Skip to content

Sterling™ H74939031453010Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Sterling (153)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729757054), received between Jul 12, 2022 and Sep 16, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture5
  • Leak/Splash2
  • Inflation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2024Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash

    IT WAS REPORTED THAT BALLOON TORN OCCURRED. THE TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND CALCIFIED CAROTID ARTERY. A 4.5MMX30MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, THE BALLOON DID NOT INFLATE. THE DEVICE WAS THEN REMOVED FROM THE BODY AND CHECKED. IT WAS CONFIRMED THAT THE BALLOON WAS TORN AND COULD NOT BE USED. THE PROCEDURE WAS COMPLETED WITH NON-BOSTON SCIENTIFIC DEVICE, AND THE TREATMENT PROGRESSED WITHOUT ANY ISSUES. NO COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20224598 ↗
  • MalfunctionJul 2, 2024Leak/Splash; Material Rupture; Leak/Splash; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED INTERNAL CAROTID ARTERY. A 4.5MMX30MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. UPON INFLATION, IT WAS NOTED THAT THE BALLOON RUPTURED, AND A LEAK OF CONTRAST MEDIUM WAS CONFIRMED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19655596 ↗
  • MalfunctionMar 19, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON HOLE OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED CAROTID ARTERY. A (B)(6) STERLING BALLOON CATHETER WAS SELECTED FOR USE, THE PREPARATION WAS PERFORMED FOR PRE-DILATION, HOWEVER, IT WAS NOTED THAT THERE WAS A HOLE IN THE BALLOON. THE USAGE WAS CONTINUED, AND THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. NO COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18937952 ↗
  • MalfunctionAug 28, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 4.5MMX30MMX135CM STERLING BALLOON CATHETER WAS SELECTED FOR USE. HOWEVER, DURING TEST INFLATION OUTSIDE THE PATIENT, THE BALLOON RUPTURED DURING FIRST INFLATION AT 2 ATMOSPHERES FOR 2 SECONDS. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17632567 ↗
  • MalfunctionJun 14, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT PRESENTED WITH PERIPHERAL ARTERIAL DISEASE. THE 100% TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY POPLITEAL ARTERY. A 4.5MMX30MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INITIAL INFLATION AT 10 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17126758 ↗
  • MalfunctionJul 12, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE INTERNAL CAROTID ARTERY. A 4.5MMX30MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. THE DEVICE UNABLE TO INFLATE AND THE PRESSURE OF THE INFLATOR DID NOT INCREASE. A BALLOON RUPTURE WAS SUSPECTED. THE DEVICE WAS REMOVED AND INFLATION WAS TRIED OUTSIDE THE PATIENT, THE CONTRAST AGENT LEAKED AND A BALLOON RUPTURE HAS OCCURRED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14996375 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page