Skip to content

Polaris™ Ultra M0061931320Recalls & reported problems

Boston Scientific Corporation · Ureteral Stent · Class 2 · Implant

All models of Polaris Ultra (40)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729754534), received between Mar 11, 2022 and Mar 11, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 11, 2022Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A POLARIS ULTRA URETERAL STENT WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2021. DURING PROCEDURE, IT WAS NOTED THAT THE SUTURE STRING BECAME BROKEN EVEN THOUGH IT WAS NOT INTENDED TO BE REMOVED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER POLARIS ULTRA URETERAL STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE AS NO PATIENT INJURY. INVESTIGATION RESULTS REVEALED THAT THE BLADDER PIG TAIL TORN, THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13733334 ↗

Other STENT, URETERAL devices

All STENT, URETERAL devices →