Skip to content

Coyote™ H74939186350810Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Coyote (55)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729796596), received between Apr 18, 2024 and Apr 18, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Entrapment of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Entrapment of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 18, 2024Entrapment of Device; Entrapment of Device

    IT WAS REPORTED THAT DEVICE ENTRAPMENT OCCURRED. THE PATIENT UNDERWENT A PERIPHERAL ANGIOGRAM ON THE PATIENT'S RIGHT LEG USING A DESTINATION CROSSOVER SHEATH IN THE LEFT GROIN VIA THE SUPERFICIAL FEMORAL ARTERY (SFA) WHICH REVEALED A CHRONIC TOTAL OCCLUSION IN THE P2-P3 SEGMENT OF POPLITEAL ARTERY. AFTER CROSSING THE LESION WITH A NON-BOSTON SCIENTIFIC (BSC) GUIDEWIRE, LASER WAS USED TO PREPARE THE LESION AND BLOOD FLOW WAS CONFIRMED. A 3.5MM X 80MM X 150CM COYOTE BALLOON CATHETER WAS ADVANCED OVER THE WIRE; HOWEVER, ONCE THE COYOTE BALLOON WAS JUST PROXIMAL TO THE TARGET AREA, THE BALLOON SEIZED AND WOULD NOT MOVE NO MATTER HOW HARD IT WAS PUSHED OR PULLED. FLUID WAS INJECTED INTO THE AREA IN ATTEMPT TO GET THE BALLOON TO MOVE OVER THE WIRE, BUT THE BALLOON WAS STUCK. THE BALLOON AND THE WIRE WERE PULLED TOGETHER AND REMOVED OUT OF THE BODY. A NEW NON-BSC GUIDEWIRE WAS PLACED, AND THE LESION WAS DILATED WITH A NON-BSC BALLOON. THE PROCEDURE WAS SUCCESSFULLY COMPLETED, AND THE PATIENT WAS RECOVERING AS NORMAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19139917 ↗