Sterling® SL H74939148251210Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Sterling SL (35)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 08714729782261), received between May 14, 2020 and Nov 20, 2025 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture4
- Inflation Problem3
- Adverse Event Without Identified Device or Use Problem2
- Break2
- Entrapment of Device1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 20, 2025Inflation Problem; Inflation Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 24OCT2025. IT WAS REPORTED THAT THE BALLOON DID NOT INFLATE PROPERLY. THE TARGET LESION WAS LOCATED IN THE RIGHT LOWER LIMB. A 2.5MM X 120MM X 150CM STERLING SL BALLOON CATHETER WAS SELECTED FOR LOWER LIMB ANGIOPLASTY TO TREAT PERIPHERAL ARTERY DISEASE (PAD). DURING THE PROCEDURE, THE BALLOON DID NOT INFLATE PROPERLY. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. HOWEVER, DEVICE ANALYSIS REVEALED A PINHOLE 102MM FROM THE TIP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23607919 ↗MalfunctionFeb 24, 2025Material Rupture; Adverse Event Without Identified Device or Use Problem; Material Rupture; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE ARTERIOSCLEROTIC OCCLUSION OF LOWER EXTREMITY. A 2.5MM X 120MM X 150CM STERLING SL WAS ADVANCED FOR DILATION. DURING INFLATION, THE BALLOON BURST. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS STABLE POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21447215 ↗MalfunctionSep 27, 2024Inflation Problem; Leak/Splash; Material Rupture; Adverse Event Without Identified Device or Use Problem; Inflation Problem; Leak/Splash; Material Rupture; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE BALLOON BURST. VASCULAR ACCESS WAS OBTAINED VIA FEMORAL ARTERY WITH A 6F SHEATH AND A PLACEMENT OF NON-BOSTON SCIENTIFIC GUIDEWIRE. THE 85% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS LOWER EXTREMITY ARTERIAL OCCLUSION. THE 6F SHEATH REACHED THE LESION AND DETERMINED THE ANGIOGRAPHY. A V-18 HYDROPHILIC WIRE WAS ADVANCED WITH THE 2.5MM X 120MM X 150CM STERLING SL BALLOON CATHETER FOR DILATION. HOWEVER, THE BALLOON COULD NOT DILATE AND WAS QUICKLY REMOVED FROM THE PATIENT. THE NORMAL SALINE WAS USED TO DETECT THE ISSUE, AND FOUND THE BALLOON WAS BUBBLING, LEAKING GAS AND COULD NOT DILATE, BUT NO VISIBLE PINHOLE WAS NOTED. IT WAS DETERMINED THAT THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT'S CONDITION FOLLOWING PROCEDURE WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20317863 ↗MalfunctionJan 26, 2024Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE PATIENT PRESENTED WITH VASCULAR SCLEROSIS AND STENOSIS OF LOWER EXTREMITY. A 2.5MM X 120MM X 150CM STERLING SL BALLOON CATHETER WAS SELECTED FOR USE. HOWEVER, AFTER UNPACKING, THE SHAFT WAS FOUND TO BE BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18587737 ↗MalfunctionFeb 9, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT WAS PRESENTED WITH ARTERIAL OCCLUSION OF THE LOWER KNEE VESSELS AND UNDERWENT ARTERY VASCULAR OPENING. THE TARGET LESION WAS LOCATED IN ANTERIOR TIBIAL ARTERY. A 2.5MM X 120MM X 150CM STERLING SL BALLOON CATHETER WAS ADVANCED FOR DILATATION. INFLATIONS WERE PERFORMED THREE TIMES. HOWEVER, DURING THIRD INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16344096 ↗MalfunctionJul 25, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. A 2.5MM X 120MM X 150CM STERLING SL AND 2.5MM X 80MM X 150CM COYOTE BALLOON CATHETER WERE ADVANCED FOR DILATION. BOTH BALLOON REACHED THE LESION SITE. HOWEVER, DURING INFLATION AT 12 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED. THE PATIENT WAS IN STABLE CONDITION FOLLOWING THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15093659 ↗MalfunctionDec 14, 2020Break; Inflation Problem; Break; Inflation Problem
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE 50% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED ANTERIOR TIBIAL ARTERY. A 2.5MM X 120MM X 150CM STERLING SL BALLOON CATHETER WAS ADVANCED FOR DILATATION. THE DEVICE WAS INFLATED THRICE AT 10 ATMOSPHERES; HOWEVER, DURING INFLATION, CONTRAST COULD NOT BE SEEN. IT WAS INITIALLY SUSPECTED THAT THE CONTRAST WAS TOO DILUTED; HOWEVER, WHEN THE BALLOON WAS REMOVED AND INFLATED OUTSIDE, IT WAS FOUND THAT IT COULD NOT INFLATE. NO VISIBLE PINHOLE NOTED WITH THE DEVICE. AS PER PHYSICIAN, THE BALLOON SHAFT WAS FAULTY OR BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11004056 ↗MalfunctionMay 14, 2020Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT THE BALLOON CATHETER WAS STUCK ON THE WIRE. THE TARGET LESION WAS LOCATED IN NON-SEVERELY CALCIFIED ANTERIOR TIBIAL ARTERY BELOW THE KNEE. PRE-DILATION WAS PERFORMED WITH A 2.0MM STERLING BALLOON. A 2.5MMX80MMX150CM STERLING SL BALLOON AND V18 GUIDEWIRE WERE SELECTED WERE SELECTED FOR USE. THE BALLOON RUPTURED UPON FIRST INFLATION AT 10 ATMOSPHERES. A 2.5MMX120MMX150CM STERLING SL BALLOON WAS SELECTED TO CONTINUE THE CASE. THE 2.5MMX120MMX150CM STERLING SL BALLOON CATHETER BECAME STUCK ON THE V18 GUIDEWIRE AND COULD NOT BE PUSHED FORWARD. BOTH BALLOON CATHETER AND GUIDE WIRE WERE THEN REMOVED FROM THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10058083 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page