Skip to content

NC Quantum Apex™ H7493912430250Recalls & reported problems

Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2

All models of NC Quantum Apex (48)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729783572), received between Nov 13, 2018 and Jul 1, 2024 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture3
  • Break1
  • Difficult to Remove1
  • Wrong Label1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • Perforation of Vessels1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 1, 2024Material Rupture; Wrong Label; Material Rupture; Wrong Label

    IT WAS REPORTED THAT A LABELING ISSUE OCCURRED. DURING A PROCEDURE, FOLLOWING ROTABLATION, A DEVICE WITH PACKAGE LABEL OF NC QUANTUM APEX 2.5 X 30MM WAS SELECTED FOR USE. HOWEVER, THE INNER PACKAGE CONTAINED A 3.5 X 30MM NC QUANTUM APEX IN SIZE. THE OVERSIZED DEVICE RESULTED IN A CORONARY PERFORATION. THE PERFORATION WAS COVERED WITH A NON-BSC STENT. THE PATIENT WAS STABILIZED AND EXPECTED TO FULLY RECOVER.

    Patient problems: Perforation of Vessels

    MAUDE report 19644072 ↗
  • MalfunctionApr 7, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE MODERATELY CALCIFIED CORONARY ARTERY. A 30MM X 2.50MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE AND THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14048311 ↗
  • InjuryOct 29, 2020Break; Break

    IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED MID LEFT ANTERIOR DESCENDING ARTERY. A 30MM X 2.50MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING REMOVAL, THE DISTAL PORTION REMAINED IN PATIENT DUE TO A BEND. THE DEVICE WAS COMPLETELY REMOVED BY TRAPPING THE WIRE AND INFLATING ANOTHER BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10757988 ↗
  • MalfunctionAug 15, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN A CORONARY ARTERY. A 30MM X 2.50MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION AT 12 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8899771 ↗
  • MalfunctionNov 13, 2018Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT BALLOON REMOVAL DIFFICULTIES WERE ENCOUNTERED. A 30MM X 2.50MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, WHEN THE BALLOON WAS ATTEMPTED TO BE REMOVED FOLLOWING DEFLATION, THE BALLOON DID NOT SLIDE EASILY AND BECAME STUCK TO THE GUIDE. AFTER SEVERAL ATTEMPTS, THE PHYSICIAN MANAGED TO WITHDRAW THE BALLOON. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8067144 ↗

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page