Coyote™ H74939186251010Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Coyote (55)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729796503), received between Mar 25, 2019 and Oct 16, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Difficult to Advance” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Advance1
- Difficult to Remove1
- Entrapment of Device1
- Failure to Deflate1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 16, 2025Difficult to Remove; Difficult to Advance; Difficult to Remove; Difficult to Advance
IT WAS REPORTED THAT BALLOON WAS DIFFICULT TO REMOVE. THE TARGET LESION WAS LOCATED IN THE RIGHT POPLITEAL/TIBIAL/PERONEAL TRUNK VESSEL. A 2.5MM X 100MM X 150CM WAS SELECTED FOR USE IN A LOWER EXTREMITY PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY. THE BALLOON EXHIBITED DIFFICULTY TRACKING OVER THE WIRE BOTH DURING ADVANCEMENT AND WITHDRAWAL. THE BALLOON WAS REMOVED, AND THE WIRE WAS FLUSHED AND WIPED, HOWEVER, TRACKING DIFFICULTY PERSISTED. THE PHYSICIAN SWITCHED TO A NON-BSC WIRE, BUT SIMILAR RESISTANCE WAS ENCOUNTERED DURING ADVANCEMENT AND REMOVAL OF THE BALLOON. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH ALTERNATE DEVICE. THERE WAS NO PATIENT COMPLICATION AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23312499 ↗MalfunctionApr 28, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 60% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED ANTERIOR TIBIAL ARTERY. A 2.5MM X 100MM X 150CM COYOTE BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION, IT WAS NOTED THAT THE PRESSURE DROPPED BEFORE 6 ATMOSPHERES AND BLOOD CAME BACK THROUGH THE INFLATION DEVICE. THE DEVICE WAS COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH A 2.5MMX150MMX150CM COYOTE BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS AND THE PATIENT CONDITION WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16834767 ↗MalfunctionJan 25, 2023Failure to Deflate; Failure to Deflate
IT WAS REPORTED THAT DEFLATION ISSUE OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA FEMORAL ARTERY. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED ANTERIOR TIBIAL ARTERY. A 2.5MM X 100MM X 150CM, OTW COYOTE BALLOON CATHETER WAS ADVANCED FOR DILATATION. THE BALLOON WAS INFLATED AT 8 ATMOSPHERES FOR 2 MINUTES WITH A SALINE CONTRAST RATIO OF 3:1. HOWEVER, DURING THE PROCEDURE, THE BALLOON SHOWS PERSISTENTLY INFLATED EVEN AFTER DEFLATING IT. DEFLATION ACTIVITY WAS DONE WITH HELP OF A NON-BOSTON SCIENTIFIC INFLATION DEVICE BY PULLING THE PLUNGER BACK RELEASE THE LOCKING LEVER. SUBSEQUENTLY, THE BALLOON WAS REMOVED THROUGH THE SHEATH CAREFULLY FOR 3 MINUTES. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16243805 ↗MalfunctionMar 25, 2019Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT A FROZE ON WIRE OCCURRED. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE ANTERIOR TIBIAL ARTERY. A 2.5MM X 100MM X 150CM COYOTE BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, THE BALLOON AND A NON BSC GUIDEWIRE WAS STUCK FIRMLY INSIDE THE PATIENT'S BODY AND CANNOT BE WITHDRAWN. BOTH DEVICES WERE REMOVED TOGETHER. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8449908 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page