EndoVive Jejunal Feeding Tube M00566340Recalls & reported problems
Boston Scientific Corporation · 3-Port Through-the-PEG J-Tube Kit · Class 2 · Implant
All models of EndoVive Jejunal Feeding Tube (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729789666), received between Mar 13, 2018 and Jan 11, 2023 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Advance1
- Kinked1
- Leak/Splash1
- Material Deformation1
- Obstruction of Flow1
- Occlusion Within Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 11, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE JEJUNAL FEEDING TUBE WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PROCEDURE ON (B)(6) 2022. DURING THE PROCEDURE, ONCE THE JEJUNAL TUBE WAS IN PLACE IT WAS TESTED, A LEAK WAS NOTED IN THE TUBING. THE PROCEDURE WAS COMPLETED WITH A NEW JEJUNAL TUBE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16139419 ↗MalfunctionJul 26, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE JEJUNAL FEEDING TUBE WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PLACEMENT PROCEDURE. THE PROCEDURE DATE IS UNKNOWN. POST PROCEDURE, THE JEJUNAL TUBE SNAPPED. A NEW ENDOVIVE JEJUNAL FEEDING TUBE WAS PLACED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15102506 ↗InjuryMar 25, 2021Detachment of Device or Device Component; Difficult to Advance; Detachment of Device or Device Component; Difficult to Advance
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE THREE-PORT THROUGH THE PEG JEJUNAL TUBE WAS USED DURING A JEJUNAL TUBE PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, WHEN THE JEJUNAL TUBE WAS INSERTED VIA AN ENDOVIVE SAFETY PEG TUBE, THE PLASTIC TIP DETACHED INSIDE THE PATIENT. THE DETACHED PORTION WAS RETRIEVED ENDOSCOPICALLY USING FORCEPS. ALSO, IT WAS NOTICED THAT THE JEJUNAL TUBE WAS DIFFICULT TO ADVANCE INTO THE SMALL INTESTINE. THEY THEN DECIDED TO DISCONTINUE THE PLACEMENT. THE PROCEDURE WAS COMPLETED WITH A NON BSC DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11567546 ↗InjuryMar 4, 2021Material Deformation; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE THREE-PORT THROUGH THE PEG JEJUNAL TUBE WAS BEING USED DURING A GASTROSTOMY REPLACEMENT PROCEDURE PERFORMED ON (B)(6) 2021. ON (B)(6) 2021, POST-PROCEDURE AND BEFORE DISCHARGE, THE NUTRIENTS COULD NOT BE INJECTED INTO THE FEEDING TUBE. AN ENDOSCOPE WAS USED TO SEE THERE WAS A BEND IN THE FEEDING TUBE. THE PHYSICIAN WAS ABLE TO UNBEND THE JEJUNAL TUBE AND THE FLOW RETURNED TO NORMAL. HOWEVER, THE PHYSICIAN DECIDED THE FEEDING TUBE NEEDED TO BE REPLACED WITH A NEW ENDOVIVE THREE-PORT THROUGH THE PEG JEJUNAL TUBE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11409006 ↗MalfunctionMar 30, 2020Obstruction of Flow; Adverse Event Without Identified Device or Use Problem; Obstruction of Flow; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE JEJUNAL FEEDING TUBE WAS PLACED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE JEJUNAL TUBE HAS AN OCCLUSION. THE PROCEDURE WAS COMPLETED WITH A NEW ENDOVIVE JEJUNAL FEEDING TUBE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. ADDITIONAL INFORMATION RECEIVED ON APRIL 6, 2020. THERE WAS A SHARP ANGULATION AT THE DISTAL END OF THE PROBE, WHICH PREVENTED THE PASSAGE OF THE GUIDEWIRE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9898487 ↗MalfunctionJun 4, 2018Occlusion Within Device; Occlusion Within Device
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE THREE-PORT THROUGH THE PEG JEJUNAL TUBE WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY JEJUNOSTOMY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE JEJUNAL TUBE WAS ADVANCED TO THE JEJUNUM. UPON PLACING THE JEJUNAL TUBE IN THE PROPER PLACEMENT, THEY WERE NOT ABLE TO ADVANCE ANY SALINE, FOOD BOLUS OR ANYTHING THROUGH THE FEEDING TUBE AS IT WAS BLOCKED/OCCLUDED. SINCE THE SAME DEVICE IS NOT AVAILABLE, THEY JUST REMOVED THE JEJUNAL TUBE AND KEPT THE PEG TUBE IN PLACE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7567114 ↗MalfunctionMar 13, 2018Kinked; Kinked
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE THREE-PORT THROUGH THE PEG JEJUNAL TUBE WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, AFTER PLACEMENT WAS PERFORMED, TUBE BLOCKAGE OCCURRED FREQUENTLY DUE TO KINK, AND THE KINK ISSUE WAS RESOLVED USING A BIOPSY FORCEPS THROUGH ENDOSCOPE. RECURRENCE OF ISSUE WITHIN A WEEK WAS NOTICED. THE PROCEDURE WAS COMPLETED WITH A NEW ENDOVIVE THREE-PORT THROUGH THE PEG JEJUNAL TUBE. THERE WERE NO PATIENT COMPLICATIONS REPORTED DUE TO THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7335969 ↗
Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-05090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page