NC Quantum Apex™ H7493912420270Recalls & reported problems
Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2
All models of NC Quantum Apex (48)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729783473), received between Jun 6, 2019 and Feb 5, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture3
- Material Separation2
- Nonstandard Device2
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
- Device Markings/Labelling Problem1
- Inflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient3
- No Patient Involvement1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 5, 2026Nonstandard Device; Material Separation
IT WAS REPORTED THAT A DETACHMENT OCCURRED. AN NC QUANTUM APEX MR 2.75 X 20MM WAS SELECTED FOR TREATMENT OF THE TARGET LESION. WHEN REMOVING THE BALLOON FROM THE PROTECTOR, THE PHYSICIAN NOTED THAT THE BALLOON WAS CUT. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24272187 ↗MalfunctionAug 20, 2025Nonstandard Device; Material Separation; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Nonstandard Device; Material Separation; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. DURING PREPARATION OF THE 20MM X 2.75MM NC QUANTUM APEX? BALLOON CATHETER, NO RESISTANCE WAS ENCOUNTERED WHILE REMOVING THE BALLOON PROTECTOR/MANDREL BUT AFTERWARD, IT WAS NOTED THAT THE TIP OF THE DEVICE WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22849498 ↗MalfunctionNov 29, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED RIGHT CORONARY ARTERY. A 20MM X 2.75MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR POST-DILATATION. HOWEVER, DURING THE FOURTH INFLATION AT 20 ATMOSPHERES FOR 15 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A 3.00MMX20MM NC QUANTUM APEX BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15878530 ↗MalfunctionJan 6, 2021Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT LABELING ISSUE OCCURRED. A 20MM X 2.75MM NC QUANTUM APEX BALLOON CATHETER WAS HANDED TO THE REPRESENTATIVE AND EXPLAINED THAT THE BALLOON CONTAINED AN EXPANSION CHART FROM AN ENTIRELY DIFFERENT BALLOON. THE BALLOON WAS NOT USED TO THE PATIENT.
Patient problems: No Patient Involvement
MAUDE report 11125323 ↗MalfunctionMay 11, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 20MM X 2.75MMM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10043419 ↗MalfunctionFeb 6, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 20MM X 2.75MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR POST DILATION. HOWEVER, THE BALLOON RUPTURED AT 14 ATMOSPHERES TO OPTIMIZE THE STENT IMPLANT. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9677202 ↗MalfunctionJun 6, 2019Inflation Problem; Inflation Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2019. IT WAS REPORTED THAT BALLOON INFLATION FAILURE OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA THE FEMORAL ARTERY. THE CONCENTRIC TARGET LESION WAS LOCATED IN THE MILDLY CALCIFIED MID RIGHT CORONARY ARTERY. THE LESION CONTAINED >=90 DEGREES BEND. A 20MM X 2.75MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR DILATATION, BUT THE BALLOON FAILED TO INFLATE. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON WITH BUDDY WIRE SUPPORT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS STABLE AND DOING WELL. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A SEPARATED HYPOTUBE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8674131 ↗
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