Skip to content

Sterling® SL H74939148201210Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Sterling SL (35)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729782223), received between Mar 19, 2019 and Apr 7, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture5
  • Adverse Event Without Identified Device or Use Problem1
  • Failure to Advance1
  • Material Twisted/Bent1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient2
  • Thrombosis/Thrombus1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 7, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON BURST OCCURRED. THE 60% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED ARTERY OF THE LOWER EXTREMITY. A 2.0MM X 120MM X 150CM STERLING SL BALLOON CATHETER WAS ADVANCED FOR BILATERAL PUNCTURE OF THE LOWER EXTREMITY PROCEDURE. HOWEVER, THE BALLOON BURST DURING SECOND INFLATION AT 20 ATMOSPHERES FOR 2 MINUTES. THE DEVICE WAS REMOVED NORMALLY, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT'S CONDITIONS WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21787399 ↗
  • MalfunctionJul 31, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LOCATION WAS LOCATED IN THE PERIPHERAL ARTERY. A 2.0MM X 120MM X 150CM STERLING SL BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT BELOW 6 ATMOSPHERES. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19874855 ↗
  • MalfunctionFeb 17, 2023Failure to Advance; Material Twisted/Bent; Physical Resistance/Sticking

    MID 60¿S MALE 3 DAY POST OP FROM RIGHT SUPERFICIAL FEMORAL ARTERY (SFA) REVASCULARIZATION AND STENTING. DECREASED CIRCULATION IN THE RIGHT LOWER EXTREMITY AND WAS TAKEN FOR PROCEDURE: RIGHT LOWER EXTREMITY (RLE) ANGIOGRAM, MECHANICAL THROMBECTOMY/PLASTY. THEN END OF THE BALLOON WAS PATENT AND FLUSHED WELL BUT THE END OF THE BALLOON WOULD BUNCH UP AND NOT LET THE WIRE GO THROUGH. THE DEVICE WAS NOT USED ON THE PATIENT, MINOR DELAY IN PROCEDURE. MANUFACTURER RESPONSE FOR CATHETER, PERCUTANEOUS, STERLING® SL (PER SITE REPORTER) WILL OBTAIN.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 16396100 ↗
  • MalfunctionJun 14, 2022Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. A 2.0MM X 120MM X 150CM STERLING SL BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON BURST BEFORE REACHING THE NOMINAL PRESSURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14687860 ↗
  • MalfunctionDec 14, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED ANTERIOR TIBIAL ARTERY. A 2.0MM X 120MM X 150CM STERLING SL BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION AT 8 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND PATIENT STATUS WAS FINE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11004210 ↗
  • MalfunctionMar 19, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE TIBIAL ARTERY. A ROTALINK WAS TESTED OUTSIDE OF THE BODY AND IT WAS WORKING CORRECTLY. THE PLAQUE WAS ENGAGED WITH A ROTABLATOR, BUT IT WAS NO LONGER SPINNING INSIDE OF THE BODY. THE ROTABLATOR WAS REMOVED AND A 2.0MM X 120MM X 150CM STERLING SL BALLOON CATHETER WAS THEN ADVANCED FOR DILATATION. HOWEVER, UPON INFLATION, THE BALLOON RUPTURED BEFORE IT GOT TO THE NOMINAL SIZE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8434246 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page