Sterling™ H74939031704010Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Sterling (153)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Sterling balloon catheter, used to open narrowed arteries in the legs, has 115 adverse event reports on file. Most describe the balloon rupturing during inflation, usually with the procedure finished using another device. The reports include no deaths and 4 injuries, and the number of reports fell from 11 to 4 over the last 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 115
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 115 adverse event reports that list this exact device (DI 08714729757443), received between Nov 13, 2018 and Apr 1, 2026 — classified as 4 injuries, 111 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (109 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture109
- Break3
- Deflation Problem1
- Detachment of Device or Device Component1
- Difficult to Advance1
- Inflation Problem1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions71
- No Consequences Or Impact To Patient41
- Foreign Body In Patient1
- No Code Available1
- No Patient Involvement1
- Vascular Dissection1
Recent adverse event reports
The 25 most recent of 115 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 1, 2026Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 7.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 5 ATMOSPHERES FOR 2 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT SIGNIFICANT CALCIFICATION WAS OBSERVED THROUGHOUT THE LESION AND THE BLOOD VESSELS. HOWEVER, IT CANNOT BE RULED OUT THAT THE BALLOON RUPTURE WAS CAUSED BY NODULAR CALCIFICATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24757293 ↗MalfunctionMar 10, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY SEVERELY SUPERFICIAL FEMORAL ARTERY. A 7.0MMX40MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE FIRST INFLATION AT 10 ATMOSPHERES. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24565320 ↗MalfunctionJan 12, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN A VESSEL OF THE ILIAC. A 7.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 8 ATMOSPHERES FOR 30 SECONDS. THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24046383 ↗MalfunctionNov 3, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 7.0MMX40MMX135CM STERLING BALLOON CATHETER WAS SELECTED FOR USE IN AN ENDOVASCULAR TREATMENT. DURING THE FIFTH INFLATION AT 10 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED AT THE CENTER. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23449892 ↗MalfunctionOct 27, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 7.0MMX40MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 6 ATMOSPHERES FOR 3 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23397498 ↗MalfunctionSep 4, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED COMMON FEMORAL ARTERY. A 7.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS SELECTED FOR USE IN AN ENDOVASCULAR THERAPY. DURING THE FIRST INFLATION AT 10 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED FROM A PINHOLE IN THE CENTER. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22961904 ↗MalfunctionAug 4, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED ILIAC ARTERY. A 7.0MMX40MMX135CM (4F) STERLING MR BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON SECOND INFLATION AT 10 ATMOSPHERES. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22698239 ↗MalfunctionJul 23, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED IN THE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA). A 7.0MMX40MMX135CM (4F) STERLING BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION SHORTLY BEFORE NOMINAL PRESSURE FOR 15 SECONDS, THE BALLOON RUPTURED. EVEN WHEN IT WAS REMOVED FROM THE BODY AND EXPANDED, NO PRESSURE WAS GENERATED, AND A PINHOLE-LIKE SPOT AT THE TIP COULD BE VISUALLY CONFIRMED. ALTHOUGH THERE WAS CALCIFICATION IN THE LESION, THERE WAS NO SNAGGING OR OTHER EFFECT ON BALLOON INSERTION. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22581216 ↗MalfunctionJul 21, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS AN AREA OF IN-STENT RESTENOSIS (ISR) LOCATED IN THE MILD TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 7.0MMX40MMX135CM (4F) FG STERLING MR BALLOON CATHETER WAS USED TO DILATE A PREVIOUSLY IMPLANTED ELUVIA DRUG-ELUTING STENT. DURING THE PROCEDURE, THE BALLOON RUPTURED ON SECOND INFLATION AT 8 ATMOSPHERES FOR 30 SECONDS. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22561577 ↗MalfunctionJul 1, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED COMMON ILIAC ARTERY. A 7.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 6 ATMOSPHERES FOR 5 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22372625 ↗MalfunctionJun 10, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED COMMON ILIAC ARTERY. A 7.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. THE BALLOON WAS INFLATED TO 10, 10, AND 12 ATMOSPHERES RESPECTIVELY. HOWEVER, DURING THE THIRD INFLATION FOR A FEW SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22177945 ↗MalfunctionMay 12, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE STENOSED TARGET LESION WAS LOCATED IN THE EXTERNAL ILIAC ARTERY. A 7.0MMX40MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE FIRST INFLATION AT NOMINAL PRESSURE. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: Foreign Body In Patient
MAUDE report 22006910 ↗MalfunctionFeb 21, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED COMMON FEMORAL ARTERY. A 7.0MMX40MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 6 ATMOSPHERES FOR 3 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21440142 ↗MalfunctionJan 9, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 7.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON INITIAL INFLATION AT 8 ATMOSPHERES FOR 10 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEMS, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21118406 ↗MalfunctionDec 27, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED ILIAC VEIN. A 7.0MMX40MMX135CM (4F) STERLING BALLOON WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON INITIAL INFLATIONS AT 8 ATMOSPHERES FOR 10 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEMS AND REPLACED IT WITH A 10.0 X 40, 75CM MUSTANG BALLOON. HOWEVER, IT ALSO RUPTURED AT FIRST INFLATION AT 15 ATMOSPHERES FOR 10 SECONDS. A NEW 6.0MMX40MMX135CM (4F) STERLING BALLOON WAS USED BUT IT ALSO RUPTURED. THE PROCEDURE WAS THEN COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21029691 ↗MalfunctionSep 24, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MILD TORTUOUS AND SEVERELY CALCIFIED COMMON FEMORAL ARTERY. A 7.0MMX40MMX135CM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 6 ATMOSPHERES. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20295725 ↗MalfunctionSep 10, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 7.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20184957 ↗MalfunctionSep 9, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED COMMON FEMORAL ARTERY. A 7.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING FIRST INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED VERTICALLY AT NEAR THE CENTER. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20172041 ↗MalfunctionSep 8, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATE TORTUOUS AND MILDLY CALCIFIED ILIAC ARTERY. A 7.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, ON THE FIRST INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM AND NO DAMAGE WAS OBSERVED. A DIFFERENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO COMPLICATIONS REPORTED, AND PATIENT STATUS WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20166066 ↗InjurySep 4, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE AND DISSECTION OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED BILATERAL COMMON ILIAC ARTERY. AN OPTICROSS 18 VASCULAR IMAGING CATHETER AND A STERLING BALLOON WERE SELECTED FOR USE. A SHEATH WAS INSERTED FROM BOTH DORSALIS PEDIS ARTERIES, AND A NON-BOSTON SCIENTIFIC 0.014-INCH GUIDEWIRE WAS CROSSED THE LESION. BILATERAL LESION DILATATION WAS PERFORMED BY A SMALL DIAMETER COYOTE 4-40 BALLOON BEFORE THE PRE-DILATION. THE LESION WAS CONFIRMED BY AN OPTICROSS TO BE A CIRCUMFERENTIAL SEVERELY CALCIFIED LESION, SO A BALLOON SIZE WITH A LITTLE EXTRA SPACE FOR THE LUMEN WAS SELECTED. THE LEFT COMMON ILIAC ARTERY WAS DILATED USING THIS STERLING DEVICE. HOWEVER, UPON SEVERAL INFLATIONS AT 6 - 10 ATMOSPHERES FOR ABOUT 10 SECONDS, THE BALLOON RUPTURED VERTICALLY. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE. THE DILATATION WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THE RIGHT COMMON ILIAC ARTERY WAS DILATED USING STERLING 6-40. AFTERWARDS, THE LUMEN WAS CHECKED USING OPTICROSS. THERE WAS A LITTLE RESISTANCE. IT WAS POSSIBLE THAT THE PHYSICIAN PUSHED A LITTLE TOO HARD AND IT KINKED. AFTE
Patient problems: Vascular Dissection
MAUDE report 20138130 ↗MalfunctionAug 5, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY A 7.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19905391 ↗MalfunctionMay 6, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED EXTERNAL ILIAC ARTERY. A 7.0MMX40MMX135CM (4F) STERLING BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE SECOND INFLATION AT 10 ATMOSPHERES FOR FEW SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19254909 ↗MalfunctionApr 8, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED COMMON FEMORAL ARTERY. A 7.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED IN A PINHOLE FORM AT NEAR THE DISTAL SHOULDER. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19066747 ↗MalfunctionMar 6, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURED OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 7.0MMX40MMX135CM (4F) FG STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED IN THE MIDDLE PART UPON SECOND INFLATION AT 10 ATMOSPHERES FOR 20 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEMS, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18852742 ↗MalfunctionFeb 15, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS AN AREA OF IN-STENT RESTENOSIS (ISR) LOCATED IN A MODERATELY TORTUOUS AND SEVERELY CALCIFIED COMMON FEMORAL ARTERY. A 7.0MMX40MMX135CM (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, UPON INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18716577 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page