3
- Defective Device2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- Hemorrhage/Blood Loss/Bleeding1
- Internal Organ Perforation1
- Laceration(s)1
- Perforation1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 9, 2023Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA URETERAL CATHETER WAS USED IN THE URETER DURING A URETEROSCOPY PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE CATHETER WAS BEING REMOVED FROM THE SCOPE; HOWEVER, THE CATHETER BROKE. THE PHYSICIAN REPORTED THAT NO UNORDINARY RESISTANCE WAS FELT AND THERE WAS NO ANY MISUSE. THE PROCEDURE WAS COMPLETED WITH ANOTHER FLEXIMA URETERAL CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16895861 ↗MalfunctionDec 8, 2022Defective Device; Defective Device
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA 5F CONE TIP URETERAL CATHETER WAS USED IN THE KIDNEY DURING A URETEROSCOPY PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, WHEN THE DEVICE WAS IMPLANTED, IT WAS NOTICED THAT THE CONE TIP PORTION OF THE CATHETER WAS SHARP AND PIERCED THROUGH THE URETERAL ORIFICE. THE DEVICE WAS THEN REMOVED IMMEDIATELY, AND THE PROCEDURE WAS COMPLETED WITH AN OPEN-ENDED CATHETER. IT IS UNKNOWN HOW EXACTLY THE CATHETER PIERCED THE URETERAL ORIFICE. THERE WAS NO REPORTED TREATMENT PERFORMED FOR THE PATIENT INJURY. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS.
Patient problems: No Clinical Signs, Symptoms or Conditions; Perforation
MAUDE report 15941269 ↗InjuryJun 28, 2021Defective Device; Defective Device
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA UERETERAL CATHETER WAS USED IN THE URETERAL ORIFICE DURING A URETEROSCOPY PROCEDURE PERFORMED ON JUNE 3, 2021. DURING THE PROCEDURE, WHEN THE PHYSICIAN INSERTED THE DEVICE, HE NOTICED A LINEAR BURR ON THE CONE PORTION OF THE CATHETER WHICH CAUSED A CUT IN THE URETERAL ORIFICE OF THE PATIENT. REPORTEDLY, THE PATIENT EXPERIENCED BLEEDING AND THEY HAD TO PUT PRESSURE ON THE SITE TO STOP THE BLEEDING. AND THE PROCEDUE WAS COMPLETED AT THIS TIME. THE PATIENT WAS REPORTED TO HAVE FULLY RECOVERED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Laceration(s)
MAUDE report 12074907 ↗MalfunctionDec 15, 2020Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA URETERAL CATHETER DRAINAGE DEVICE WAS USED IN THE URETER DURING A URETEROSCOPY PROCEDURE PERFORMED ON (B)(6), 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THE CATHETER BROKE OFF AT THE END AND CREATED A BARBED SHARP TIP. REPORTEDLY, THE BROKEN CATHETER WAS REMOVED FROM THE SCOPE, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER FLEXIMA URETERAL CATHETER DRAINAGE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. **ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2020** REPORTEDLY, THE TIP BROKE OFF DURING UNPACKING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11010942 ↗InjuryAug 29, 2017Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA URETERAL CATHETER WAS USED IN THE BLADDER AND URETER DURING A CYSTOSCOPY PROCEDURE PERFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CATHETER BROKE INSIDE THE PATIENT BUT WERE SUCCESSFULLY RETRIEVED, THOUGH IT IS UNKNOWN HOW THE PIECES WERE RETRIEVED. NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AFTER THE PROCEDURE WAS REPORTED TO BE OKAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6827311 ↗InjuryJun 9, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A FLEXIMA 5F CONE TIP URETERAL CATHETER WAS USED DURING A URETEROSCOPY PROCEDURE PERFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE URETER WAS PERFORATED WHEN ADVANCING THE FLEXIMA 5F CONE TIP URETERAL CATHETER. THE CONE TIP WAS NOTED TO BE TOO SHARP. IT IS UNKNOWN IF THERE WAS ADDITIONAL INTERVENTION PERFORMED FOR THE PUNCTURED URETER AND HOW THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Internal Organ Perforation
MAUDE report 6629317 ↗
Other CATHETER, UROLOGICAL devices
- Magic3™ Coude Intermittent Catheter with Sure-Grip™ Sleeve 50614C. R. Bard, Inc. · CATHETER, UROLOGICAL883 reports
- Magic3™ Intermittent Catheter with Sure-Grip™ Sleeve 53614GC. R. Bard, Inc. · CATHETER, UROLOGICAL692 reports
- Magic3™ Coude Intermittent Catheter with Sure-Grip™ Sleeve 50616C. R. Bard, Inc. · CATHETER, UROLOGICAL2 deaths443 reports
- GENTLECATH GLIDE 421567Convatec, Purchasing Department · Catheter, Urological342 reports
- Magic3™ Intermittent Catheter with Sure-Grip™ Sleeve 53616GC. R. Bard, Inc. · CATHETER, UROLOGICAL301 reports
- Bardex® Foley Catheter Silicone Coated 0167V22SC. R. Bard, Inc. · CATHETER, UROLOGICAL187 reports
- Bardex® Foley Catheter Silicone Coated 0167V20SC. R. Bard, Inc. · CATHETER, UROLOGICAL175 reports
- Bardia® Coude Olive Tip Urethral Foley Catheter 802516C. R. Bard, Inc. · CATHETER, UROLOGICAL170 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page