Skip to content

NC Quantum Apex™ H7493912420270Recalls & reported problems

Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2

All models of NC Quantum Apex (48)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729783473), received between Jun 6, 2019 and Feb 5, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture3
  • Material Separation2
  • Nonstandard Device2
  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1
  • Device Markings/Labelling Problem1
  • Inflation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient3
  • No Patient Involvement1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 5, 2026Nonstandard Device; Material Separation

    IT WAS REPORTED THAT A DETACHMENT OCCURRED. AN NC QUANTUM APEX MR 2.75 X 20MM WAS SELECTED FOR TREATMENT OF THE TARGET LESION. WHEN REMOVING THE BALLOON FROM THE PROTECTOR, THE PHYSICIAN NOTED THAT THE BALLOON WAS CUT. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24272187 ↗
  • MalfunctionAug 20, 2025Nonstandard Device; Material Separation; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Nonstandard Device; Material Separation; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. DURING PREPARATION OF THE 20MM X 2.75MM NC QUANTUM APEX? BALLOON CATHETER, NO RESISTANCE WAS ENCOUNTERED WHILE REMOVING THE BALLOON PROTECTOR/MANDREL BUT AFTERWARD, IT WAS NOTED THAT THE TIP OF THE DEVICE WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22849498 ↗
  • MalfunctionNov 29, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED RIGHT CORONARY ARTERY. A 20MM X 2.75MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR POST-DILATATION. HOWEVER, DURING THE FOURTH INFLATION AT 20 ATMOSPHERES FOR 15 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A 3.00MMX20MM NC QUANTUM APEX BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15878530 ↗
  • MalfunctionJan 6, 2021Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT LABELING ISSUE OCCURRED. A 20MM X 2.75MM NC QUANTUM APEX BALLOON CATHETER WAS HANDED TO THE REPRESENTATIVE AND EXPLAINED THAT THE BALLOON CONTAINED AN EXPANSION CHART FROM AN ENTIRELY DIFFERENT BALLOON. THE BALLOON WAS NOT USED TO THE PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 11125323 ↗
  • MalfunctionMay 11, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 20MM X 2.75MMM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10043419 ↗
  • MalfunctionFeb 6, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 20MM X 2.75MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR POST DILATION. HOWEVER, THE BALLOON RUPTURED AT 14 ATMOSPHERES TO OPTIMIZE THE STENT IMPLANT. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9677202 ↗
  • MalfunctionJun 6, 2019Inflation Problem; Inflation Problem

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2019. IT WAS REPORTED THAT BALLOON INFLATION FAILURE OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA THE FEMORAL ARTERY. THE CONCENTRIC TARGET LESION WAS LOCATED IN THE MILDLY CALCIFIED MID RIGHT CORONARY ARTERY. THE LESION CONTAINED >=90 DEGREES BEND. A 20MM X 2.75MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR DILATATION, BUT THE BALLOON FAILED TO INFLATE. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON WITH BUDDY WIRE SUPPORT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS STABLE AND DOING WELL. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A SEPARATED HYPOTUBE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8674131 ↗

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page