Sterling™ H74939031602010Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Sterling (153)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Sterling balloon dilatation catheter is used to widen narrowed arteries. The 49 reports mostly describe the balloon bursting during inflation, sometimes on the first try, plus a few cases of the balloon leaking or failing to deflate and a puncture found in the packaging. One injury was reported and there were no deaths, and 48 of the 49 reports are malfunctions. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 49
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 08714729757252), received between Aug 30, 2018 and Jun 10, 2026 — classified as 1 injury, 48 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (45 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture45
- Adverse Event Without Identified Device or Use Problem2
- Deflation Problem1
- Detachment of Device or Device Component1
- Fluid/Blood Leak1
- Leak/Splash1
- Material Puncture/Hole1
Patient problems
- No Clinical Signs, Symptoms or Conditions28
- No Consequences Or Impact To Patient21
- Embolism/Embolus1
- No Known Impact Or Consequence To Patient1
- Stroke/CVA1
Recent adverse event reports
The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 10, 2026Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. A 6.0MMX20MMX135CM (4F) STERLING WAS SELECTED FOR USE. DURING THE PROCEDURE, THE BALLOON RUPTURED AT NOMINAL PRESSURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT AND THE PATIENT STATUS WAS STABLE POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25459256 ↗MalfunctionNov 17, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED VESSEL IN THE FOREARM. A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE INITIAL INFLATION AT 6 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23566019 ↗MalfunctionNov 3, 2025Fluid/Blood Leak; Leak/Splash; Fluid/Blood Leak; Leak/Splash
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 08OCT2025. IT WAS REPORTED THAT BALLOON LEAK OCCURRED. THE TARGET LESION WAS LOCATED IN THE THORACIC AORTA AND SUBCLAVIAN ARTERY. A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS SELECTED FOR USE IN A THORACIC AORTIC STENT-GRAFT IMPLANTATION (BRANCH REVASCULARIZATION). DURING THE PROCEDURE, A BOSTON SCIENTIFIC PRESSURE PUMP WAS USED TO INJECT CONTRAST AGENT INTO A BALLOON TO PRE-DILATE THE STENT. UNDER BALLOON ANGIOGRAPHY, THERE WAS AN OBVIOUS MANIFESTATION OF CONTRAST AGENT LEAKAGE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. HOWEVER, DEVICE ANALYSIS REVEALED A RUPTURE AT 9CM FROM THE TIP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23450035 ↗MalfunctionJul 25, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL. A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION UNTIL IT REACH AT 6 ATMOSPHERES. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22621077 ↗MalfunctionMar 7, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND MILD CALCIFIED PULMONARY ARTERY. A 6.0MMX20MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON SECOND INFLATION AT 14 ATMOSPHERES FOR 15 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21551519 ↗MalfunctionOct 31, 2024Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE NON-CALCIFIED PULMONARY ARTERY. A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON INITIAL INFLATION AT 5 ATMOSPHERES FOR 30 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEMS, AND THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20578584 ↗MalfunctionJul 10, 2024Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT DEVICE PERFORATION OCCURRED. A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS SELECTED FOR TREATMENT. DURING PREPARATION, ASPIRATION WAS DONE, AND BUBBLES WERE NOTICED CONTINUOUSLY COMING BACK INTO THE SYRINGE. PER THE REPORT, IT IS LIKELY AN UNDERLYING PERFORATION IN THE PACKAGING. DUE TO THIS, PREPARATION WAS NOT DONE. NO PATIENT COMPLICATIONS WERE REPORTED. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19714882 ↗MalfunctionApr 29, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION AT 4 ATMOSPHERES FOR 2 SECONDS, THE BALLOON RUPTURED IN A PINHOLE FORM AT NEAR THE CENTER OF THE BALLOON. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19201021 ↗MalfunctionFeb 1, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 10 ATMOSPHERES FOR LESS THAN 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18628501 ↗MalfunctionDec 20, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 6.0MMX20MMX135CM (4F) STERLING BALLOON WAS ADVANCED FOR DILATATION. HOWEVER, THE BALLOON WAS PIERCED. NO COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18369359 ↗MalfunctionAug 7, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE EXTERNAL CAROTID ARTERY (ECA). A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, WHILE INFLATING TO THE RATED BURST PRESSURE, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT CONDITION WAS FINE. IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS 70% STENOSED AND WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED ECA. THE BALLOON RUPTURED DURING THE SECOND INFLATION AT 8 ATMOSPHERES FOR 1 MINUTE AND 30 SECONDS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17474286 ↗MalfunctionJun 19, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT PRESENTED WITH PERIPHERAL ARTERY DISEASE (PAD). THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED POPLITEAL ARTERY. A 6.0MM X 20MM X 135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17160045 ↗MalfunctionFeb 26, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 4 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED VERTICALLY. