NC Quantum Apex™ H7493912430250Recalls & reported problems
Boston Scientific Corporation · PTCA Dilatation Catheter · Class 2
All models of NC Quantum Apex (48)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729783572), received between Nov 13, 2018 and Jul 1, 2024 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture3
- Break1
- Difficult to Remove1
- Wrong Label1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- Perforation of Vessels1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJul 1, 2024Material Rupture; Wrong Label; Material Rupture; Wrong Label
IT WAS REPORTED THAT A LABELING ISSUE OCCURRED. DURING A PROCEDURE, FOLLOWING ROTABLATION, A DEVICE WITH PACKAGE LABEL OF NC QUANTUM APEX 2.5 X 30MM WAS SELECTED FOR USE. HOWEVER, THE INNER PACKAGE CONTAINED A 3.5 X 30MM NC QUANTUM APEX IN SIZE. THE OVERSIZED DEVICE RESULTED IN A CORONARY PERFORATION. THE PERFORATION WAS COVERED WITH A NON-BSC STENT. THE PATIENT WAS STABILIZED AND EXPECTED TO FULLY RECOVER.
Patient problems: Perforation of Vessels
MAUDE report 19644072 ↗MalfunctionApr 7, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE MODERATELY CALCIFIED CORONARY ARTERY. A 30MM X 2.50MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE AND THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14048311 ↗InjuryOct 29, 2020Break; Break
IT WAS REPORTED THAT SHAFT BREAK OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED MID LEFT ANTERIOR DESCENDING ARTERY. A 30MM X 2.50MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING REMOVAL, THE DISTAL PORTION REMAINED IN PATIENT DUE TO A BEND. THE DEVICE WAS COMPLETELY REMOVED BY TRAPPING THE WIRE AND INFLATING ANOTHER BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10757988 ↗MalfunctionAug 15, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN A CORONARY ARTERY. A 30MM X 2.50MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION AT 12 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8899771 ↗MalfunctionNov 13, 2018Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT BALLOON REMOVAL DIFFICULTIES WERE ENCOUNTERED. A 30MM X 2.50MM NC QUANTUM APEX BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, WHEN THE BALLOON WAS ATTEMPTED TO BE REMOVED FOLLOWING DEFLATION, THE BALLOON DID NOT SLIDE EASILY AND BECAME STUCK TO THE GUIDE. AFTER SEVERAL ATTEMPTS, THE PHYSICIAN MANAGED TO WITHDRAW THE BALLOON. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8067144 ↗
Other Catheters, transluminal coronary angioplasty, percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page