Skip to content

Carotid WALLSTENT® M001719000Recalls & reported problems

Boston Scientific Corporation · Closed Cell Self-Expanding Stent · Class 3 · Implant

All models of Carotid WALLSTENT (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jul 7, 2025Class IOpen, Classified

    Certain batches are being removed due to an increase in complaints received in which physicians encountered greater than anticipated resistance while attempting to withdraw the stent delivery system (SDS) from the guidewire or embolic protection device (EPD) after successful stent deployment. The most serious potential adverse health consequence is stroke. Other risks include delay to procedure, vessel injury, vessel spasm, or stent disruption/damage necessitating additional intervention.

    Z-2273-2025 · Root cause: Process change control · Recall details · FDA record ↗

Other STENT, CAROTID devices

All STENT, CAROTID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page