Skip to content

Sterling™ H74939031552010Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Sterling (153)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729755579), received between Jul 29, 2020 and Nov 10, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 10, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE RIGHT INTERNAL CAROTID ARTERIES. A 5.5MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, A BALLOON RUPTURED WAS NOTED DURING INFLATION CHECK AT THE TIME OF SETUP. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23518729 ↗
  • MalfunctionJul 29, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON PINHOLE OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND NON-CALCIFIED INTERNAL CAROTID TO COMMON CAROTID ARTERY. AFTER A FILTERWIRE EZ GUIDEWIRE CROSSED THE LESION, A 4-30 STERLNG BALLOON CATHETER WAS USED FOR PRE-DILATATION, FOLLOWED BY PLACEMENT OF A 10-31 CAROTID WALLSTENT. A 5.5MMX20MMX135CM STERLING BALLOON CATHETER WAS THEN ADVANCED FOR POST-DILATATION. HOWEVER, DURING INFLATION, LEAKAGE OF CONTRAST AGENT AND PINHOLE LEAK WAS OBSERVED. THE PROCEDURE WAS COMPLETED AS THE REQUIRED EXPANSION WAS OBTAINED. POST-PROCEDURE, VISUAL DILATATION WITH SALINE SOLUTION WAS PERFORMED BUT NO MAJOR LEAK OR IMAGE WAS OBSERVED IN THE BALLOON MATERIAL. NO KNOWN PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 10340868 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →