Skip to content

ZIPwire™ M006630209B0Recalls & reported problems

Lake Region Medical · Class 1

All models of ZIPwire (24)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729761730), received between Oct 8, 2020 and Dec 17, 2021 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Device Damaged by Another Device” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Device Damaged by Another Device4
  • Break1
  • Entrapment of Device1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 17, 2021Device Damaged by Another Device; Material Split, Cut or Torn; Device Damaged by Another Device; Material Split, Cut or Torn

    IT WAS REPORTED THAT: PATIENT COMPLICATIONS: NO PATIENT COMPLICATIONS. FAILURE MODES RELATED TO THIS DEVICE INCLUDE: COATING ISSUE. HOW WAS THE PROCEDURE COMPLETED (WITH ANOTHER OF THE SAME DEVICE, WITH A DIFFERENT DEVICE, ETC.)? COMPLETED WITH SAME DEVICE. ADDITIONAL INFORMATION PROVIDED (B)(6) 2021: FAILURE MODES RELATED TO THIS DEVICE INCLUDE: COATING ISSUE. WERE THE ISSUES PRESENT UPON OPENING THE PACKAGE?: NO DO YOU HAVE A PHOTO, VIDEO, OR SCREENSHOT DEPICTING THE ISSUE?: YES HOW WAS THE PROCEDURE COMPLETED (WITH ANOTHER OF THE SAME DEVICE, WITH A DIFFERENT DEVICE, ETC.)? COMPLETED WITH SAME DEVICE. IF THERE WAS A PROCEDURE INVOLVED, WAS IT FOR TREATMENT OR DIAGNOSIS? TREATMENT IF THERE WAS A PROCEDURE INVOLVED, WHAT WAS THE ANATOMY LOCATION? KIDNEY/URETER WAS THIS THE FIRST TIME THE DEVICE WAS USED (NOT RESTERILIZED OR REPROCESSED)? YES PLEASE DESCRIBE WHAT PORTION OF THE GUIDEWIRE DETACHED/PEELED HYDROPHILIC TIP, PTFE COATING, OTHER?

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13025536 ↗
  • MalfunctionApr 26, 2021Device Damaged by Another Device; Device Damaged by Another Device

    EVENT DESCRIPTION: PER EMAIL, IT WAS REPORTED THAT: THE GUIDEWIRE BROKE DURING A PROCEDURE. A SMALL PIECE OF IT WAS IN THE PATIENT THANKFULLY STAFF WERE ABLE TO GET THE SMALL PIECE OUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11726609 ↗
  • MalfunctionFeb 22, 2021Device Damaged by Another Device; Device Damaged by Another Device

    THE END OF THE ZIPWIRE WAS TORN AFTER ONE PASS THROUGH OF THE WIRE. THE EVENT OCCURED DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING ANOTHER ZIPWIRE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11361150 ↗
  • InjuryOct 26, 2020Device Damaged by Another Device; Device Damaged by Another Device

    AFTER COMPLETION OF PROCEDURE THE PATIENT CHEST XRAY WAS ABNORMAL. WHEN THE ZIPWIRE WAS RETRIEVED, INSPECTED, AND COMPARED TO A NEW UNUSED ZIPWIRE IT WAS NOTED THAT A VERY THIN PART OF THE COATING WAS MISSING. PATIENT WAS BROUGHT BACK TO SURGERY TO REMOVE FOREIGN BODY FROM THE GUIDEWIRE THAT WAS SEEN ON POST-OP XRAY.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 10734993 ↗
  • InjuryOct 8, 2020Break; Entrapment of Device

    AFTER COMPLETION OF PROCEDURE (INSERTION OF VASCULAR PORT) A CHEST XRAY WAS PERFORMED. RESULTS SUSPECTED RETAINED FOREIGN BODY. THE ZIPWIRE USED IN THE CASE WAS RETRIEVED AND INSPECTED. WHEN COMPARED TO A NEW (UNUSED) ZIPWIRE, IT WAS NOTED THAT A VERY THIN PART OF THE COATING WAS MISSING. THE PATIENT WAS RETURNED TO SURGERY AND AN APPROXIMATELY 15 CM PIECE OF THE GUIDEWIRE WAS RETRIEVED. FDA SAFETY REPORT ID# (B)(4).

    Patient problems: Foreign Body In Patient

    MAUDE report 10657915 ↗

FDA clearance and classification

The FDA regulates product code EYA (stylet, ureteral) under 21 CFR 876.5130 as a Class I device. 21 CFR 876.5130 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the ZIPwire™ M006630209B0

Has the ZIPwire™ M006630209B0 been recalled?

No FDA recall record in our data names this exact device identifier (DI 08714729761730), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the ZIPwire™ M006630209B0?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Oct 8, 2020 and Dec 17, 2021. In those reports, the most common specific device problems are “Device Damaged by Another Device” (4), “Break” (1) and “Entrapment of Device” (1); the most common patient problems are “Device Embedded In Tissue or Plaque” (1) and “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ZIPwire™ M006630209B0?

None of the reports are classified as deaths, 2 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ZIPwire™ M006630209B0 FDA approved?

Devices of this type (product code EYA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Lake Region Medical.

Who makes the ZIPwire™ M006630209B0, and is it still sold?

The FDA's device database lists Lake Region Medical as the company responsible for it (brand ZIPwire™, model M006630209B0). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Stylet, Ureteral devices

All Stylet, Ureteral devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page