Sterling™ H74939031404010Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Sterling (153)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Sterling balloon dilatation catheter is used to open narrowed arteries, most often in the leg, forearm, or dialysis shunt. Most of the 68 reports describe the balloon bursting during inflation or while being pushed through a bent or narrowed vessel, with a few mentioning difficulty inflating or deflating it or trouble removing it. There are no deaths and 1 injury. No recalls name this device, and reports have slowed from 9 in the earlier 12 months to 3 in the most recent12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 68
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 68 adverse event reports that list this exact device (DI 08714729756996), received between Aug 17, 2018 and Apr 1, 2026 — classified as 1 injury, 67 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (62 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture62
- Deflation Problem3
- Detachment of Device or Device Component2
- Inflation Problem2
- Break1
- Contamination1
- Difficult to Remove1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions51
- No Consequences Or Impact To Patient17
Recent adverse event reports
The 25 most recent of 68 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 1, 2026Nonstandard Device; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SHUNT. A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, RESISTANCE WAS ENCOUNTERED DURING INSERTION, AND THE SHEATH WAS NOTED TO BE BENT. THE BALLOON WAS FORCEFULLY ADVANCED, AND RUPTURE OCCURRED WITHIN THE SHEATH PRIOR TO INFLATION. DAMAGE TO THE DISTAL TIP OF THE BALLOON WAS OBSERVED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT REMAINED STABLE POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24758869 ↗MalfunctionMar 30, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND MILDLY CALCIFIED VESSEL IN THE FOREARM. A 4.0MMX40MMX135CM (4F) STERLING BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE SECOND INFLATION AT 6 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM USING THE NORMAL METHOD, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24737549 ↗MalfunctionNov 5, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS SELECTED FOR USE IN AN ENDOVASCULAR TREATMENT. DURING THE FIRST INFLATION AT 6 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23480033 ↗MalfunctionAug 1, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 4.0MMX40MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE FIRST INFLATION AT 6 ATMOSPHERES. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22691775 ↗MalfunctionJul 30, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED RIGHT SUPERFICIAL FEMORAL ARTERY. A 4.0MMX40MMX135CM (4F) STERLING BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 8 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22666115 ↗MalfunctionJul 16, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION AT NOMINAL ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED AT THE MIDDLE PART OF THE BALLOON. THE DEVICE WAS REMOVED, AND A PINHOLE WAS NOTED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT INJURIES REPORTED, AND THE PATIENT CONDITION WAS GOOD POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22524630 ↗MalfunctionJul 11, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION AT 12 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT INJURIES REPORTED, AND THE PATIENT CONDITION WAS GOOD POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22482068 ↗MalfunctionJun 16, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING SECOND INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED IN A PINHOLE FORM AT NEAR THE TIP. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22210843 ↗MalfunctionJun 13, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THIRD INFLATION AT 6 ATMOSPHERES FOR SEVERAL SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22203451 ↗MalfunctionMay 12, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. A 4.0MM X 40MM X 135CM STERLING MR BALLOON CATHETER WAS ADVANCE FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 8 ATMOSPHERES FOR 30 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WAS NO PATIENT COMPLICATIONS NOTED AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22005862 ↗MalfunctionApr 3, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LEFT POPLITEAL ARTERY. A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING FIRST INFLATION AT 12 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED IN A PINHOLE FORM. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21761153 ↗MalfunctionJan 16, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY CALCIFIED VESSEL. A 4.0MMX40MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE SECOND INFLATION AT NOMINAL PRESSURE FOR 10 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21170009 ↗MalfunctionOct 11, 2024Contamination; Contamination
IT WAS REPORTED THAT CONTAMINATION OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED RIGHT SUPERFICIAL FEMORAL ARTERY. A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT A HAIR WAS FOUND IN THE PRODUCT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20429039 ↗MalfunctionJul 22, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED POPLITEAL ARTERY. A 4.0MMX40MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE SECOND INFLATION AT 14 ATMOSPHERES FOR 30 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19793307 ↗MalfunctionJun 20, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED POPLITEAL ARTERY. A 4.0MMX40MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 6 ATMOSPHERES FOR 3 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19577147 ↗MalfunctionJun 13, 2024Deflation Problem; Inflation Problem; Deflation Problem; Inflation Problem
IT WAS REPORTED THAT THE BALLOON WAS SLOW TO DEFLATE. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND CALCIFIED CAROTID ARTERY. A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION AND INFLATED TO 12 ATMOSPHERES FOR 1 MIN. HOWEVER, DURING THE PROCEDURE, THE BALLOON WAS TOO SLOW TO INFLATE AND DEFLATE. THE BALLOON WAS PULLED OUT AND REMOVED IN A DEFLATED STATE, AND THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19531738 ↗MalfunctionMay 15, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED COMMON FEMORAL ARTERY. A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 6 ATMOSPHERES FOR LESS THAN 10 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19323076 ↗MalfunctionMar 5, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 4.0MMX40MMX135CM (4F) STERLING BALLOON WAS ADVANCED FOR DILATATION. HOWEVER, UPON FIRST INFLATION AT 5 ATMOSPHERES FOR 2 SECONDS THE BALLOON RUPTURED NEAR THE MIDDLE IN PINHOLE FORM. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18835958 ↗MalfunctionJan 4, 2024Deflation Problem; Inflation Problem; Deflation Problem; Inflation Problem
IT WAS REPORTED THAT SLOW DEFLATION OCCURRED. THE PATIENT UNDERWENT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY OF THE TARGET LESION LOCATED IN THE CAROTID ARTERY. A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS USED FOR DILATATION. HOWEVER, DURING THE PROCEDURE, IT WAS NOTED THAT THE INFLATION AND DEFLATION OF THE BALLOON WAS TOO SLOW. THE PROCEDURE WAS COMPLETED WITH THE ORIGINAL DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18446317 ↗MalfunctionJan 2, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN SUPERFICIAL FEMORAL ARTERY. A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18430676 ↗MalfunctionAug 7, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED VESSEL. A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION AT THE RATED BURST PRESSURE, THE BALLOON BURST. THE DEVICE WAS RETRIEVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17474263 ↗MalfunctionJul 5, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY BELOW THE KNEE. A 4.0MM X 40MM X 135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 14 ATMOSPHERES FOR ABOUT 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17262371 ↗MalfunctionJun 29, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY BELOW THE KNEE. A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS STABLE POST- PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17230373 ↗MalfunctionJun 22, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY (SFA). A 4.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 8 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH DIFFERENT DEVICE WITH NO PATIENT COMPLICATIONS WERE REPORTED. THE PATIENT'S CONDITION WAS GOOD POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17181633 ↗MalfunctionJun 6, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY BELOW KNEE. A 4.0MM X 40MM X 135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INITIAL INFLATION AT 14 ATMOSPHERES FOR ABOUT 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17075444 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page