Skip to content

Blazer™ Dx-20 M00420SL21020Recalls & reported problems

Boston Scientific Corporation · Bidirectional Steerable Diagnostic Catheter · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 08714729780526), received between Jun 12, 2019 and Nov 5, 2025 — classified as 7 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem6
  • Break1
  • Device Contamination with Chemical or Other Material1
  • Difficult to Remove1
  • Tear, Rip or Hole in Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Chest Pain2
  • Pericardial Effusion2
  • Cardiac Perforation1
  • Cardiac Tamponade1
  • Cardiopulmonary Arrest1
  • Congestive Heart Failure1
  • Dyspnea1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED PERICARDIAL EFFUSION. A BLAZER DX-20 CATHETER WAS USED DURING AN ATRIAL FIBRILLATION (AF) PROCEDURE. DURING THE PROCEDURE, A NON-SUSTAINED ATRIAL TACHYCARDIA WAS OBSERVED IN THE LATERAL WALL OF THE RIGHT ATRIUM. AN INTELLAMAP ORION CATHETER WAS UTILIZED TO MAP THE PATIENT'S CARDIAC ACTIVITY. THE PHYSICIAN SUSPECTS THAT THE DIAGNOSTIC CATHETER MAY HAVE PUSHED TOO HARD CAUSING A SMALL PERFORATION NEAR THE TIP OF THE LATERAL ANNULUS DURING THE TACHYCARDIA EPISODE. NO MEDICAL OR SURGICAL INTERVENTION WAS PERFORMED, AS THE PHYSICIAN IS CONFIDENT THAT THE COMPLICATION WILL RESOLVE SPONTANEOUSLY. THE ISSUE WAS RESOLVED, AND THE PROCEDURE WAS COMPLETED WITH THIS DEVICE, THE ABLATION WORKFLOW WAS NORMAL AND NOT PROLONGED. THE PATIENT REMAINS ADMITTED FOR MONITORING AND IS EXPECTED TO RECOVER. THE NEXT DAY, THE PATIENT WAS DISCHARGED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS, AS IT WAS DISPOSED OF BY THE FACILITY.

    Patient problems: Cardiac Perforation; Pericardial Effusion; Tachycardia

    MAUDE report 23476668 ↗
  • InjuryMar 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CLINICAL INTERRUPT AF CLINICAL STUDY, CLINICAL STUDY ID: (B)(4), SUBJECT ID: (B)(4). IT WAS REPORTED THAT DURING A ABLATION PROCEDURE INVOLVING AN INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER, BLAZER DX-20 DIAGNOSTIC CATHETER AND A INTELLANAV MIFI O WERE SELECTED FOR USE. AFTER PULMONARY VEIN ISOLATIONS, BLOOD PRESSURE DECREASED AND INCREASED PRESSOR SUPPORT WAS REQUIRED. A LARGE PERICARDIAL EFFUSION WAS NOTED ON INTRACARDIAC ECHOCARDIOGRAM (ICE) IMAGING. EMERGENT PERICARDIOCENTESIS WAS PERFORMED.

    Patient problems: Cardiac Tamponade; Pericardial Effusion

    MAUDE report 16632048 ↗
  • MalfunctionNov 16, 2022Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    DURING PREPARATION OF A RIGHT ATRIAL FLUTTER PROCEDURE, A BLAZER DX CATHETER WAS SELECTED FOR USE. UPON OPENING THE PACKAGE, SOMETHING BLACK WAS OBSERVED BEHIND THE STEERABLE BI-WING KNOB. AN EXCHANGE OF THE CATHETER RESOLVED THE ISSUE TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15809413 ↗
  • MalfunctionSep 10, 2021Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT THE TECHNICIAN WAS OPENING THE BOX FOR THE BLAZER DX-20, AND NOTICED A TEAR IN THE PLASTIC PACKAGING PROTECTING THE STERILE CATHETER. THIS CATHETER WAS NOT USED, AND NEW ONE WAS PULLED OUT TO USE. THE PACKAGING WAS FINE ON THE SECOND CATHETER, AND IT WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12453067 ↗
  • InjuryJul 27, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NEWTON AF CLINICAL STUDY: IT WAS REPORTED THAT A BLAZER DX-20 WAS USED IN AN ATRIAL FIBRILLATION ABLATION PROCEDURE. AFTER THE PROCEDURE THE PATIENT EXPERIENCED CHEST TIGHTNESS THAT EXTENDS UP TO THEIR THROAT AT TIMES AND MAKES IT HARD TO BREATH WHICH BECAME WORSE AT NIGHT. THE PATIENT WAS GIVEN COLCHICINE AND HAD STAT ECHOCARDIOGRAM.