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT INJURIES REPORTED AND THE PATIENT CONDITION WAS GOOD POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16441262 ↗MalfunctionFeb 1, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED, TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED RIGHT SUPERFICIAL FEMORAL ARTERY. A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT ABOUT 10 ATMOSPHERES FOR A FEW SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16279310 ↗MalfunctionDec 13, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN MODERATELY TORTUOUS AND MODERATELY CALCIFIED ILIAC ARTERY. A 6.0MM X 20MM X 135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. INFLATION WAS PERFORMED TWO TIMES. HOWEVER, DURING THE SECOND INFLATION AT 14 ATMOSPHERES IN 10 SECONDS, THE BALLOON WAS RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15970497 ↗MalfunctionAug 9, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED EXTERNAL ILIAC ARTERY. A 6.0MMX20MMX135CM FG STERLING MR BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING SECOND INFLATION AT 10 ATMOSPHERES FOR 3 SECONDS, THE BALLOON RUPTURED AT THE MIDDLE PART. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A 6.0X40 NON-BSC DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT WAS IN GOOD CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15195678 ↗InjuryAug 9, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT EMBOLISM OCCURRED AND THE PATIENT EXPERIENCED CEREBRAL VASCULAR ACCIDENT. THE PATIENT UNDERWENT LEFT CAROTID ANGIOPLASTY. AFTER CROSSING THE LESION WITH A HYDROPHILIC GUIDEWIRE, A NON-BOSTON SCIENTIFIC STENT WAS IMPLANTED FOLLOWED BY POST-DILATATION WITH A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER. THE PATIENT WAS ASSESSED NEUROLOGICALLY AND EXHIBITED DEFICITS IN SPEECH AND CONSCIOUSNESS. CEREBRAL PAN ANGIOGRAPHY WAS PERFORMED WHICH REVEALED AN OCCLUSION AT THE MIDDLE CEREBRAL ARTERY DUE TO POSSIBLE EMBOLISM. THROMBOLYTIC TREATMENT WAS THEN STARTED AND THE PATIENT RECOVERED PART OF MOBILITY AND SPEECH AFTER ONE HOUR. NO FURTHER COMPLICATIONS WERE REPORTED AND THE PATIENT WAS FULLY RECOVERED.
Patient problems: Embolism/Embolus; Stroke/CVA
MAUDE report 15195229 ↗MalfunctionJun 14, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED COMMON ILIAC ARTERY. A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INITIAL INFLATION AT 10 ATMOSPHERES FOR 40 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14689952 ↗MalfunctionMay 12, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED PULMONARY ARTERY. A 6.0MMX20MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 10 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED VERTICALLY. THE DEVICE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED. THE PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14381036 ↗MalfunctionApr 29, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT UNDERWENT PERIPHERAL ANGIOPLASTY. THE TARGET LESION WAS LOCATED IN THE RIGHT SUPERFICIAL FEMORAL ARTERY. A 6.0MM X 20MM X 135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION AT 12 ATMOSPHERES, THE BALLOON RUPTURED. THERE WERE NO PATIENT COMPLICATIONS REPORTED. THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14254490 ↗MalfunctionApr 26, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL. A 6.0MM X 20MM X 135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION BEFORE NOMINAL PRESSURE, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14213467 ↗MalfunctionFeb 15, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED RENAL ARTERY. A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING FIRST INFLATION AT 14 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED ON THE HUB SIDE OF THE CATHETER FROM THE CENTER OF THE BALLOON. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13538145 ↗MalfunctionDec 29, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND NON-CALCIFIED ILIAC ARTERY. A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE INITIAL INFLATION, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13112995 ↗MalfunctionDec 1, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 7X80, 130 CM ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM WAS DEPLOYED, AND IT WAS NOTICED THAT THE EDGE OF THE STENT WAS LIFTED. A 6.0MM X 20MM X 135CM STERLING BALLOON CATHETER WAS THEN ADVANCED TO EXPAND THE ELUVIA STENT. HOWEVER, DURING THE INITIAL INFLATION AT 4 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND REPLACED WITH ANOTHER 6.0MM X 40MM X 135CM STERLING BALLOON CATHETER WHICH ALSO RUPTURED DURING THE INITIAL INFLATION AT 4 ATMOSPHERES FOR 5 SECONDS. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH IMPLANTATION OF AN ADDITIONAL 7X40 ELUVIA STENT. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED. THE PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12913912 ↗MalfunctionOct 28, 2021Deflation Problem; Deflation Problem
IT WAS REPORTED THAT BALLOON DEFLATION FAILURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE CAROTID ARTERY. A 6.0MMX20MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR POST-DILATATION. HOWEVER, AFTER THE BALLOON WAS INFLATED, IT WOULD NOT DEFLATE EVEN AFTER MULTIPLE ATTEMPTS. ANOTHER INFLATION DEVICE WAS TRIED BUT STILL WOULDN'T DEFLATE. FINALLY, THE PHYSICIAN WAS ABLE TO GET THE BALLOON IN THE TRANSCAROTID ARTERY REVASCULARIZATION SHEATH AND PULL THE WHOLE SYSTEM OUT WITH NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12711764 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page