    Patient problems: Chest Pain

    MAUDE report 12233997 ↗
  • InjuryApr 28, 2021Break; Difficult to Remove; Break; Difficult to Remove

    IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE TO TREAT PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIA (PSVT), IN THE PROCESS OF APPROACHING THE BLAZER DX CATHETER TO THE CORONARY SINUS (CS) OSTIUM, THE CATHETER "WAS NOT CURVED." THE CATHETER DIDN'T MOVE AT THE TRICUSPID VALVE (TV) MUSCLE DUE TO IT BEING UNABLE TO CURVE FOR CS APPROACH. THE CURVE WAS JAMMED OR UNABLE TO BE STRAIGHTENED. THE CATHETER WAS "STUCK" AT THE MUSCLE AND WOULD NOT MOVE. THE PHYSICIAN TRIED MANY TIMES TO EXTRACT THE DEVICE, BUT WAS UNABLE TO. THEREFORE, THE PHYSICIAN NEEDED TO USE A TRIO SHEATH CONNECTOR TO EXTRACT THE BLAZER DX CATHETER FROM THE MUSCLE. THE PROCEDURE WAS THEN COMPLETED WITH ANOTHER DEVICE. THE DEVICE WAS RETURNED TO BOSTON SCIENTIFIC FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11741321 ↗
  • InjuryMay 12, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    (B)(6) CLINICAL STUDY. IT WAS REPORTED THE PATIENT EXPERIENCED ACUTE DIASTOLIC CONGESTIVE HEART FAILURE (CHF). THREE DAYS FOLLOWING AN ABLATION PROCEDURE WITH AN INTELLANAV MIFI OPEN-IRRIGATED CATHETER, A BLAZER DX-20 AND AN INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER, THE PATIENT WENT TO THE HOSPITAL FOR SHORTNESS OF BREATH ON EXERTION AND WEAKNESS. THE SUSPECTED CAUSE WAS ACUTE DIASTOLIC CONGESTIVE HEART FAILURE. LASIX WAS ADDED TO THE PATIENT'S ORAL MEDICATION TREATMENT AND THE EVENT RESOLVED.

    Patient problems: Congestive Heart Failure; Dyspnea; Weakness

    MAUDE report 10046568 ↗
  • InjuryFeb 12, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    (B)(6) CLINICAL STUDY. IT WAS REPORTED THAT THE PATIENT EXPERIENCED CHEST PAIN. AN ABLATION PROCEDURE WAS COMPLETED USING AN INTELLANAV MIFI OPEN-IRRIGATED CATHETER, AN INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER, A BLAZER DX-20 CATHETER, A NON-BOSTON SCIENTIFIC DIAGNOSTIC CATHETER AND FOUR NON-BOSTON SCIENTIFIC SHEATHS. ON (B)(6) 2019, THE PATIENT PRESENTED TO OUTSIDE HOSPITAL WITH CHEST PAIN WHICH DID NOT RESPOND TO NITROGLYCERIN. AN ECHOCARDIOGRAM WAS COMPLETED WHICH DID NOT SHOW AN EFFUSION. THE PATIENT WAS MONITORED OVERNIGHT AND DISCHARGED WITH COLCRYS FOR PAIN MANAGEMENT. THE SUSPECTED CAUSE WAS MILD PERICARDITIS. THE EVENT RESOLVED ON (B)(6) 2020.

    Patient problems: Chest Pain

    MAUDE report 9701288 ↗
  • InjuryJun 12, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED HYPOTENSION. DURING AN ABLATION PROCEDURE FOR ATRIAL FLUTTER, A BLAZER DX-20 CATHETER WAS INSERTED IN THE PATIENT WITH NO ISSUES. THEN, AN INTELLAMAP ORION CATHETER WAS INSERTED INTO THE PATIENT WITH NO ISSUES. ABOUT 5 MINUTES LATER, THE PATIENT'S BLOOD PRESSURE DROPPED. THEY HAD TO DO COMPRESSIONS AND THEN THE PATIENT'S PRESSURE WAS FINE. THE CASE WAS THEN CANCELLED. THE DOCTOR CHECKED WITH ULTRASOUND TO CHECK IF THERE WAS AN EFFUSION FROM THE CATHETERS AND THERE WAS NOT. THE PHYSICIAN BELIEVED THAT THE EVENT WAS DUE AN ANESTHESIA ISSUE AND NOT THE DEVICES. THE PATIENT WAS FINE AND THERE WERE NO OTHER COMPLICATIONS. IT WAS FURTHER REPORTED THAT THE "COMPRESSIONS" MEANT THE PATIENT REQUIRED CARDIOPULMONARY RESUSCITATION (CPR).

    Patient problems: Cardiopulmonary Arrest; Low Blood Pressure/ Hypotension

    MAUDE report 8691221 ↗

FDA clearance and classification

The FDA regulates product code DRF (catheter, electrode recording, or probe, electrode recording) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Blazer™ Dx-20 M00420SL21020

Has the Blazer™ Dx-20 M00420SL21020 been recalled?

No FDA recall record in our data names this exact device identifier (DI 08714729780526), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Blazer™ Dx-20 M00420SL21020?

The FDA's MAUDE database lists 9 adverse event reports for this device, received between Jun 12, 2019 and Nov 5, 2025. In those reports, the most common specific device problems are “Break” (1), “Device Contamination with Chemical or Other Material” (1) and “Difficult to Remove” (1); the most common patient problems are “Chest Pain” (2), “Pericardial Effusion” (2) and “Cardiac Perforation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Blazer™ Dx-20 M00420SL21020?

None of the reports are classified as deaths, 7 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Blazer™ Dx-20 M00420SL21020, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Blazer™ Dx-20, model M00420SL21020). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices

All CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page